全膝关节假体存在时下肢旋转对线:超低剂量光子计数CT扭转分析的可重复性

Rotational alignment of the lower extremity in the presence of total knee endoprosthesis: Reproducibility of torsion analyses using ultra-low-dose photon-counting CT.
PubMed 文献 · Eur J Radiol · 2023-08-19

研究采用第一代光子计数CT结合锡预滤过光谱整形,对8例尸体标本及3例单侧膝关节假体供体行髋-膝-踝三层扫描,比较三种剂量水平下股骨、胫骨及整体下肢扭转测量。结果显示各剂量方案间测量值无显著差异,观察者间信度优良(ICC 0.915–0.967),超低剂量下仍可可靠完成扭转分析。

要点

  • 超低剂量光子计数CT(髋0.25、膝0.06、踝0.06 mGy)与高质量方案间下肢扭转测量无显著差异。
  • 股骨、胫骨及整体下肢扭转的观察者间ICC分别为0.915、0.960、0.967,信度优良。
  • 膝关节假体存在时,调整管电压(膝水平Sn 140 kVp)可提高同侧及对侧诊断信心(p < 0.001)。

核心结论

  • 结合光子计数CT与光谱整形优势,即使存在全膝关节假体,超低辐射剂量下仍可行可靠的下肢扭转分析。

研究设计与样本

  • 采用第一代光子计数CT,8例尸体标本按既定三层扫描方案(髋Sn 140、膝Sn 100、踝Sn 100 kVp)检查,另对3例单侧膝关节假体供体追加膝水平Sn 140 kVp修改设置,三种剂量水平(高质量5.0-3.0-2.0 mGy、低剂量0.80-0.30-0.26 mGy、超低剂量0.25-0.06-0.06 mGy)由6名放射科医生行扭转分析并报告诊断信心。

关键影像表现/征象

  • 全膝关节假体标本中,绝对旋转测量值不受剂量水平和管电压影响。
  • 修改设置下同侧及对侧诊断信心更优(p < 0.001)。

诊断标准与关键数值

  • 股骨扭转观察者间ICC 0.915(95%CI 0.871-0.947)。
  • 胫骨扭转观察者间ICC 0.960(95%CI 0.938-0.976)。
  • 整体下肢扭转观察者间ICC 0.967(95%CI 0.945-0.981)。
  • 股骨(p = 0.330)、胫骨(p = 0.177)及整体下肢旋转测量(p = 0.358)在高质量、低剂量与超低剂量方案间无显著差异。

临床应用与阅片要点

  • 全膝关节置换术后有症状患者可行下肢扭转分析,超低剂量光子计数CT结合锡预滤过光谱整形可降低辐射暴露。
  • 膝水平采用Sn 140 kVp修改设置可提高假体存在时同侧及对侧的诊断信心。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Leg torsion analysis can provide valuable information in symptomatic patients after total knee arthroplasty. However, extensive beam-hardening and photon-starvation artifacts limit diagnostic assessability and dose reduction potential. For this study, we investigated the reproducibility of rotational measurements in ultra-low-dose photon-counting CT with spectral shaping via tin prefiltration.

MATERIAL AND METHODS: Employing a first-generation photon-counting CT, eight cadaveric specimens were examined with an established three-level scan protocol (hip: Sn 140, knee: Sn 100, ankle: Sn 100 kVp). In three body donors with unilateral knee endoprostheses, additional modified settings were applied (Sn 140 kVp at knee level). Protocols were executed with three dose levels (hip-knee-ankle, high-quality: 5.0-3.0-2.0 mGy, low-dose: 0.80-0.30-0.26 mGy, ultra-low-dose: 0.25-0.06-0.06 mGy). Six radiologists performed torsion analyses, additionally reporting their diagnostic confidence. Intraclass correlation coefficients (ICC) were calculated to assess interrater reliability.

RESULTS: No significant differences were ascertained for femoral (p = 0.330), tibial (p = 0.177), and overall leg rotation measurements (p = 0.358) among high-quality, low-dose, and ultra-low-dose protocols. Interrater reliability was excellent for torsion of the femur (ICC 0.915, 95% confidence interval 0.871-0.947), tibia (0.960, 0.938-0.976), and overall leg (0.967, 0.945-0.981). In specimens with total knee endoprostheses, absolute rotational measurements were unaffected by dose level and tube voltage despite superior diagnostic confidence on the ipsilateral and contralateral sides with modified settings (p < 0.001).

CONCLUSIONS: Combining the advantages of photon-counting CT and spectral shaping, reliable leg torsion analyses are feasible with ultra-low radiation exposure even in the presence of total knee endoprostheses.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Patzer TS, Kunz AS, Huflage H, et al. Rotational alignment of the lower extremity in the presence of total knee endoprosthesis: Reproducibility of torsion analyses using ultra-low-dose photon-counting CT. Eur J Radiol. 2023. PMID: 37632998. doi:10.1016/j.ejrad.2023.111055

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。