文献速递技术与安全腹部US

谐波运动成像引导聚焦超声消融:三种聚焦超声干扰滤波方法的比较

Harmonic Motion Imaging-Guided Focused Ultrasound Ablation: Comparison of Three Focused Ultrasound Interference Filtering Methods.
PubMed 文献 · Ultrasound Med Biol · 2023-10-21

该研究在离体犬肝脏的20次谐波运动成像引导聚焦超声消融采集中,比较陷波滤波、交错脉冲序列和FUS-net三种滤除聚焦超声干扰的方法。交错法在所有测试条件下均显著避免干扰,位移振幅对比噪声比显著高于另两法;FUS-net与陷波滤波的位移对比噪声比无显著趋势,但FUS-net的B模式均方误差显著高于陷波滤波,提示其作为干扰滤波方法的进一步潜力。

要点

  • 交错HMIgFUS在高声压和长时程消融监测中避免聚焦超声干扰的稳健性显著优于陷波滤波和FUS-net。
  • 交错法位移振幅图的对比噪声比在所有测试条件下均显著高于陷波滤波和FUS-net数据。
  • FUS-net与陷波滤波的位移对比噪声比之间无显著趋势,但FUS-net的B模式均方误差显著高于陷波滤波,提示其进一步潜力。

核心结论

  • 交错HMIgFUS在HMIgFUS监测中避免聚焦超声干扰具有稳健性,FUS-net和陷波滤波各有优势、局限和未来潜力。

研究设计与样本

  • 离体犬肝脏研究,共20次HMIgFUS采集,峰值正压范围11至18 MPa,持续时间范围60至180秒,比较陷波滤波、交错HMIgFUS和FUS-net三种聚焦超声干扰滤波方法。

诊断标准与关键数值

  • 交错法在所有测试病例中显著避免聚焦超声干扰,尤其高声压和长持续时间病例;交错法位移振幅图对比噪声比在所有病例中显著高于陷波滤波和FUS-net;FUS-net与陷波滤波的位移对比噪声比之间无显著趋势;FUS-net与交错法比较的B模式均方误差显著高于陷波滤波与交错法比较。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Harmonic motion imaging (HMI) is an acoustic radiation force-based elasticity imaging technique, which can be used to monitor changes in tissue mechanical properties caused by focused ultrasound (FUS)-induced thermal ablation. In conventional HMI, the amplitude-modulated FUS sequence and imaging pulse are transmitted simultaneously. With this method, the high-amplitude FUS signal must be separated from the imaging data for tissue displacement estimation. Frequency domain notch filtering and bandpass filtering were previously used to filter FUS interference from imaging data. However, FUS interference becomes more pronounced at the increased intensities and durations necessary for thermal ablation, rendering frequency domain filtering challenging. In this study, three methods for FUS interference filtering during HMI-guided FUS ablation (HMIgFUS) were compared. The methods were notch filtering; interleaved HMIgFUS, with interleaved FUS and imaging pulses; and FUS-net, a convolutional neural network-based U-Net autoencoder developed by our group for FUS interference filtering in radiofrequency data. FUS-net was applied here for the first time for the purpose of ablation displacement monitoring.

METHODS: The three filtering methods were tested during 20 HMIgFUS acquisitions in an ex vivo canine liver using a range of peak positive pressures from 11 to 18 MPa and durations ranging from 60 to 180 seconds. The B-mode mean squared errors (MSEs) and displacement amplitude contrast-to-noise ratios (CNRs) of the three methods were calculated and compared.

RESULTS: The interleaved method for HMIgFUS was found to be significantly robust in avoiding FUS interference in all tested cases for FUS ablation monitoring, especially cases with high FUS pressure and long durations, as opposed to traditional notch filtering and FUS-net filtering. CNRs obtained from displacement amplitude maps in the interleaved data set were significantly higher in all cases than those obtained from the notch filtered and FUS-net data sets. There was not a significant trend in displacement CNR between the FUS-net and notch filtered data sets. However, B-mode MSE was found to be significantly higher when comparing the FUS-net and interleaved data sets as opposed to the notch filtered and interleaved data sets, suggesting further potential of FUS-net as an FUS interference filtering method.

CONCLUSION: These findings indicate the robustness of interleaved HMIgFUS in avoiding FUS interference during HMIgFUS monitoring and the advantages, limitations and future potential of FUS-net and notch filtering.