通过调整检查床高度使头部偏离中心以降低脑三维旋转血管造影中眼晶状体辐射剂量的前瞻性研究
本研究采用RANDO头部体模及20例颅内动脉瘤患者的前瞻性自身对照设计,评估三维旋转血管造影中通过抬高检查床使头部偏离中心对眼晶状体剂量的影响。体模研究中检查床每升高1 cm,晶状体剂量平均降低38%;患者研究中剂量降低方案(检查床平均抬高2.3 cm)使中位剂量从4.65 mGy降至0.79 mGy,降低83%(P < 0.001),而面积乘积剂量、空气比释动能及图像质量均无显著差异。
要点
- 三维旋转血管造影中检查床高度调整可显著影响眼晶状体辐射剂量。
- 患者研究中检查床平均抬高2.3 cm使中位晶状体剂量从4.65 mGy降至0.79 mGy,降低83%。
- 剂量降低方案与常规方案在面积乘积剂量、空气比释动能及图像质量上无显著差异。
核心结论
- 三维旋转血管造影中眼晶状体辐射剂量受检查床高度调整的显著影响;通过抬高检查床有意使头部偏离中心是临床实践中降低晶状体剂量的简单有效方法。
研究设计与样本
- 前瞻性研究,采用RANDO头部体模(Alderson Research Labs)进行体模实验,并前瞻性纳入20例(58.0 ± 9.4岁)拟行双侧三维旋转血管造影的颅内动脉瘤患者;患者双侧三维旋转血管造影中,一侧颈内动脉应用抬高检查床的晶状体剂量降低方案,另一侧应用常规方案;使用光致发光玻璃剂量计(GD-352M,AGC Techno Glass Co., LTD)测量晶状体剂量。
诊断标准与关键数值
- 体模研究中检查床每升高1 cm,晶状体剂量平均降低38%。
- 患者研究中剂量降低方案(检查床平均抬高2.3 cm)使中位晶状体剂量从4.65 mGy降至0.79 mGy,降低83%(P < 0.001)。
- 剂量降低方案与常规方案的面积乘积剂量为7.34 vs. 7.40 Gy·cm²(P = 0.892),无显著差异。
- 剂量降低方案与常规方案的空气比释动能75.7 vs. 75.1 mGy(P = 0.872),无显著差异。
- 源图像图像噪声、信噪比、对比噪声比及三位评审者采用五点Likert量表的定性图像质量评估在两种方案间无显著差异。
临床应用与阅片要点
- 作者认为通过抬高检查床有意使头部偏离中心是临床实践中降低三维旋转血管造影眼晶状体剂量的简单有效方法;患者研究中检查床平均抬高2.3 cm。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Three-dimensional rotational angiography (3D-RA) is increasingly used for the evaluation of intracranial aneurysms (IAs); however, radiation exposure to the lens is a concern. We investigated the effect of head off-centering by adjusting table height on the lens dose during 3D-RA and its feasibility in patient examination.
MATERIALS AND METHODS: The effect of head off-centering during 3D-RA on the lens radiation dose at various table heights was investigated using a RANDO head phantom (Alderson Research Labs). We prospectively enrolled 20 patients (58.0 ± 9.4 years) with IAs who were scheduled to undergo bilateral 3D-RA. In all patients' 3D-RA, the lens dose-reduction protocol involving elevation of the examination table was applied to one internal carotid artery, and the conventional protocol was applied to the other. The lens dose was measured using photoluminescent glass dosimeters (GD-352M, AGC Techno Glass Co., LTD), and radiation dose metrics were compared between the two protocols. Image quality was quantitatively analyzed using source images for image noise, signal-to-noise ratio, and contrast-to-noise ratio. Additionally, three reviewers qualitatively assessed the image quality using a five-point Likert scale.
RESULTS: The phantom study showed that the lens dose was reduced by an average of 38% per 1 cm increase in table height. In the patient study, the dose-reduction protocol (elevating the table height by an average of 2.3 cm) led to an 83% reduction in the median dose from 4.65 mGy to 0.79 mGy (P < 0.001). There were no significant differences between dose-reduction and conventional protocols in the kerma area product (7.34 vs. 7.40 Gy·cm², P = 0.892), air kerma (75.7 vs. 75.1 mGy, P = 0.872), and image quality.
CONCLUSION: The lens radiation dose was significantly affected by table height adjustment during 3D-RA. Intentional head off-centering by elevation of the table is a simple and effective way to reduce the lens dose in clinical practice.
引用
- Ryu JC, Yoon JT, Kim BJ, et al. Reduction of Radiation Dose to Eye Lens in Cerebral 3D Rotational Angiography Using Head Off-Centering by Table Height Adjustment: A Prospective Study. Korean J Radiol. 2023. PMID: 37404110. doi:10.3348/kjr.2023.0169
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。