光子计数探测器CT与能量积分探测器CT在腰椎检查中的辐射剂量与图像质量比较
前瞻性纳入39例行研究性光子计数探测器(PCD)CT腰椎检查的患者,回顾性匹配39例行临床适应症能量积分探测器(EID)CT腰椎检查的患者,两组均使用锡预滤过平扫技术。PCD CT较EID CT的CTDIvol和SSDE显著降低,图像噪声和SNR无显著差异,CNR显著更高,主观图像质量无显著差异。
要点
- PCD CT腰椎平均CTDIvol为4.4 mGy,EID CT为11.1 mGy,PCD CT显著降低(p < .001)。
- PCD CT与EID CT的CTDIvol和SSDE均显著低于EID CT(p < .001),图像噪声和SNR无显著差异。
- PCD CT的CNR显著高于EID CT(33.6 vs 29.3,p < .001),主观图像质量评分两组中位数均为4分。
核心结论
- 与EID CT相比,PCD CT在保持图像质量的同时显著降低辐射剂量,支持扩大PCD CT在腰椎评估中的应用。
研究设计与样本
- 前瞻性研究性PCD CT样本39例(22男17女,平均年龄27.2岁)与回顾性临床EID CT样本39例(22男17女,平均年龄34.9岁),两组按年龄、性别和BMI匹配;所有检查均为2022年6月至2023年1月间使用锡预滤过平扫技术;使用自定义高斯曲线拟合算法自动计算图像噪声、SNR和CNR;三名放射科医生独立进行主观视觉评估。
关键影像表现/征象
- PCD CT平均CTDIvol为4.4 ± 1.0 mGy,EID CT为11.1 ± 1.9 mGy(p < .001)。
- PCD CT平均SSDE为6.2 ± 1.0 mGy,EID CT为14.2 ± 1.8 mGy(p < .001)。
- PCD CT与EID CT图像噪声和SNR无显著差异(均p > .05)。
- PCD CT的CNR显著高于EID CT(33.6 ± 3.3 vs 29.3 ± 4.1,p < .001)。
- 三名读者对PCD CT和EID CT的总体图像质量中位评分均为4分(范围3-4),任何读者的任何定性指标均无显著差异(均p > .05)。
临床应用与阅片要点
- PCD CT腰椎检查可在显著降低辐射剂量的同时保持与EID CT相当的图像噪声、SNR和主观图像质量。
- PCD CT的CNR显著高于EID CT,可能有助于腰椎结构的显示。
- 研究结果支持扩大PCD CT在腰椎评估中的应用。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Photon-counting detector (PCD) CT could be useful to help address the typically high radiation doses of conventional energy-integrating detector (EID) CT of the lumbar spine. OBJECTIVE. The purpose of our study was to compare PCD CT and EID CT of the lumbar spine, both performed using tin filtration, in terms of radiation dose and image quality. METHODS. This study included a prospective sample of 39 patients (22 men, 17 women; mean age, 27.2 years) who underwent investigational PCD CT of the lumbar spine as part of a separate study and a retrospective sample of 39 patients (22 men, 17 women; mean age, 34.9 years) who underwent clinically indicated EID CT of the lumbar spine. In both groups, all examinations were performed using unenhanced technique with tin prefiltration between June 2022 and January 2023. Patients were matched between groups using age, sex, and BMI. A custom gaussian curve-fitting algorithm was used to automatically calculate image noise, SNR, and CNR for each examination, on the basis of all voxels within the image set. Three radiologists independently reviewed examinations to perform a subjective visual assessment of visualization of trabecular architecture, cortical bone, neuroforaminal content, paraspinal muscles, and intervertebral disk, as well as overall image quality, using a 4-point Likert scale (1 = poor, 4 = excellent). PCD CT and EID CT examinations were compared. RESULTS. Mean CTDIvol was 4.4 ± 1.0 (SD) mGy for PCD CT versus 11.1 ± 1.9 mGy for EID CT (p < .001). Mean size-specific dose estimate (SSDE) was 6.2 ± 1.0 (SD) mGy for PCD CT versus 14.2 ± 1.8 mGy for EID CT (p < .001). PCD CT and EID CT examinations were not significantly different in terms of image noise or SNR (both p > .05). PCD CT, in comparison with EID CT, showed significantly higher CNR (mean ± SD, 33.6 ± 3.3 vs 29.3 ± 4.1; p < .001). For all three readers, the median score for overall image quality was 4 (range, 3-4) for both PCD CT and EID CT. PCD CT and EID CT examinations showed no significant difference in terms of any qualitative measure for any reader (all p > .05). CONCLUSION. PCD CT, in comparison with EID CT, yielded significantly lower radiation dose with preserved image quality. CLINICAL IMPACT. The findings support expanded use of PCD CT for lumbar spine evaluation.