双层探测器头部CT在小儿患者中维持图像质量的同时降低辐射剂量
回顾性评估106例双层探测器CT小儿头部平扫,分250 mAs与180 mAs两组各53例;60 keV虚拟单能图像较常规图像噪声更低、SNR与CNR更高且主观评分更优,两组间60 keV虚拟单能图像质量无显著差异,180 mAs组剂量参数显著更低。
要点
- 双层探测器CT 60 keV虚拟单能图像可较常规图像降低噪声并提高SNR与CNR。
- 180 mAs组剂量长度乘积显著低于250 mAs组,60 keV虚拟单能图像质量两组间无显著差异。
- 作者结论为60 keV虚拟单能图像可在小儿颅脑CT中降低28%辐射剂量而不损失图像质量。
核心结论
- 双层探测器CT 60 keV虚拟单能图像在小儿颅脑CT中可实现28%的辐射剂量降低且无图像质量损失。
研究设计与样本
- 回顾性评估106例经双层探测器CT采集的小儿头部平扫,分为250 mAs组53例与180 mAs组53例,比较剂量长度乘积、噪声、SNR、CNR及两名盲法放射科医师的5分Likert主观评分,统计采用ANOVA与Wilcoxon检验并校正多重比较。
诊断标准与关键数值
- 250 mAs组平均剂量长度乘积717.47(SD,41.52)mGy×cm,180 mAs组平均剂量长度乘积520.74(SD,42)mGy×cm;60 keV虚拟单能图像较常规图像噪声显著更低、SNR与CNR显著更高、主观评分显著更优(均P < .001);60 keV虚拟单能图像的SNR、CNR及主观评分在两组间无显著差异(P > .05);180 mAs组辐射剂量参数显著低于250 mAs组(P < .001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Radiation exposure in the CT diagnostic imaging process is a conspicuous concern in pediatric patients. This study aimed to evaluate whether 60-keV virtual monoenergetic images of the pediatric cranium in dual-layer CT can reduce the radiation dose while maintaining image quality compared with conventional images.
MATERIALS AND METHODS: One hundred six unenhanced pediatric head scans acquired by dual-layer CT were retrospectively assessed. The patients were assigned to 2 groups of 53 and scanned with 250 and 180 mAs, respectively. Dose-length product values were retrieved, and noise, SNR, and contrast-to-noise ratio were calculated for each case. Two radiologists blinded to the reconstruction technique used evaluated image quality on a 5-point Likert scale. Statistical assessment was performed with ANOVA and the Wilcoxon test, adjusted for multiple comparisons.
RESULTS: Mean dose-length product values were 717.47 (SD, 41.52) mGy×cm and 520.74 (SD, 42) mGy×cm for the 250- and 180-mAs groups, respectively. Irrespective of the radiation dose, noise was significantly lower, SNR and contrast-to-noise ratio were significantly higher, and subjective analysis revealed significant superiority of 60-keV virtual monoenergetic images compared with conventional images (all P < .001). SNR, contrast-to-noise ratio, and subjective evaluation in 60-keV virtual monoenergetic images were not significantly different between the 2 scan groups (P > .05). Radiation dose parameters were significantly lower in the 180-mAs group compared with the 250-mAs group (P < .001).
CONCLUSIONS: Dual-layer CT 60-keV virtual monoenergetic images allowed a radiation dose reduction of 28% without image-quality loss in pediatric cranial CT.
引用
- Tan Z, Zhang L, Sun X, et al. Dual-Layer Detector Head CT to Maintain Image Quality While Reducing the Radiation Dose in Pediatric Patients. AJNR Am J Neuroradiol. 2023. PMID: 37735089. doi:10.3174/ajnr.A7999
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。