磁共振弹性成像无创检测肝纤维化:三维采集是否具有附加价值
这项单中心前瞻性横断面研究纳入49例慢性肝病并同期肝活检患者,比较2D与3D磁共振弹性成像参数及FIB-4对进展期纤维化(F3-F4)和肝硬化(F4)的诊断效能。2D与3D肝硬度诊断效能相似,3D附加参数效能不一,FIB-4诊断效能较低。
要点
- 2D与3D肝硬度诊断F3-F4的AUC分别为0.87与0.88,诊断F4均为0.81,3D未见附加价值。
- 3D附加参数(储能模量、损耗模量、阻尼比)诊断F3-F4的AUC范围为0.59-0.78,F4为0.52-0.70。
- FIB-4诊断F3-F4与F4的AUC分别为0.66与0.68,低于2D及3D肝硬度。
核心结论
- 与2D磁共振弹性成像相比,3D磁共振弹性成像检测进展期肝纤维化和肝硬化未见附加价值,而FIB-4诊断效能较低。
研究设计与样本
- 单中心、前瞻性、经IRB批准的横断面研究,纳入49例慢性肝病并同期肝活检患者(男/女23/26,平均年龄50.8岁),采用1.5T自旋回波-EPI序列行磁共振弹性成像。
诊断标准与关键数值
- 2D肝硬度诊断F3-F4的AUC为0.87,敏感度为0.71,特异度为0.89;3D肝硬度AUC为0.88,敏感度为0.81,特异度为0.89。
- 2D肝硬度诊断F4的AUC为0.81,敏感度为0.75,特异度为0.84;3D肝硬度AUC为0.81,敏感度为0.83,特异度为0.73。
- 3D附加参数诊断F3-F4的AUC范围为0.59-0.78,诊断F4的AUC范围为0.52-0.70。
- FIB-4诊断F3-F4的AUC为0.66,诊断F4的AUC为0.68。
临床应用与阅片要点
- 原文结论显示3D磁共振弹性成像较2D磁共振弹性成像在检测进展期肝纤维化和肝硬化方面未见附加价值。
- 原文结论显示FIB-4在检测进展期肝纤维化和肝硬化方面诊断效能较低。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: (1) Assess the diagnostic performance of liver 3D magnetic resonance elastography (MRE) parameters (including stiffness, storage/loss modulus and damping ratio) compared to liver stiffness measured with 2D MRE for noninvasive detection of advanced liver fibrosis (F3-F4) and cirrhosis (F4) in patients with chronic liver disease. (2) Assess the value of serum markers (FIB-4) in detecting advanced liver fibrosis and cirrhosis in the same patients.
METHODS: This was a single center, prospective IRB-approved cross-sectional study that included 49 patients (M/F: 23/26, mean age 50.8 y) with chronic liver disease and concomitant liver biopsy. MRE was acquired at 1.5T using a spin echo-EPI sequence. The following parameters were measured: liver stiffness using 2D MRE (LS-2D) and 3D MRE parameters (LS-3D, liver storage, loss modulus and damping ratio). The Mann-Whitney U test, ROC curve analysis, Spearman correlation and logistic regression were performed to evaluate diagnostic performance of MRE parameters and FIB-4.
RESULTS: LS-2D and LS-3D had similar diagnostic performance for diagnosis of F3-F4, with AUCs of 0.87 and 0.88, sensitivity of 0.71 and 0.81, specificity of 0.89 for both. For diagnosis of F4, LS-2D and LS-3D had similar performance with AUCs of 0.81 for both, sensitivity of 0.75 and 0.83, and specificity of 0.84 and 0.73, respectively. Additional 3D parameters (storage modulus, loss modulus, damping ratio) had variable performance, with AUC range of 0.59-0.78 for F3-F4; and 0.52-0.70 for F4. FIB-4 had lower diagnostic performance, with AUCs of 0.66 for F3-F4, and 0.68 for F4.
CONCLUSION: Our study shows no added value of 3D MRE compared to 2D MRE for detection of advanced fibrosis and cirrhosis, while FIB-4 had lower diagnostic performance.