文献速递胸部CTPET

双源与分束滤线器多能CT评估肺气肿肺叶灌注:与SPECT/CT的头对头比较

Head-to-Head Comparison of Dual-Source and Split-Beam Filter Multi-Energy CT versus SPECT/CT for Assessing Lobar Lung Perfusion in Emphysema.
PubMed 文献 · Radiol Cardiothorac Imaging · 2023-07-13

回顾性纳入59例肺气肿患者,分别接受双源或分束滤线器多能CT,以SPECT/CT为参考标准计算肺叶灌注分数。两种多能CT方法与SPECT/CT各肺叶均显著相关(P<.001),双源CT总体相关性(r=0.88)显著优于分束滤线器CT(r=0.78;P=.006)。

要点

  • 双源与分束滤线器多能CT均与SPECT/CT肺叶灌注分数显著相关(P<.001)。
  • 全肺叶合并分析,双源CT相关性r=0.88,优于分束滤线器CT的r=0.78。
  • 仅右上叶相关性差异有统计学意义(双源r=0.88 vs 分束滤线器r=0.58,P=.008)。

核心结论

  • 双源与分束滤线器多能CT在肺叶水平评估肺气肿患者肺灌注时与SPECT/CT相比均准确,双源多能CT总体相关性更高。

研究设计与样本

  • 回顾性纳入2016年5月至2021年2月间评估肺减容治疗的肺气肿患者59例(男30例、女29例;双源CT 31例、分束滤线器CT 28例;平均年龄67岁±8);以SPECT与胸部CT共注册(SPECT/CT)半手动分割计算肺叶灌注分数为参考标准,多能CT用U-Net自动分割肺叶并经人工核实,由碘图导出肺叶灌注分数;统计采用Pearson相关系数、组内相关系数和Bland-Altman分析。

诊断标准与关键数值

  • 双源多能CT与SPECT/CT全肺叶合并Pearson相关系数r=0.88。
  • 分束滤线器多能CT与SPECT/CT全肺叶合并Pearson相关系数r=0.78。
  • 右上叶Pearson相关系数:双源CT r=0.88 vs 分束滤线器CT r=0.58(P=.008)。
  • 两种多能CT方法与SPECT/CT在各单个肺叶均显著相关(P<.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate dual-source and split-beam filter multi-energy chest CT in assessing pulmonary perfusion on a lobar level in patients with lung emphysema, using perfusion SPECT as the reference standard.

MATERIALS AND METHODS: Patients with emphysema evaluated for lung volume reduction therapy between May 2016 and February 2021 were retrospectively included. All patients underwent SPECT and either dual-source or split-beam filter (SBF) multi-energy CT. To calculate the fractional lobar lung perfusion (FLLP), SPECT acquisitions were co-registered with chest CT scans (hereafter, SPECT/CT) and semi-manually segmented. For multi-energy CT scans, lung lobes were automatically segmented using a U-Net model. Segmentations were manually verified. The FLLP was derived from iodine maps computed from the multi-energy data. Statistical analysis included Pearson and intraclass correlation coefficients and Bland-Altman analysis.

RESULTS: Fifty-nine patients (30 male, 29 female; 31 underwent dual-source CT, 28 underwent SBF CT; mean age for all patients, 67 years ± 8 [SD]) were included. Both multi-energy methods significantly correlated with the SPECT/CT acquisitions for all individual lobes (P < .001). Pearson correlation concerning all lobes combined was significantly better for dual-source (r = 0.88) than for SBF multi-energy CT (r = 0.78; P = .006). On the level of single lobes, Pearson correlation coefficient differed for the right upper lobe only (dual-source CT, r = 0.88; SBF CT, r = 0.58; P = .008).

CONCLUSION: Dual-source and SBF multi-energy CT accurately assessed lung perfusion on a lobar level in patients with emphysema compared with SPECT/CT. The overall correlation was higher for dual-source multi-energy CT.Keywords: Chronic Obstructive Pulmonary Disease, Comparative Studies, Computer Applications, CT Spectral Imaging, Image Postprocessing, Lung, Pulmonary Perfusion© RSNA, 2023.

来源:Radiol Cardiothorac Imaging,查看原文。版权归原作者所有。

引用

  1. Strotzer QD, Heidemanns S, Mayr V, et al. Head-to-Head Comparison of Dual-Source and Split-Beam Filter Multi-Energy CT versus SPECT/CT for Assessing Lobar Lung Perfusion in Emphysema. Radiol Cardiothorac Imaging. 2023. PMID: 37693196. doi:10.1148/ryct.220273

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。