在爱尔兰建立放射摄影、乳腺X线摄影和双能X线吸收测量法服务的国家诊断参考水平并与欧洲诊断参考水平比较
该研究基于爱尔兰全国放射设施普查数据,建立放射摄影、乳腺X线摄影和双能X线吸收测量法(DXA)的国家诊断参考水平(DRL),并比较欧洲现有DRL。DRL取中位数分布的第75百分位,建立了12项成人放射摄影投照、11项儿科投照、单侧双位乳腺X线摄影及乳腺断层合成、4项DXA临床适应证的全国DRL,除一项外均低于或等于欧洲数据。
要点
- DRL取各设施中位剂量分布的第75百分位,用于识别患者剂量异常偏高并辅助优化。
- 按探测器类型差异显著时建立设备特异性DRL:计算机放射摄影与数字放射摄影各4项成人投照。
- 首次建立乳腺断层合成、DXA扫描和儿科放射摄影的国家DRL。
核心结论
- 该工作基于普查数据建立了覆盖放射摄影、乳腺X线摄影和DXA多种检查与临床任务的国家DRL,除一项外所有爱尔兰DRL数值均低于或等于欧洲数据。
研究设计与样本
- 通过在线调查普查爱尔兰所有提供放射摄影、乳腺X线摄影和DXA服务的放射设施,DRL设为各设施中位值分布的第75百分位,并建立国家中位剂量。
诊断标准与关键数值
- 国家DRL定义为各设施中位剂量分布的第75百分位;当不同探测器类型间DRL差异有统计学意义时建立设备特异性国家DRL。
临床应用与阅片要点
- 国家DRL是患者辐射剂量优化的重要组成部分,可用于识别患者剂量异常偏高的检查。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this work was to establish national diagnostic reference levels (DRLs) in Ireland and compare these to existing European DRLs where available. This work surveyed all radiological facilities providing radiography, mammography, and dual-energy x-ray absorptiometry (DXA) services in Ireland.
METHODS: A list of common procedures and clinical tasks was established. A national database of service providers was used to identify the appropriate medical radiological facilities providing these services. These facilities were issued with an online survey. National DRLs were set as the 75th percentile of the distribution of median values obtained. A national median dose was also established. The broad categorisation of equipment type was also considered. Where differences between DRLs established using different detector types were deemed statistically significant, equipment-specific national DRLs were established.
RESULTS: National DRLs were established for 12 adult radiography projections. Equipment-specific (computed radiography and digital radiography) adult DRLs were established for four radiography projections. Paediatric DRLs were established for 11 radiography projections, including two based on clinical indications, for a range of paediatric weight categories. National DRLs were established for unilateral two-view mammography and breast tomosynthesis as well as for four DXA clinical indications and projections. All but one Irish DRL figure was found to be below or equal to European data.
CONCLUSIONS: This work provided a unique opportunity to establish national DRLs based on census data for a range of procedures and clinical tasks across radiography, mammography and DXA and compare these with European levels.
CLINICAL RELEVANCE STATEMENT: This work established national diagnostic reference levels (DRLs) based on census data for a range of procedures and clinical tasks across radiography, mammography and dual-energy x-ray absorptiometry. The establishment of national DRLs is an essential component in the optimisation of patient radiation dose.
KEY POINTS: • Diagnostic reference levels are easily measured quantities intended for use as an aid to optimise patient dose and to identify when levels of patient dose are unusually high. • Data from all medical radiological facilities in Ireland was obtained to establish national diagnostic reference level (DRL) values and national median dose values in radiography, x-ray breast imaging and dual-energy x-ray absorptiometry (DXA) scanning and these were compared to existing European DRLs where available. • National DRL values were established for the first time in breast tomosynthesis, DXA scanning, and paediatric radiography.
引用
- O'Hora L, Neville N, Tuffy J, et al. Establishing national diagnostic reference levels in radiography, mammography, and dual-energy x-ray absorptiometry services in Ireland and comparing these with European diagnostic reference levels. Eur Radiol. 2023. PMID: 37505250. doi:10.1007/s00330-023-09992-4
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。