聚乙烯醇异质性斑块体模:利用脉搏波成像估测狭窄区域脉搏波速度

Polyvinyl Alcohol Phantoms With Heterogeneous Plaques: Estimation of Pulse Wave Velocity at the Stenotic Region Using Pulse Wave Imaging.
PubMed 文献 · Ultrasound Med Biol · 2023-10-12

研究制备了5个含不同硬度异质性斑块的聚乙烯醇狭窄动脉体模,用脉搏波成像估测脉搏波速度。均质软斑块、异质性软加中间斑块、均质中间斑块、异质性硬加中间斑块和均质硬斑块的脉搏波速度分别为2.06、2.21、2.49、2.67和3.31 m/s。结果表明脉搏波成像可区分不同硬度和组成的斑块。

要点

  • 脉搏波成像在狭窄动脉体模中估测的脉搏波速度随斑块硬度增加而升高。
  • 均质软斑块脉搏波速度2.06 m/s,均质硬斑块3.31 m/s,提示可区分不同硬度斑块。
  • 异质性斑块脉搏波速度介于其组成成分的均质斑块之间,提示可反映斑块组成。

核心结论

  • 研究报道了一种新型动脉体模制造技术,可制备不同硬度和组成的异质性斑块。脉搏波成像在狭窄动脉中评估斑块硬度的可行性得到确定,且发现其可区分不同硬度和组成的斑块。

研究设计与样本

  • 研究制备了5个狭窄度50%的体模,通过将聚乙烯醇溶液倒入3-D打印模具制成;其中2个为异质性斑块,分别为软材料(E0=13 kPa)与中间材料(E0=40 kPa)组成、硬材料(E0=54 kPa)与中间材料组成。体模连接脉动泵后采集超声序列,采用归一化互相关追踪脉搏诱导的体模壁扩张传播,通过拟合壁扩张波形峰值加速度到达时间与对应位置的线性回归线估测脉搏波速度。

关键影像表现/征象

  • 均质软斑块体模的脉搏波速度为2.06 m/s。
  • 异质性软加中间斑块体模的脉搏波速度为2.21 m/s。
  • 均质中间斑块体模的脉搏波速度为2.49 m/s。
  • 异质性硬加中间斑块体模的脉搏波速度为2.67 m/s。
  • 均质硬斑块体模的脉搏波速度为3.31 m/s。

诊断标准与关键数值

  • 脉搏波速度估测方法为拟合壁扩张波形峰值加速度到达时间与对应位置的线性回归线。
  • 体模狭窄度为50%。
  • 软材料弹性模量E0=13 kPa,中间材料E0=40 kPa,硬材料E0=54 kPa。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Plaque characterization is essential for stroke prevention. In the study reported herein, we describe a heterogeneous phantom manufacturing technique with varying plaque compositions of different stiffness using polyvinyl alcohol (PVA) to emulate stenotic arteries and evaluated the use of pulse wave imaging (PWI) to assess plaque stiffness by comparing derived pulse wave velocities, with the goal of assessing plaque vulnerability and identifying high-risk patients for stroke.

METHODS: Five stenotic phantoms (50% stenosis) were fabricated by pouring PVA solutions into 3-D-printed molds. Two of the phantoms had heterogeneous plaque compositions of soft (E0 = 13 kPa) and intermediate (E0 = 40 kPa) materials and of stiff (E0 = 54 kPa) and intermediate materials. Ultrasound sequences were acquired as the arterial phantoms were connected to a pulsating pump, and PWI was performed on the ultrasound acquisition using normalized cross-correlation to track the pulse-induced phantom wall distension propagations. Pulse wave velocities were estimated by fitting a linear regression line between the arrival time of the peak acceleration of the wall distension waveform and the corresponding location.

RESULTS: Arterial phantoms with heterogeneous plaque stiffness were successfully fabricated. Pulse wave velocities of 2.06, 2.21, 2.49, 2.67 and 3.31 m/s were found in the phantom experiments using PWI for homogeneous soft plaque, the heterogeneous soft and intermediate plaque, homogeneous intermediate plaque, the heterogeneous stiff and intermediate plaque and homogeneous stiff plaque, respectively.

CONCLUSION: A novel arterial phantom building technique was reported with varying heterogenous plaque compositions of different stiffness. The feasibility of using PWI to evaluate plaque stiffness in stenotic arteries was determined and found that PWI can distinguish between plaques of distinct stiffness and composition.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。