四维CT评估不稳定性舟月分离诊断准确性的前瞻性ACTION试验
这项前瞻性研究纳入临床怀疑舟月韧带撕裂的连续受试者,以腕关节镜为参考标准,计算四维CT诊断不稳定性舟月韧带撕裂的敏感度和特异度,并与历史影像增强透视样本进行比较;四维CT敏感度为74.3%、特异度为80.0%,阅片者间一致性良好至优秀。
要点
- 四维CT诊断不稳定性舟月韧带撕裂的敏感度为74.3%、特异度为80.0%。
- 四维CT阅片者间测量一致性ICC范围为0.89至0.96,可重复性高。
- 腕部四维CT有效辐射剂量为67至72 mSv,头部小于1 mSv。
核心结论
- 四维CT结果具有高度可重复性;在探索性试验中,诊断舟月关节间隙大于3 mm的不稳定关节的敏感度为74.3%、特异度为80%。
- 作者认为需要更大规模随机试验的进一步证据。
研究设计与样本
- 前瞻性研究,纳入2020年7月至2022年5月临床怀疑舟月韧带撕裂并接受四维CT的连续受试者40例(平均年龄43岁±14[SD],24例男性)。
- 以历史研究中经影像增强透视诊断的78例患者(平均年龄45岁±11,50例男性)作为对照,腕关节镜为参考标准。
关键影像表现/征象
- 四维CT和影像增强透视均将舟月关节间隙大于3 mm判读为不稳定。
诊断标准与关键数值
- 四维CT在35例不稳定性撕裂中检出26例,敏感度74.3%(95%CI:56.7,87.5)。
- 影像增强透视在63例不稳定性撕裂中检出52例,敏感度82.5%(95%CI:70.9,91)。
- 四维CT在5例稳定舟月关节中识别4例,特异度80.0%(95%CI:28.4,99.5)。
- 影像增强透视在15例稳定舟月韧带中识别12例,特异度80.0%(95%CI:51.9,95.7)。
- 四维CT测量阅片者间一致性ICC范围为0.89至0.96。
临床应用与阅片要点
- 四维CT和影像增强透视均以舟月关节间隙大于3 mm作为不稳定的判读阈值。
- 四维CT的有效辐射剂量在腕部为67至72 mSv,在头部小于1 mSv。
局限性
- 作者指出该探索性试验需要更大规模随机试验的进一步证据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Timely treatment of scapholunate instability depends on early identification, but current imaging methods are either intricate or fail to demonstrate the dynamic stages. Purpose To calculate the diagnostic accuracy of four-dimensional (4D) CT for diagnosing instable scapholunate ligament (SLL) tears. Materials and Methods This prospective study enrolled consecutive participants with clinically suspected SLL tears who underwent 4D CT from July 2020 to May 2022. A historical study sample diagnosed at cineradiography served as a comparison, and wrist arthroscopy was the reference standard. Scapholunate joints greater than 3 mm were interpreted as instable at index 4D CT and cineradiography. Diagnostic accuracy was expressed as sensitivity and specificity. Areas under the receiver operating characteristic curve and cutoff values for both index tests were calculated. Intraclass correlation coefficients (ICCs) were computed to compare interrater reliability. Effective radiation doses at 4D CT were measured with thermoluminescent dosimeters. Results The study included 40 participants (mean age, 43 years ± 14 [SD]; 24 male) evaluated at 4D CT and 78 patients (mean age, 45 years ± 11; 50 male) historically evaluated at cineradiography. Four-dimensional CT helped detect instable tears in 26 of 35 participants (sensitivity, 74.3% [95% CI: 56.7, 87.5]. Cineradiography revealed instable tears in 52 of 63 patients (sensitivity, 82.5% [95% CI: 70.9, 91]). Four of five participants with stable scapholunate joints were identified at 4D CT (specificity, 80.0% [95% CI: 28.4, 99.5]), and 12 of 15 patients with stable SLLs were identified at cineradiography (specificity, 80.0% [95% CI: 51.9, 95.7]). Interrater agreement of radiologic measurements on 4D CT scans was good to excellent (ICC range, 0.89-0.96). The effective radiation dose ranged from 67 to 72 mSv at the wrist and was less than 1 mSv at the head. Conclusion Four-dimensional CT results are highly reproducible. Instable scapholunate joints greater than 3 mm were detected with a sensitivity of 74.3% and a specificity of 80% in an exploratory trial. Further evidence from larger randomized trials is warranted. German Register for Clinical Trials no. DRKS00021110 (Universal Trial Number U1111-1249-7884) Published under a CC BY 4.0 license. Supplemental material is available for this article. See also the editorial by Demehri and Ibad in this issue.