肝铁过载患者磁共振弹性成像的技术成功率与可靠性
回顾性纳入37例经R2*值确认的肝铁过载患者,比较二维自旋回波(SE)MRE与二维SE-EPI MRE的成功率、质量评分及肝硬度。SE MRE总体成功率91.89%,SE-EPI MRE为78.38%;重度铁过载时SE MRE成功率100%,SE-EPI MRE为78.57%。轻中度铁过载时两种序列均可提供可靠肝硬度测量,重度铁过载时SE MRE可能更优。
要点
- 肝铁过载患者SE MRE总体成功率91.89%,SE-EPI MRE为78.38%。
- 重度铁过载时SE MRE成功率100%,SE-EPI MRE为78.57%。
- 轻中度铁过载时两种MRE序列肝硬度测量无显著差异。
核心结论
- 轻度和中度铁过载肝脏中,SE MRE与快速SE-EPI MRE均可提供成功且可靠的肝硬度测量;重度铁过载肝脏中,SE MRE可能优于SE-EPI MRE。
研究设计与样本
- 回顾性研究,纳入37例经R2*值测量确认的肝铁过载患者,均行二维自旋回波(SE)MRE和二维SE-EPI MRE检查。
诊断标准与关键数值
- SE MRE与SE-EPI MRE在肝铁过载患者中的总体成功率分别为91.89%和78.38%。
- 重度铁过载患者中SE MRE与SE-EPI MRE成功率分别为100%和78.57%。
- 全部患者MRE质量评分SE MRE为54、SE-EPI MRE为48(P = 0.107)。
- 轻、中、重度铁过载肝脏中两种MRE方法肝硬度测量均无显著差异(P > 0.6)。
- 两种MRE方法中R2*值对肝硬度测量均无显著影响(相关系数<0.1,P >0.6)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: This study aimed to evaluate the technical success rate and stiffness measurement reliability of two specific hepatic magnetic resonance elastography (MRE) sequences dedicated to solving susceptibility artifacts in patients with various degrees of hepatic iron overload.
MATERIALS AND METHODS: Thirty-seven patients with iron-overloaded liver confirmed by R2* value measurement who underwent two-dimensional (2D) spin-echo (SE) MRE and 2D SE-echo-planar-imaging (EPI) MRE were reviewed retrospectively. According to four categories based on R2* value (mild, moderate, severe elevation, and extremely severe iron overload), we compared the success rate, quality score, and liver stiffness of the two sequences. In addition, Spearman's correlation was performed to evaluate the relationship between the R2* value and liver stiffness.
RESULTS: The overall success rates of SE MRE and SE-EPI MRE in patients with hepatic iron overload were 91.89% and 78.38%, respectively, and 100% and 78.57%, respectively, for severe elevation iron overload. In all patients, the MRE quality scores were 54 and 48 for SE MRE and SE-EPI MRE, respectively (P = 0.107). There were no significant differences in liver stiffness measurements between the two MRE methods in patients with mild, moderate, and severe elevation iron-overloaded livers (P > 0.6 for all), respectively. For both MRE methods, R2* value had no significant effect on the liver stiffness measurements (correlation coefficient <0.1, P >0.6 for both).
CONCLUSION: In the mild and moderate elevation iron-overloaded liver, both SE MRE and fast SE-EPI MRE can provide successful and reliable liver stiffness measurement. In severe elevation iron-overloaded livers, SE MRE may be a better choice than SE-EPI MRE.