肝脏多采样超声点剪切波弹性成像:可重复性及与传统点剪切波弹性成像的一致性

Multisample Ultrasound Point Shear Wave Elastography of the Liver: Repeatability and Agreement With Conventional Point Shear Wave Elastography.
PubMed 文献 · J Ultrasound Med · 2023-08-24

研究在56名超重和肥胖成人中比较传统点剪切波弹性成像与单次屏气最多15次测量的多采样点剪切波弹性成像,两种技术间及基线与重复检查间呈中至高度相关、中度一致,多采样技术可在单次屏气内完成最多15次测量。

要点

  • 多采样点剪切波弹性成像可在单次屏气内完成最多15次肝脏硬度测量。
  • 多采样技术与传统点剪切波弹性成像的相关系数为r=0.59至0.73,一致性ICC为0.58至0.72。
  • 传统点剪切波弹性成像使用5C1与DAX探头在基线和重复检查时均呈高相关、中至良好一致。

核心结论

  • 允许单次屏气最多15次测量的多采样点剪切波弹性成像与传统点剪切波弹性成像呈中至高度相关和中度一致。

研究设计与样本

  • 在56名超重和肥胖成人中进行研究超声检查,使用传统点剪切波弹性成像(5C1和DAX探头,连续5次屏气各采集5次独立测量)和单次屏气多采样点剪切波弹性成像(DAX探头,最多15次独立测量),所有成像在基线及「咖啡休息」后重复进行。

诊断标准与关键数值

  • 传统点剪切波弹性成像5C1与DAX探头基线相关r=0.75(95%CI:0.61-0.85,P<.0001)、重复相关r=0.88(95%CI:0.78-0.92,P<.0001),基线ICC=0.69(95%CI:0.52-0.81)、重复ICC=0.81(95%CI:0.65-0.90)。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To characterize the technical performance of a multisample ultrasound point shear wave elastography (pSWE) technique that allows 15 individual measurements of liver stiffness in a single acquisition.

METHODS: In 56 overweight and obese adults, research ultrasound examinations were performed using conventional pSWE and multisample pSWE (Sequoia; Siemens Healthineers). Five independent measurements were acquired with the conventional technique over five consecutive breath holds (5C1 and DAX transducers). A single multisample acquisition (DAX transducer) of up to 15 unique measurements was acquired during a single breath hold. All imaging was performed before (baseline) and after a "coffee break" (repeat). Median liver stiffness measurements between techniques and between baseline and repeat imaging were compared using Pearson correlation (r) and intra-class correlation (ICC) coefficients.

RESULTS: Mean participant age was 33.7 ± 11.4 years; 40 participants were female. There was high correlation between conventional pSWE measurements obtained using the 5C1 and DAX transducers at baseline (r = .75 [95% CI: 0.61-0.85], P < .0001) and repeat (r = .88 [95% CI: 0.78-0.92], P < .0001). There was moderate agreement between conventional pSWE measurements obtained using the 5C1 and DAX transducers at baseline (ICC = 0.69 [95% CI: 0.52-0.81]), and good agreement at repeat (ICC = 0.81 [95% CI: 0.65-0.90]). There was moderate correlation (r = .59 [95% CI: 0.39-0.74], P < .0001) and moderate agreement (ICC = 0.58 [95% CI: 0.38-0.73]) between baseline conventional and multisample pSWE measurements acquired using the DAX transducer; there was high correlation (r = .73 [95% CI: 0.57-0.83], P < .0001) and moderate agreement (ICC = 0.72 [95% CI: 0.56-0.82] between techniques at repeat. There was moderate correlation (r = .65 [95% CI: 0.46-0.78], P < .0001) and moderate agreement (ICC = 0.64 [95% CI: 0.45-0.77]) between baseline and repeat multisample pSWE measurements.

CONCLUSIONS: Multisample pSWE, allowing up to 15 measurements in a single breath hold, showed moderate to high correlation and moderate agreement with conventional pSWE.

来源:J Ultrasound Med,查看原文。版权归原作者所有。