声触诊弹性成像在慢性肝病肝纤维化评估中的作用:以APRI为参考标准

The role of sound touch elastography in assessment of liver fibrosis in chronic liver disease keeping APRI as the reference standard.
PubMed 文献 · Abdom Radiol (NY) · 2023-08-24

前瞻性短期研究纳入60例慢性肝病患者,以APRI值为参考标准分为非显著纤维化、显著纤维化和肝硬化,用声触诊弹性成像(STE)测量肝硬度。STE值与APRI值呈正相关(rs=0.837,p<0.001),组内相关系数0.97,鉴别显著纤维化和肝硬化的最佳阈值分别为7.26 kPa和13.79 kPa,预测临床显著胃食管静脉曲张的阈值为18.84 kPa。

要点

  • STE测值随APRI分期升高而升高,与APRI值呈正相关(rs=0.837,p<0.001)。
  • STE鉴别显著纤维化阈值7.26 kPa(敏感度85.19%,特异度87.5%),肝硬化阈值13.79 kPa(敏感度96.0%,特异度88.89%)。
  • STE预测临床显著胃食管静脉曲张阈值18.84 kPa(敏感度92.85%,特异度95.65%)。

核心结论

  • 新型STE技术在以APRI值为参考标准时对肝纤维化分期和预测临床显著胃食管静脉曲张显示出良好准确性及极佳可靠性,前景可期,可广泛整合于临床实践用于慢性肝病纤维化评估和分期。

研究设计与样本

  • 前瞻性短期研究,纳入60例符合条件的慢性肝病患者,以APRI值为参考标准分为非显著纤维化、显著纤维化和肝硬化,每位患者按预定标准行STE检查获取多次读数,用Youden指数确定最佳阈值。

诊断标准与关键数值

  • 以APRI为参考标准,STE鉴别显著纤维化的最佳阈值为7.26 kPa(J=0.73,敏感度85.19%,特异度87.5%,95%CI),鉴别肝硬化的最佳阈值为13.79 kPa(J=0.84,敏感度96.0%,特异度88.89%,95%CI),对应AUROC分别为0.926(0.785-0.987)和0.976(0.890-0.999);预测临床显著胃食管静脉曲张的阈值为18.84 kPa(J=0.88,敏感度92.85%,特异度95.65%)。

临床应用与阅片要点

  • STE测值与APRI值呈正相关(rs=0.837,p<0.001),基于平均评分(k=5)的组内相关系数为0.97(0.95-0.99),提示测量间一致性极佳;作者认为该技术前景可期,可广泛整合于临床实践用于慢性肝病纤维化评估和分期。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: The study aims to determine the role of Sound Touch Elastography [STE] technique in staging liver fibrosis and predicting clinically significant gastro-esophageal varices among patients with chronic liver disease [CLD] keeping aspartate aminotransferase to platelet ratio index [APRI] as the reference standard.

METHODS: A prospective short-term study including 60 eligible patients with CLD were staged as non-significant fibrosis [NSF], significant fibrosis [SF] and cirrhosis [C] based on APRI values. STE was performed on each patient obtaining multiple readings as per pre-defined standards. The intra-observer reliability between each measurement and its association with APRI staging was evaluated using relevant statistical variables. Further, Youden's index was used to define the optimum cut-off values on STE in differentiating the stages of fibrosis and in predicting clinically significant gastro-esophageal varices.

RESULTS: Based on APRI cut-off values, 41.7% [n = 25] of the study population had cirrhosis, while 45% [n = 27] had significant fibrosis and 13.3% [n = 8] had NSF. The STE values in kPa showed a positive correlation with APRI values [(rs) = 0.837, p < 0.001]. The intra-class correlation estimates based on a mean rating [k = 5] was found to be 0.97 [0.95-0.99], implying an excellent agreement between the measurements. Optimum cut-off values in staging SF and C were 7.26 kPa [J = 0.73, sensitivity-85.19%, specificity-87.5%; 95% CI] and 13.79 kPa [J = 0.84, sensitivity-96.0%, specificity-88.89%; 95% CI]. The AUROC for each of these stages were 0.926 [0.785-0.987] and 0.976 [0.890-0.999], respectively. 23.3% [n = 14] of the study population had clinically significant gastro-esophageal varices with a value above 18.84 kPa [J = 0.88] showing a sensitivity of 92.85% and a specificity of 95.65% in predicting the same.

CONCLUSION: The novel STE technique shows good accuracy in staging liver fibrosis as determined by APRI values and in prediction of clinically significant gastro-esophageal varices with excellent reliability. It shows promising prospects and can be integrated widely in clinical practice for assessment and staging of fibrosis in CLD.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。