文献速递骨肌技术与安全US

新型脊柱装置超声触发释放的体外与体内评价

In Vitro and In Vivo Evaluation of Ultrasound-Triggered Release From Novel Spinal Device.
PubMed 文献 · J Ultrasound Med · 2023-05-30

研究将载有万古霉素和Sonazoid微泡的聚乳酸夹在体外经多普勒超声触发释放,72小时累积释放达101.8±2.8%(81.4±2.8 mg),未辐照夹仅0.3±0.1%(P<.0001);兔体内夹未破裂,猪体内膜可见破裂并见亚甲蓝组织染色,结论为超声触发释放提供概念验证。

要点

  • 体外多普勒超声辐照载万古霉素聚乳酸夹72小时累积释放101.8±2.8%(81.4±2.8 mg),未辐照仅0.3±0.1%(P<.0001)。
  • 兔椎旁植入夹辐照后未破裂且无亚甲蓝组织染色,猪椎旁植入夹辐照后聚乳酸膜可见破裂并见亚甲蓝组织染色。
  • Sonazoid微泡在万古霉素中较蒸馏水中更易保存(4°C、25°C、37°C,P=.0131)。

核心结论

  • 结果证明封装预防性溶液可被超声触发释放,并为继续开展大动物评价以推动转化价值提供重要概念验证。

研究设计与样本

  • 体外研究将载万古霉素和Sonazoid微泡的聚乳酸夹置水中用Logiq E10超声仪C6凸阵探头多普勒模式辐照并以分光光度法定量释放;体内将载亚甲蓝和Sonazoid的夹植入兔L2、L5及猪L1、L3椎旁后辐照。

关键影像表现/征象

  • 兔回收夹辐照后未破裂,周围组织无亚甲蓝染色。
  • 猪辐照后聚乳酸膜肉眼可见破裂并见亚甲蓝组织染色,未辐照夹保持完整。
  • 体外万古霉素溶液中Sonazoid微泡在4°C、25°C、37°C孵育时较蒸馏水保存更好(P=.0131)。
  • 4°C储存条件下两种溶液均观察到造影增强。

诊断标准与关键数值

  • 体外辐照夹72小时平均累积万古霉素释放101.8±2.8%(81.4±2.8 mg),未辐照夹仅0.3±0.1%(0.3±0.1 mg),P<.0001。

临床应用与阅片要点

  • 作者认为结果提供超声触发释放封装预防性溶液的概念验证,支持继续开展大动物评价以评估转化价值。

局限性

  • 原文结论指出需继续开展大动物评价以评估转化价值。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Bacterial infection following spinal fusion is a major clinical concern with up to 20% incidence. An ultrasound-triggered bulk-release system to combat postsurgical bacterial survival was designed and evaluated.

METHODS: Polylactic acid (PLA) clips were loaded with vancomycin (VAN) and microbubbles (Sonazoid, GE HealthCare) in vitro. Stability was determined over 14 days. VAN-loaded clips were submerged in water and insonated using a Logiq E10 scanner (GE HealthCare) with a curvilinear C6 probe. Doppler-induced VAN release was quantified using spectrophotometry. For in vivo testing, clips were loaded with methylene blue (MeB) solution and Sonazoid. These clips were implanted into a rabbit along the spine at L2 and L5, as well as a pig at L1 and L3, then insonated in Doppler mode using the C6 probe.

RESULTS: Sonazoid microbubbles were better preserved when incubated in VAN compared with distilled water at 4°C, 25°C, and 37°C incubation temperatures (P = .0131). Contrast enhancement was observed from both solutions when incubated at 4°C storage conditions. Insonated clips achieved average cumulative VAN release of 101.8 ± 2.8% (81.4 ± 2.8 mg) after 72 hours. Uninsonated clips had only 0.3 ± 0.1% (0.3 ± 0.1 mg) average cumulative VAN release (P < .0001). Clips retrieved from the rabbit did not rupture with insonation nor produce MeB staining of surrounding tissues. In the pig, the PLA film was visibly ruptured and MeB tissue was observed following insonation, whereas the uninsonated clip was intact.

CONCLUSION: These results demonstrate ultrasound-triggered release of an encapsulated prophylactic solution and provide an important proof-of-concept for continuing large animal evaluations for translational merit.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Delaney LJ, Isguven S, Hilliard R, et al. In Vitro and In Vivo Evaluation of Ultrasound-Triggered Release From Novel Spinal Device. J Ultrasound Med. 2023. PMID: 37249416. doi:10.1002/jum.16263

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。