为肝脏组织适配的专用虚拟平扫图像在临床光子计数CT中改善了肝脏以外多个器官的虚拟平扫成像
回顾性纳入54例患者,从真平扫和门静脉期增强图像重建虚拟平扫(VNCn、VNCv),并评估肝脏专用VNC(VNCl)。各VNC在肝、肾皮质、主动脉和脂肪的CT值与真平扫差异显著,但肝、脾、肌肉平均偏移<10HU;VNCl在皮下脂肪偏移显著更低,提示其对脂肪组织测量更优。
要点
- 肝、脾、肌肉的VNC与真平扫平均CT值偏移均<10HU,临床可接受。
- VNCl在皮下脂肪的CT值偏移显著低于其他VNC,适合脂肪组织测量。
- 肝、肾皮质、主动脉和脂肪的VNC与真平扫CT值差异有统计学意义。
核心结论
- 尽管某些组织在真平扫与虚拟平扫图像上测得的HU值存在统计学显著差异,但肝和脾的微小偏移提示VNCv和VNCl图像具有良好的临床适用性;VNCl图像上皮下脂肪的偏移显著更低,提示其在脂肪组织测量方面更优。
研究设计与样本
- 回顾性纳入54例患者,从真平扫图像重建VNCn、从门静脉期增强图像重建VNCv,并评估肝脏专用VNC(VNCl),在肝、脾、肾皮质、主动脉、肌肉和皮下脂肪勾画感兴趣区进行比较。
诊断标准与关键数值
- 肝、脾、肌肉:真平扫与所有VNC图像的平均CT值偏移均<10HU。
- 皮下脂肪:真平扫与VNCl图像的平均CT值偏移<10HU。
- 真平扫与VNCv差异≤10HU者占68%,差异≤15HU者占79%;真平扫与VNCl差异≤10HU者占75%,差异≤15HU者占92%。
- 真平扫与VNCn差异≤10HU者占79%,差异≤15HU者占85%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Purpose of this study is to re-evaluate the accuracy and diagnostic reliability of virtual non-contrast (VNC) images acquired with the photon-counting computed tomography (PCCT) after an update of the CT scanner software.
METHODS: Fifty-four patients were retrospectively enrolled. VNC images were reconstructed from true non-contrast (TNC) images (VNCn) and contrast-enhanced images in portal venous contrast phase (VNCv). Additionally, a liver-specific VNC (VNCl) was assessed. Quantitative image properties of VNC and TNC images were compared and consistency between VNC images was evaluated. Regions of interest were drawn in the liver, spleen, renal cortex, aorta, muscle and subcutaneous fat.
RESULTS: Attenuation values on all VNC images differed significantly from TNC images in the liver, renal cortex, aorta and fat. A mean offset of <10HU between TNC and all VNC images was found in the liver, spleen and muscle. The comparison of TNC and VNCl images revealed an offset < 10HU in fat. Differences ≤ 10HU between TNC and VNCv and between TNC and VNCl were found in 68%, respectively in 75%. Differences ≤ 15HU were found in 79%, respectively in 92% of all measurements. Differences ≤ 10HU between TNC and VNCn were found in 79% and differences ≤ 15HU in 85%.
CONCLUSION: Although there are statistically significant differences between HU values measured on TNC and VNC images in certain tissues, the minor offsets measured in liver and spleen suggest a good clinical applicability of VNCv and VNCl images. The significantly lower offset in subcutaneous fat on VNCl images suggests a superiority for measurements in adipose tissues.