文献速递泌尿生殖MR

反向极性梯度多路敏感编码扩散加权成像在前列腺癌诊断中的价值提升:与单次激发扩散加权成像及多路敏感编码扩散加权成像的比较研究

Improved Value of Multiplexed Sensitivity Encoding DWI with Reversed Polarity Gradients in Diagnosing Prostate Cancer: A Comparison Study with Single-Shot DWI and MUSE DWI.
PubMed 文献 · Acad Radiol · 2023-09-29

本研究纳入73例前列腺疾病患者,比较反向极性梯度多路敏感编码扩散加权成像(RPG-MUSE DWI)、多路敏感编码扩散加权成像(MUSE DWI)与单次激发回波平面扩散加权成像(ssEPI DWI)的图像质量与前列腺癌诊断效能。结果显示RPG-MUSE DWI在伪影抑制、解剖细节、边缘锐利度、信噪比和对比噪声比上优于MUSE和ssEPI,几何畸变率更低,且不影响ADC值测量;RPG-MUSE与MUSE的诊断效能高于ssEPI。

要点

  • RPG-MUSE DWI的图像质量优于MUSE和ssEPI,伪影更少、几何畸变更低。
  • RPG-MUSE DWI的信噪比和对比噪声比高于MUSE和ssEPI。
  • RPG-MUSE与MUSE DWI对前列腺癌的诊断效能高于ssEPI。

核心结论

  • RPG-MUSE可减少前列腺DWI图像的伪影和几何畸变,提高信噪比和对比噪声比,改善解剖细节和边界的清晰度,且不影响ADC值测量,有潜力提高前列腺病变的诊断效能,有助于前列腺病变的清晰显示和准确评估。

研究设计与样本

  • 本研究为对比研究,纳入73例前列腺疾病患者,均行RPG-MUSE、MUSE和ssEPI三种DWI序列扫描;由3名放射科医师进行定性图像分析,2名放射科医师测量信噪比、对比噪声比、几何畸变率和ADC值,2名放射科医师共同进行前列腺影像报告和数据系统评分并比较三种DWI序列对前列腺癌的诊断效能。

关键影像表现/征象

  • RPG-MUSE DWI在伪影最小化、解剖细节和前列腺边缘锐利度方面的评分高于MUSE和ssEPI(P < 0.05)。
  • RPG-MUSE DWI的信噪比和对比噪声比高于MUSE和ssEPI(P < 0.05)。
  • RPG-MUSE DWI的几何畸变率低于MUSE和ssEPI(P < 0.05)。
  • 三种DWI序列的ADC值无统计学差异(P > 0.05)。

诊断标准与关键数值

  • RPG-MUSE与MUSE DWI的诊断效能高于ssEPI(P < 0.017)。
  • 三位放射科医师对三种DWI序列图像评估和测量的一致性良好(组内相关系数 >0.75,P < 0.05)。

临床应用与阅片要点

  • RPG-MUSE DWI可减少前列腺DWI图像的伪影和几何畸变,提高信噪比和对比噪声比。
  • RPG-MUSE DWI改善解剖细节和边界清晰度,且不影响ADC值测量。
  • RPG-MUSE DWI有潜力提高前列腺病变的诊断效能,有助于前列腺病变的清晰显示和准确评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: This study aimed to investigate the value of multiplexed sensitivity encoding with reversed polarity gradients in improving the quality of diffusion-weighted imaging (DWI) images of the prostate and the diagnostic efficacy of prostate cancer.

MATERIALS AND METHODS: Seventy-three patients with prostate disease underwent multiplexed sensitivity encoding with reversed polarity gradients (RPG-MUSE), multiplexed sensitivity encoding (MUSE), and single-shot echo-planar imaging (ssEPI) DWI. Three radiologists performed a qualitative image analysis of the three DWI sequences. Qualitative image analysis included artifact minimization, anatomical detail, and sharpness of prostate edges. Two radiologists measured the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), geometric distortion rate, and the apparent diffusion coefficient (ADC) values of the prostate disease tissue. Two radiologists jointly performed Prostate Imaging Reporting and Data System scoring of prostate lesions and compared the diagnostic efficacy of the three DWI sequences for prostate cancer.

RESULTS: There was good agreement among radiologists in the evaluation and measurement of the three DWI sequence images (intraclass correlation coefficient >0.75, P < 0.05). The RPG-MUSE DWI images were rated higher than those of MUSE and ssEPI in terms of artifact minimization, anatomical details, and sharpness of prostate edges (P < 0.05). The SNR and CNR of the RPG-MUSE DWI images were higher than those of MUSE and ssEPI (P < 0.05), and the geometric distortion rate was lower than that of the other two sequences (P < 0.05). There were no statistical differences in ADC values between the three DWI sequences (P > 0.05). The diagnostic efficacy of RPG-MUSE and MUSE DWI was higher than that of ssEPI (P < 0.017).

CONCLUSION: RPG-MUSE can reduce the artifacts and geometric distortion in DWI images of the prostate, improve the SNR and CNR of the images, improve the clarity of anatomical details and boundaries without affecting the measurement of ADC values, has the potential to improve the diagnostic efficacy of prostate lesions, and facilitates the clear display and accurate assessment of prostate lesions.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Zhao W, Ju S, Yang H, et al. Improved Value of Multiplexed Sensitivity Encoding DWI with Reversed Polarity Gradients in Diagnosing Prostate Cancer: A Comparison Study with Single-Shot DWI and MUSE DWI. Acad Radiol. 2023. PMID: 37778902. doi:10.1016/j.acra.2023.08.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。