基于压缩感知与共享速度编码的高度加速自由呼吸实时二维血流成像

Highly accelerated free-breathing real-time 2D flow imaging using compressed sensing and shared velocity encoding.
PubMed 文献 · Eur Radiol · 2023-09-02

研究在3T临床系统上评估压缩感知(CS)加速的二维实时相位对比(PC)MRI技术,采用共享速度编码(SVE),在体模(n=8)和健康志愿者(n=7)中与常规GRPPA(R=2)及全采样PC序列比较血流参数。结果显示体模中相关性极好(R²≥0.93,ICC≥0.97),志愿者中相关性良好(R²≥0.80,ICC≥0.90),但CS实时技术对最大流速和流率约低估12%。该技术无需屏气即可逐搏评估血流,适用于心律失常或屏气能力差的患者。

要点

  • CS实时PC技术无需屏气,可在1–5个心动周期内以51–56.5 ms时间分辨率完成逐搏血流测量。
  • 体模中CS实时与常规PC及全采样序列相关性极好(R²≥0.93,ICC≥0.97),平均差异可忽略。
  • 健康志愿者中相关性良好(R²≥0.80,ICC≥0.90),但CS实时对最大流速和流率约低估12%。

核心结论

  • 高度加速的CS实时PC技术无需屏气即可评估血流模式,可对心律失常或屏气能力差的患者进行快速血流评估。

研究设计与样本

  • 原型研究,采用单次激发快速射频扰相梯度回波序列结合对称速度编码的SVE,在3T临床系统上对主动脉夹层体模(n=8)和健康志愿者(n=7)进行CS实时(R=8)、常规GRPPA(R=2)及全采样PC序列扫描,时间分辨率51–56.5 ms,采集1–5个心动周期。

关键影像表现/征象

  • CS实时PC技术采用单次激发快速射频扰相梯度回波与SVE,时间分辨率51–56.5 ms,可在1–5个心动周期内完成采集。
  • 体模中所有位置血流测量与常规及全采样序列比较,相关性极好(所有R²≥0.93),一致性极好(所有ICC≥0.97),平均差异可忽略。
  • 健康志愿者中CS实时与常规及全采样序列相关性良好(所有R²≥0.80),一致性良好(所有ICC≥0.90),但CS实时对最大流速和流率约低估12%。

诊断标准与关键数值

  • 体模研究:CS实时、常规GRPPA(R=2)与全采样PC序列比较,所有R²≥0.93,所有ICC≥0.97,平均差异可忽略。
  • 志愿者研究:CS实时与常规及全采样序列比较,所有R²≥0.80,所有ICC≥0.90,最大流速和流率约低估12%。

临床应用与阅片要点

  • CS实时PC技术无需屏气,适用于心律失常或屏气能力差患者的逐搏血流评估。
  • 临床相关性声明指出,自由呼吸实时血流MRI技术相比常规PC技术,在心律失常尤其心房颤动患者中图像质量更优。
  • 作者指出CS实时技术对最大流速和流率约低估12%,解读结果时需注意此偏差。

局限性

  • CS实时技术对健康志愿者的最大流速和流率约低估12%。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: 2D real-time (RT) phase-contrast (PC) MRI is a promising alternative to conventional PC MRI, which overcomes problems due to irregular heartbeats or poor respiratory control. This study aims to evaluate a prototype compressed sensing (CS)-accelerated 2D RT-PC MRI technique with shared velocity encoding (SVE) for accurate beat-to-beat flow measurements.

METHODS: The CS RT-PC technique was implemented using a single-shot fast RF-spoiled gradient echo with SVE by symmetric velocity encoding, and acquired with a temporal resolution of 51-56.5 ms in 1-5 heartbeats. Both aortic dissection phantom (n = 8) and volunteer (n = 7) studies were conducted using the prototype CS RT (CS, R = 8), the conventional (GRAPPA, R = 2), and the fully sampled PC sequences on a 3T clinical system. Flow parameters including peak velocity, peak flow rate, net flow rate, and maximum velocity were calculated to compare the performance between different methods using linear regression, intraclass correlation (ICC), and Bland-Altman analyses.

RESULTS: Comparisons of the flow measurements at all locations in the phantoms demonstrated an excellent correlation (all R2 ≥ 0.93) and agreement (all ICC ≥ 0.97) with negligible means of differences. In healthy volunteers, a similarly good correlation (all R2 ≥ 0.80) and agreement (all ICC ≥ 0.90) were observed; however, CS RT slightly underestimated the maximum velocities and flow rates (~ 12%).

CONCLUSION: The highly accelerated CS RT-PC technique is feasible for the evaluation of flow patterns without requiring breath-holding, and it allows for rapid flow assessment in patients with arrhythmia or poor breath-hold capacity.

CLINICAL RELEVANCE STATEMENT: The free-breathing real-time flow MRI technique offers improved spatial and temporal resolutions, as well as the ability to image individual cardiac cycles, resulting in superior image quality compared to the conventional PC technique when imaging patients with arrhythmias, especially those with atrial fibrillation.

KEY POINTS: • The highly accelerated prototype CS RT-PC MRI technique with improved temporal resolution by the concept of SVE is feasible for beat-to-beat flow evaluation without requiring breath-holding. • The results of the phantom and in vivo quantitative flow evaluation show the ability of the prototype CS RT-PC technique to obtain reliable flow measurements similarly to the conventional PC MRI. • With less than 12% underestimation, excellent agreements between the two techniques were shown for the measurements of peak velocities and flow rates.