文献速递胸部CT

辐射剂量对低剂量胸部CT间质性肺改变检出及图像质量的影响:基于单次患者扫描多剂量水平的评估

Impact of radiation dose on the detection of interstitial lung changes and image quality in low-dose chest CT - Assessment in multiple dose levels from a single patient scan.
PubMed 文献 · Eur J Radiol · 2023-07-17

回顾性纳入228例疑似间质性肺病患者,单次双源100 kVp采集后经线性混合重建出100%、67%、45%、39%、33%五档剂量,两名盲法阅片者评估网状影、磨玻璃影和蜂窝影。蜂窝影百分比一致率最高(>96%)且不随剂量降低变化,网状影和磨玻璃影一致率在100%降至67%时下降,后续进一步降低剂量未见显著变化;主观图像噪声随剂量降低而增加。

要点

  • 蜂窝影检出一致率>96%且不随剂量降低变化,网状影和磨玻璃影在100%降至67%时一致率下降。
  • 100%剂量为2.34 ± 0.97 mGy,主观图像噪声随剂量降低而增加(ρ=0.34和0.53,P<0.001)。
  • 剂量降低对细微间质改变影响更大,提示维持诊断准确性需要最低剂量。

核心结论

  • 剂量降低对细微间质性肺改变的影响更大;检出能力在初始剂量降低时即下降,提示疑似间质性肺病胸部CT需要最低剂量以维持诊断准确性。

研究设计与样本

  • 回顾性研究,连续纳入228例疑似间质性肺病成人患者,行单次双源100 kVp平扫胸部CT(Dual-split模式),管A和管B分别分配总管电流时间积的67%和33%。

关键影像表现/征象

  • 蜂窝影百分比一致率最高(>96%)且不随剂量降低变化(P>0.8)。
  • 网状影和磨玻璃影百分比一致率在剂量从100%降至67%时下降:阅片者1网状影83.3%、磨玻璃影87.7%,阅片者2网状影93.9%、磨玻璃影79.8%。
  • 进一步降低剂量未显著改变两名阅片者的百分比一致率(均P>0.05)。
  • 主观图像噪声随辐射剂量降低而增加(阅片者1 ρ=0.34,阅片者2 ρ=0.53,P<0.001)。

诊断标准与关键数值

  • 以100%剂量为参考,计算各降低剂量水平下网状影、磨玻璃影和蜂窝影的百分比一致率。
  • 100%剂量为2.34 ± 0.97 mGy;五档剂量为100%、67%、45%、39%、33%。

临床应用与阅片要点

  • 阅片时应注意剂量降低对网状影和磨玻璃影检出的影响大于蜂窝影。
  • 剂量降低后主观图像噪声增加,可能影响细微间质改变的判读。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To assess image quality and detectability of interstitial lung changes using multiple radiation doses from the same chest CT scan of patients with suspected interstitial lung disease (ILD).

METHOD: Retrospective study of consecutive adult patients with suspected ILD receiving unenhanced chest CT as single-energy dual-source acquisition at 100 kVp (Dual-split mode). 67% and 33% of the overall tube current time product were assigned to tube A and B, respectively. 100%-dose was 2.34 ± 0.97 mGy. Five different radiation doses (100%, 67%, 45%, 39%, 33%) were reconstructed from this single acquisition using linear-blending technique. Two blinded radiologists assessed reticulations, ground-glass opacities (GGO) and honeycombing as well as subjective image noise. Percentage agreement (PA) as compared to 100%-dose were calculated. Non-parametric statistical tests were used.

RESULTS: A total of 228 patients were included (61.2 ± 14.6 years,146 female). PA was highest for honeycombing (>96%) and independent of dose reduction (P > 0.8). PA for reticulations and GGO decreased when reducing the radiation dose from 100% to 67% for both readers (reticulations: 83.3% and 93.9%; GGO: 87.7% and 79.8% for reader 1 and 2, respectively). Additional dose reduction did not significantly change PA for both readers (all P > 0.05). Subjective image noise increased with decreasing radiation dose (Spearman Rho of ρ = 0.34 and ρ = 0.53 for reader 1 and 2, respectively, P < 0.001).

CONCLUSIONS: Radiation dose reduction had a stronger impact on subtle interstitial lung changes. Detectability decreased with initial dose reduction indicating that a minimum dose is needed to maintain diagnostic accuracy in chest CT for suspected ILD.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Martini K, Jungblut L, Sartoretti T, et al. Impact of radiation dose on the detection of interstitial lung changes and image quality in low-dose chest CT - Assessment in multiple dose levels from a single patient scan. Eur J Radiol. 2023. PMID: 37478655. doi:10.1016/j.ejrad.2023.110981

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。