光子计数探测器CT联合去噪在低辐射剂量下骨盆骨成像的体模研究
该体模研究比较了光子计数探测器CT与能量积分探测器CT在有无去噪工具重建时骨盆骨图像质量,辐射剂量0.05–5.00 mGy。PCD CT噪声低于EID CT,去噪进一步降低噪声;读者认为PCD CT去噪重建在低于EID CT最低可接受剂量时图像质量更优。
要点
- 在0.05 mGy、100 kV加锡滤过时,PCD CT噪声38.4 HU,低于EID CT的48.8 HU。
- EID CT最低可接受剂量为0.10 mGy,两位读者总分均为15分中的11分。
- PCD CT去噪重建在0.10 mGy和0.05 mGy时图像质量相对优于参考重建。
核心结论
- PCD CT在EID CT最低可接受辐射剂量下图像质量优于EID CT;PCD CT联合去噪可在低于EID CT最低可接受剂量的水平获得更好图像质量。
研究设计与样本
- 体模研究,使用含人骨和模拟软组织的丙烯酸材料骨盆体模,行PCD CT与EID CT在不同管电压和0.05至5.00 mGy辐射剂量下的扫描,并生成商业去噪工具重建。
关键影像表现/征象
- 在0.05 mGy、100 kV加锡滤过时,PCD CT噪声为38.4 HU,EID CT为48.8 HU。
- 在100 kV加锡滤过、0.25 mGy时,PCD CT去噪前噪声19.9 HU,去噪后9.7 HU。
诊断标准与关键数值
- EID CT最低可接受剂量为0.10 mGy,两位读者总分均为15分中的11分。
- 两位读者均 somewhat 同意PCD CT未去噪在0.10 mGy时在骨结构、伪影和图像质量方面相对优于参考重建。
- 两位读者均 somewhat 同意PCD CT去噪重建在0.10 mGy和0.05 mGy时在图像质量指标上相对优于参考重建。
临床应用与阅片要点
- PCD CT联合去噪可能有助于降低骨盆成像的辐射剂量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Photon-counting detector (PCD) CT may allow lower radiation doses than used for conventional energy-integrating detector (EID) CT, with preserved image quality. OBJECTIVE. The purpose of this study was to compare PCD CT and EID CT, reconstructed with and without a denoising tool, in terms of image quality of the osseous pelvis in a phantom, with attention to low radiation doses. METHODS. A pelvic phantom comprising human bones in acrylic material mimicking soft tissue underwent PCD CT and EID CT at various tube potentials and radiation doses ranging from 0.05 to 5.00 mGy. Additional denoised reconstructions were generated using a commercial tool. Noise was measured in the acrylic material. Two readers performed independent qualitative assessments that entailed determining the denoised EID CT reconstruction with the lowest acceptable dose and then comparing this reference reconstruction with PCD CT reconstructions without and with denoising, using subjective Likert scales. RESULTS. Noise was lower for PCD CT than for EID CT. For instance, at 0.05 mGy and 100 kV with tin filter, noise was 38.4 HU for PCD CT versus 48.8 HU for EID CT. Denoising further reduced noise; for example, for PCD CT at 100 kV with tin filter at 0.25 mGy, noise was 19.9 HU without denoising versus 9.7 HU with denoising. For both readers, lowest acceptable dose for EID CT was 0.10 mGy (total score, 11 of 15 for both readers). Both readers somewhat agreed that PCD CT without denoising at 0.10 mGy (reflecting reference reconstruction dose) was relatively better than the reference reconstruction in terms of osseous structures, artifacts, and image quality. Both readers also somewhat agreed that denoised PCD CT reconstructions at 0.10 mGy and 0.05 mGy (reflecting matched and lower doses, respectively, with respect to reference reconstruction dose) were relatively better than the reference reconstruction for the image quality measures. CONCLUSION. PCD CT showed better-quality images than EID CT when performed at the lowest acceptable radiation dose for EID CT. PCD CT with denoising yielded better-quality images at a dose lower than lowest acceptable dose for EID CT. CLINICAL IMPACT. PCD CT with denoising could facilitate lower radiation doses for pelvic imaging.