现代低场强MRI(0.55 T)与CT对照检测肺结节
前瞻性单中心研究纳入46例已知肺结节患者,同日行0.55 T低场强MRI与CT检查。MRI对≥6 mm肺结节检出准确率100%(126/126),≥4-<6 mm为80%(159/200),<4 mm为23%(147/638);MRI与CT结节大小测量Spearman相关系数r=0.87(P<0.001),平均差值0.16±0.9 mm。
要点
- MRI对≥6 mm肺结节检出准确率100%,对<4 mm结节检出准确率仅23%。
- MRI与CT结节大小测量高度相关(r=0.87),平均差值0.16±0.9 mm。
- 低场强MRI对≥6 mm肺结节检出准确率优异,但无法与CT竞争小结节检出。
核心结论
- 现代低场强MRI对≥6 mm肺结节检出准确率优异,与CT大小测量相关性很强,但对小结节检出无法与CT相比。
研究设计与样本
- 前瞻性临床试验,单中心学术医疗中心,纳入46例已知肺结节患者(平均年龄66±11岁,26例女性),2020年11月至2021年7月同日行0.55 T低场强MRI与CT检查,以CT为参考标准,采用横断面周期性旋转重叠平行线增强重建联合冠状面半傅里叶采集单次激发涡轮自旋回波序列。
诊断标准与关键数值
- MRI对≥6 mm肺结节检出准确率100%(126/126),对≥4-<6 mm结节准确率80%(159/200),对<4 mm结节准确率23%(147/638);MRI与CT大小测量Spearman相关系数r=0.87(P<0.001),平均差值0.16±0.9 mm。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to evaluate the accuracy of modern low-field magnetic resonance imaging (MRI) for lung nodule detection and to correlate nodule size measurement with computed tomography (CT) as reference.
MATERIALS AND METHODS: Between November 2020 and July 2021, a prospective clinical trial using low-field MRI at 0.55 T was performed in patients with known pulmonary nodules from a single academic medical center. Every patient underwent MRI and CT imaging on the same day. The primary aim was to evaluate the detection accuracy of pulmonary nodules using MRI with transversal periodically rotated overlapping parallel lines with enhanced reconstruction in combination with coronal half-Fourier acquired single-shot turbo spin-echo MRI sequences. The secondary outcome was the correlation of the mean lung nodule diameter with CT as reference according to the Lung Imaging Reporting and Data System. Nonparametric Mann-Whitney U test, Spearman rank correlation coefficient, and Bland-Altman analysis were applied to analyze the results.
RESULTS: A total of 46 participants (mean age ± SD, 66 ± 11 years; 26 women) were included. In a blinded analysis of 964 lung nodules, the detection accuracy was 100% for those ≥6 mm (126/126), 80% (159/200) for those ≥4-<6 mm, and 23% (147/638) for those <4 mm in MRI compared with reference CT. Spearman correlation coefficient of MRI and CT size measurement was r = 0.87 ( P < 0.001), and the mean difference was 0.16 ± 0.9 mm.
CONCLUSIONS: Modern low-field MRI shows excellent accuracy in lesion detection for lung nodules ≥6 mm and a very strong correlation with CT imaging for size measurement, but could not compete with CT in the detection of small nodules.
引用
- Hinsen M, Nagel AM, May MS, et al. Lung Nodule Detection With Modern Low-Field MRI (0.55 T) in Comparison to CT. Invest Radiol. 2024. PMID: 37490031. doi:10.1097/RLI.0000000000001006