文献速递乳腺US

正则化背向散射定量超声表征乳腺肿块的操作者间与操作者内变异性

Inter- and Intra-Operator Variability of Regularized Backscatter Quantitative Ultrasound for the Characterization of Breast Masses.
PubMed 文献 · J Ultrasound Med · 2023-07-25

研究对29个BI-RADS 4和5类乳腺病灶采用两种成像协议(协议1径向与反径向切面,协议2短轴与长轴切面)采集射频回波信号,评估基于正则化的低方差定量超声背向散射系数的操作者内与操作者间变异性。协议1操作者间变异显著(P<.007),ICCinter=.77(95%CI .71-.82);协议2操作者间变异不显著(P>.05),ICCinter=.92(.89-.94),两协议中操作者内变异均不显著。

要点

  • 协议2(短轴与长轴切面)操作者间一致性优秀,ICCinter=.92(.89-.94),操作者间变异不显著(P>.05)。
  • 协议1(径向与反径向切面)操作者间变异显著(P<.007),ICCinter=.77(95%CI .71-.82),一致性良好。
  • 两种协议下操作者内变异均不显著,提示标准化采集协议可降低操作者间变异。

核心结论

  • 研究提示基于背向散射的定量超声需要标准化图像采集协议以减少操作者间变异,从而确保其成功转化用于可疑乳腺肿块的表征。

研究设计与样本

  • 对29个BI-RADS 4和5类乳腺病灶在2个序贯队列中按2种成像协议采集射频回波信号:队列1为径向与反径向切面,队列2为短轴与长轴切面;每例患者由3名放射科医师中至少2名扫描,每名医师在两次采集之间重新定位探头和患者,共进行3次采集;采用基于正则化的低方差定量超声方法估计背向散射系数,以多因素方差分析(P值)和组内相关系数(ICC)评估病灶内中位背向散射系数的操作者内与操作者间变异性。

诊断标准与关键数值

  • 协议1操作者间变异显著(P<.007),ICCinter=.77(95%CI .71-.82),提示操作者间一致性良好。
  • 协议2操作者间变异不显著(P>.05),ICCinter=.92(95%CI .89-.94),提示操作者间一致性优秀。
  • 两种协议中操作者内变异均不显著。

临床应用与阅片要点

  • 作者结论指出,基于背向散射的定量超声需要标准化图像采集协议以减少操作者间变异,并确保其成功转化用于可疑乳腺肿块的表征。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Here we report on the intra- and inter-operator variability of the backscatter coefficient (BSC) estimated with a new low-variance quantitative ultrasound (QUS) approach applied to breast lesions in vivo.

METHODS: Radiofrequency (RF) echo signals were acquired from 29 BIRADS 4 and 5 breast lesions in 2 sequential cohorts following 2 imaging protocols: cohort 1) radial and antiradial views, and cohort 2) short- and long-axis views. Protocol 2 was implemented after retraining and discussion on how to improve reproducibility. Each patient was scanned by at least 2 of 3 radiologists; each performed 3 acquisitions with transducer and patient repositioning in between acquisitions. BSC was estimated using a low-variance QUS approach based on regularization. Intra- and inter-operator variability of the intra-lesion median BSC was evaluated with a multifactorial ANOVA test (P-values) and the intraclass correlation coefficient (ICC).

RESULTS: Inter-operator variability was only significant in the first protocol (P < .007); ICCinter = .77 (95% CI .71-.82), indicating good inter-operator agreement. In the second protocol, the inter-operator variability was not significant (P > .05) and agreement was excellent (ICCinter = .92 [.89-.94]). In both protocols, the intra-operator variability was not significant.

CONCLUSIONS: Our findings demonstrate the need for standardizing image acquisition protocols for backscatter-based QUS to reduce inter-operator variability and ensure its successful translation to the characterization of suspicious breast masses.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Argueta-Lozano AK, Castañeda-Martinez L, Bass V, et al. Inter- and Intra-Operator Variability of Regularized Backscatter Quantitative Ultrasound for the Characterization of Breast Masses. J Ultrasound Med. 2023. PMID: 37490582. doi:10.1002/jum.16292

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。