肺超声技术与胸膜线深度对胸膜线厚度体外和体内测量的影响

Effects of Lung Ultrasound Technique and Pleural Line Depth on In Vitro and In Vivo Measurements of Pleural Line Thickness.
PubMed 文献 · Ultrasound Med Biol · 2023-10-10

该研究在体模和健康志愿者中评估成像方案、探头类型及胸膜线深度对胸膜线厚度(PLT)测量的影响。体外(n=250幅声像图)和体内(n=160幅声像图)均显示,相控阵探头测得的PLT大于线阵探头,且PLT随胸膜线深度增加而增大;胸膜线与探头夹角及焦距不影响PLT。作者认为未来将PLT作为疾病标志物的研究应校正这些变量的混杂影响。

要点

  • 相控阵探头测得的胸膜线厚度大于线阵探头,体外增加3.8 mm,体内增加2.56 mm。
  • 胸膜线深度每增加1 cm,PLT在相控阵探头下增加0.96 mm(体外)和0.67 mm(体内)。
  • 胸膜线与探头夹角及焦距未改变PLT测量值。

核心结论

  • PLT测量受探头类型和胸膜线深度影响,未来评估PLT作为疾病标志物的研究应校正这些变量的混杂影响。

研究设计与样本

  • 在体模和健康志愿者中进行的体外与体内声像图研究,体外n=250幅声像图,体内n=160幅声像图;使用5-1 MHz相控阵探头与13-6 MHz线阵探头,以电子卡尺测量PLT,采用回归方程描述自变量对PLT的影响。

关键影像表现/征象

  • 体外相控阵探头测得的PLT较线阵探头大3.8 mm(标准误±0.24,p<0.001)。
  • 体内相控阵探头测得的PLT较线阵探头大2.56 mm(±0.06,p<0.001)。
  • 胸膜线与探头夹角及焦距均未改变PLT。

诊断标准与关键数值

  • 体外胸膜线深度每增加1 cm,PLT在相控阵探头下增加0.96 mm(±0.081,p<0.001),在线阵探头下增加0.26 mm(±0.019,p<0.001)。
  • 体内探头与胸膜间距每增加1 cm,PLT在相控阵探头下增加0.67 mm(±0.060,p<0.001),在线阵探头下增加0.25 mm(±0.019,p<0.001)。
  • 体内变量共同解释PLT变异的93%(p<0.001)。

临床应用与阅片要点

  • 作者认为未来将PLT作为疾病标志物的研究应校正探头类型和胸膜线深度这两个变量的混杂影响。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The aim of the work described here was to determine the effects of imaging protocol, technique and pleural line depth on measured pleural line thickness (PLT).

METHODS: Sonograms were performed on a phantom and healthy volunteers. In vitro, pleural line depth, transducer type (5-1 MHz phased array vs. 13-6 MHz linear array), angle of the pleural line relative to the transducer and distance between the pleural line and focal length were explicitly modified. PLT was measured using electronic calipers. Regression equations described the effects of independent variables on PLT. Factors influencing PLT in vitro were tested in vivo.

RESULTS: In vitro (n = 250 sonograms), PLT was 3.8 (standard error: ±0.24) mm greater when using the phased array compared with the linear transducer (p < 0.001). For every additional centimeter of pleural line depth, PLT increased by 0.96 (±0.081) mm for the phased array (p < 0.001) and 0.26 (±0.019) mm for the linear transducer (p < 0.001). Neither pleural angle nor focal length altered PLT. In vivo (n = 160 sonograms), PLT was 2.56 (±0.06) mm greater when using the phased array (p < 0.001) compared with the linear transducer. PLT increased by 0.67 (±0.060) mm with the phased array (p < 0.001) and 0.25 mm (±0.019) with the linear transducer (p < 0.001) for every additional centimeter between the transducer and the pleura. Together the variables explained 93% of PLT variance in vivo (p < 0.001).

CONCLUSION: PLT measurements are affected by transducer type and pleural line depth. Future studies evaluating PLT as a disease marker should account for confounding by these variables.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Fox TH, Chansangavej S, Kirby K, et al. Effects of Lung Ultrasound Technique and Pleural Line Depth on In Vitro and In Vivo Measurements of Pleural Line Thickness. Ultrasound Med Biol. 2023. PMID: 37821245. doi:10.1016/j.ultrasmedbio.2023.09.010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。