正确封堵左心耳可减少淤滞血流并降低血栓形成风险:基于4D血流磁共振成像的概念验证实验研究

Correct Closure of the Left Atrial Appendage Reduces Stagnant Blood Flow and the Risk of Thrombus Formation: A Proof-of-Concept Experimental Study Using 4D Flow Magnetic Resonance Imaging.
PubMed 文献 · Korean J Radiol · 2023-01-01

本研究基于一位86岁长程持续性房颤患者的CT图像构建了封堵前、正确封堵和错误封堵三种左心房3D打印模型,用4D血流MRI比较血流淤滞和血栓形成指标。结果显示正确封堵模型的淤滞体积、壁面剪切应力和内皮细胞活化潜能均最低,提示正确封堵左心耳可能最大程度减少左房血流淤滞和血栓形成倾向。

要点

  • 正确封堵左心耳模型的左房血流淤滞体积较封堵前和错误封堵模型更低。
  • 正确封堵模型的壁面剪切应力和内皮细胞活化潜能均最低,提示血栓形成倾向更低。
  • 4D血流MRI可在3D打印左房模型中直观显示左房血流空间分布和方向。

核心结论

  • 正确封堵左心耳可最大程度减少左房血流淤滞和血栓形成倾向,为房颤患者提供了暂定的操作目标。

研究设计与样本

  • 基于一位86岁长程持续性房颤男性患者的心脏CT图像构建封堵前、正确封堵和错误封堵三种全尺寸左心房3D打印模型,连接闭环搏动性肺静脉血流回路,用3T磁共振行4D血流成像并用MATLAB软件分析。

关键影像表现/征象

  • 4D血流MRI可直接显示三种左心房模型内左房血流的不同空间分布、方向和幅度。
  • 正确封堵模型中左房血流淤滞的时间平均体积及其与相应左房总体积的比值均降低。

诊断标准与关键数值

  • 血流淤滞定义为速度阈值|V̅| < 3 cm/s。
  • 正确封堵模型的时间平均淤滞体积为70.82 mL,淤滞体积与相应左房总体积之比为39.0%。
  • 错误封堵模型的时间平均淤滞体积为73.17 mL,淤滞体积与相应左房总体积之比为39.0%。
  • 封堵前模型的时间平均淤滞体积为79.11 mL,淤滞体积与相应左房总体积之比为39.7%。
  • 正确封堵模型的表面和时间平均壁面剪切应力为0.048 Pa、内皮细胞活化潜能为4.004 Pa-1,错误封堵模型分别为0.059 Pa和4.792 Pa-1,封堵前模型分别为0.072 Pa和5.861 Pa-1。

临床应用与阅片要点

  • 4D血流MRI结合3D打印模型可用于评估左心耳封堵前后左房血流动力学改变。
  • 正确封堵左心耳可能最大程度减少左房血流淤滞和血栓形成倾向,为房颤患者提供暂定的操作目标。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The study was conducted to investigate the effect of correct occlusion of the left atrial appendage (LAA) on intracardiac blood flow and thrombus formation in patients with atrial fibrillation (AF) using four-dimensional (4D) flow magnetic resonance imaging (MRI) and three-dimensional (3D)-printed phantoms.

MATERIALS AND METHODS: Three life-sized 3D-printed left atrium (LA) phantoms, including a pre-occlusion (i.e., before the occlusion procedure) model and correctly and incorrectly occluded post-procedural models, were constructed based on cardiac computed tomography images from an 86-year-old male with long-standing persistent AF. A custom-made closed-loop flow circuit was set up, and pulsatile simulated pulmonary venous flow was delivered by a pump. 4D flow MRI was performed using a 3T scanner, and the images were analyzed using MATLAB-based software (R2020b; Mathworks). Flow metrics associated with blood stasis and thrombogenicity, such as the volume of stasis defined by the velocity threshold (|V̅| < 3 cm/s), surface-and-time-averaged wall shear stress (WSS), and endothelial cell activation potential (ECAP), were analyzed and compared among the three LA phantom models.

RESULTS: Different spatial distributions, orientations, and magnitudes of LA flow were directly visualized within the three LA phantoms using 4D flow MRI. The time-averaged volume and its ratio to the corresponding entire volume of LA flow stasis were consistently reduced in the correctly occluded model (70.82 mL and 39.0%, respectively), followed by the incorrectly occluded (73.17 mL and 39.0%, respectively) and pre-occlusion (79.11 mL and 39.7%, respectively) models. The surface-and-time-averaged WSS and ECAP were also lowest in the correctly occluded model (0.048 Pa and 4.004 Pa-1 , respectively), followed by the incorrectly occluded (0.059 Pa and 4.792 Pa-1 , respectively) and pre-occlusion (0.072 Pa and 5.861 Pa-1 , respectively) models.

CONCLUSION: These findings suggest that a correctly occluded LAA leads to the greatest reduction in LA flow stasis and thrombogenicity, presenting a tentative procedural goal to maximize clinical benefits in patients with AF.