文献速递乳腺MR

植入式循环记录器患者的乳腺MRI:初步经验

Breast MRI in patients with implantable loop recorder: initial experience.
PubMed 文献 · Eur Radiol · 2023-08-09

本回顾性研究分析了2008年4月至2022年9月期间植入式循环记录器(ILR)患者的乳腺MRI检查,共40/82,778例(0.049%)包含ILR。所有检查均完成且无不良事件;ILR主要位于左乳内下象限(64.6%),产生特征性伪影,包括偶极子样信号缺失及周边高信号,后者可模拟强化。8例患者MRI发现可疑病灶,其中4例为额外恶性病变;6例左乳癌中5例肿瘤完全可见、1例部分遮挡。结论认为乳腺MRI在ILR患者中可行且安全,但存在局部特征性伪影。

要点

  • ILR患者乳腺MRI检查可行且安全,本组无早期终止及不良事件。
  • ILR主要位于左乳内下象限(64.6%),产生偶极子样信号缺失及周边高信号伪影。
  • 周边高信号伪影在增强后及减影T1WI上可模拟可疑强化,需注意鉴别。

核心结论

  • 乳腺MRI在ILR患者中可行且安全,可能具有重要诊断价值,但存在局部特征性伪影。

研究设计与样本

  • 回顾性研究,2008年4月至2022年9月,共40例包含ILR的乳腺MRI(占同期82,778例的0.049%),复习影像报告和电子病历,并对不同脉冲序列上的伪影进行定量统计比较。

关键影像表现/征象

  • ILR主要位于左乳内下象限(64.6%);主要伪影为ILR床内偶极子形成的信号强度缺失,伴或不伴周边高信号区;伪影在头足轴最大(p < 0.001),其次为前后轴(p < 0.001),再者为左右轴;周边高信号伪影见于增强后及减影T1加权图像,模拟可疑强化;伪影在扩散加权成像上最显著(p < 0.001),其次为T2加权和T1加权图像;6例左乳癌中5例肿瘤完全可见、1例部分遮挡。

诊断标准与关键数值

  • 8例患者MRI发现可疑病灶,其中4例为额外恶性病变。
  • 6例左乳癌中,5例肿瘤完全可见,1例部分遮挡。

临床应用与阅片要点

  • 作者指出,有指征的乳腺MRI及引导活检在ILR患者中可安全进行;但放射科医生应意识到相关局限,包括左乳内侧评估受限和假强化伪影;作者认为有指征时应进行乳腺MRI,并考虑典型相关伪影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the feasibility of breast MRI exams and guided biopsies in patients with an implantable loop recorder (ILR) as well as the impact ILRs may have on image interpretation.

MATERIALS AND METHODS: This retrospective study examined breast MRIs of patients with ILR, from April 2008 to September 2022. Radiological reports and electronic medical records were reviewed for demographic characteristics, safety concerns, and imaging findings. MR images were analyzed and compared statistically for artifact quantification on the various pulse sequences.

RESULTS: Overall, 40/82,778 (0.049%) MRIs during the study period included ILR. All MRIs were completed without early termination. No patient-related or device-related adverse events occurred. ILRs were most commonly located in the left lower-inner quadrant (64.6%). The main artifact was a signal intensity (SI) void in a dipole formation in the ILR bed with or without areas of peripheral high SI. Artifacts appeared greatest in the cranio-caudal axis (p < 0.001), followed by the anterior-posterior axis (p < 0.001), and then the right-left axis. High peripheral rim-like SI artifacts appeared on the post-contrast and subtracted T1-weighted images, mimicking suspicious enhancement. Artifacts were most prominent on diffusion-weighted (p < 0.001), followed by T2-weighted and T1-weighted images. In eight patients, suspicious findings were found on MRI, resulting in four additional malignant lesions. Of six patients with left breast cancer, the tumor was completely visible in five cases and partially obscured in one.

CONCLUSION: Breast MRI is feasible and safe among patients with ILR and may provide a significant diagnostic value, albeit with localized, characteristic artifacts.

CLINICAL RELEVANCE STATEMENT: Indicated breast MRI exams and guided biopsies can be safely performed in patients with implantable loop recorder. Nevertheless, radiologists should be aware of associated limitations including limited assessment of the inner left breast and pseudo-enhancement artifacts.

KEY POINTS: • Breast MRI in patients with an implantable loop recorder is an infrequent, feasible, and safe procedure. • Despite limited breast visualization of the implantable loop recorder bed and characteristic artifacts, MRI depicted additional lesions in 8/40 (20%) of cases, half of which were malignant. • Breast MRI in patients with an implantable loop recorder should be performed when indicated, taking into consideration typical associated artifacts.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Nissan N, Ochoa-Albiztegui RE, Fruchtman H, et al. Breast MRI in patients with implantable loop recorder: initial experience. Eur Radiol. 2023. PMID: 37555957. doi:10.1007/s00330-023-10025-3