一种用于经颅多普勒检查中鉴别假定固体与气体微栓子信号的算法的诊断准确性
本研究评估了一种新算法(SGA-a)在经颅多普勒音频信号时域分析中鉴别假定固体与气体微栓子信号及伪差的诊断准确性。在心脏手术前1小时监测的伪差队列(20例患者,n=24,429)中,伪差分类准确率为94.5%;在主动脉瓣置换术后4小时监测的栓子验证队列(10例患者,n=12,328)中,区分固体/气体栓子与伪差的准确率为85.6%。固体测试文件误分类率为2%,气体测试文件误分类率为12%。
要点
- SGA-a算法在伪差队列中分类准确率达94.5%,在栓子验证队列中鉴别固体/气体栓子与伪差的准确率为85.6%。
- 固体测试文件误分类率为2%,气体测试文件误分类率为12%。
- 假定固体栓子数、气体栓子数及来源不明微栓子信号的95%一致性界限分别为[-10,10]、[-14,7]和[-9,16],组内相关系数均为0.99。
核心结论
- SGA-a能够检测假定的固体和气体微栓子信号并将其与伪差区分开;在固体和气体栓子均可能危及脑功能的情形(如心脏瓣膜和/或主动脉弓置换术中)可能有价值。
研究设计与样本
- 研究论文;采用两个队列验证SGA-a算法:队列1为心脏手术前接受1小时经颅多普勒检查的20例患者(伪差队列,n=24,429),由人类专家验证;队列2为主动脉瓣置换术后接受4小时监测的10例患者(栓子验证队列,n=12,328);另通过无伪差的固体和气体栓子测试文件评估误分类率。
诊断标准与关键数值
- 队列1(n=24,429)中,SGA-a对伪差的分类准确率为94.5%。
- 队列2(n=12,328)中,SGA-a区分固体/气体栓子与伪差的准确率为85.6%。
- 固体测试文件误分类率为2%,气体测试文件误分类率为12%。
- 假定固体栓子数、气体栓子数及来源不明微栓子信号的95%一致性界限分别为[-10,10]、[-14,7]和[-9,16],组内相关系数均为0.99。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The aim of the work described here was to assess the diagnostic accuracy of a new algorithm (SGA-a) for time-domain analysis of transcranial Doppler audio signals to discriminate presumed solid and gaseous microembolic signals and artifacts (SGAs).
METHODS: SGA-a was validated by human experts in an artifact cohort of 20 patients subjected to a 1-h transcranial Doppler exam before cardiac surgery (cohort 1). Emboli were validated in a cohort of 10 patients after aortic valve replacement in a 4-h monitoring period (cohort 2). The SGA misclassification rate was estimated by testing SGA-a on artifact-free test files of solid and gaseous emboli.
RESULTS: In cohort 1 (n = 24,429), artifacts were classified with an accuracy of 94.5%. In cohort 2 (n = 12,328), the accuracy in discriminating solid/gaseous emboli from artifacts was 85.6%. The 95% limits of agreement for, respectively, the numbers of presumed solids and gaseous emboli, artifacts and microembolic signals of undetermined origin were [-10, 10], [-14, 7] and [-9, 16], and the intra-class correction coefficients were 0.99, 0.99 and 0.99, respectively. The rate of misclassification of solid test files was 2%, and the rate of misclassification of gaseous test files was 12%.
CONCLUSION: SGA-a can detect presumed solid and gaseous microembolic signals and differentiate them from artifacts. SGA-a could be of value when both solid and gaseous emboli may jeopardize brain function such as seen during cardiac valve and/or aortic arch replacement procedures.
引用
- Keunen RWM, Daal SM, Romers GJ, et al. Diagnostic Accuracy of an Algorithm for Discriminating Presumed Solid and Gaseous Microembolic Signals During Transcranial Doppler Examinations. Ultrasound Med Biol. 2023. PMID: 37709563. doi:10.1016/j.ultrasmedbio.2023.08.011
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。