相位校正信号平均:两种监护仪间的相关性及其与胎心率短期变异的关联

Phase-rectified signal averaging: correlation between two monitors and relationship with short-term variation of fetal heart rate.
PubMed 文献 · Ultrasound Obstet Gynecol · 2023-01-01

前瞻性观察研究纳入285名孕28周以上孕妇,共采集306对胎心监护(cCTG)与新型自贴式无创胎心电(NIFECG)同步记录,评估相位校正信号平均(PRSA)输出在两种设备间的相关性及其与短期变异(STV)的关系。完全滤波下cAAC与eAAC(r=0.879)、cADC与eADC(r=0.895)强相关;cAAC、cADC与STV呈极强线性相关(r=0.928、0.911)。作者认为PRSA似乎是等效于STV的胎儿评估方法,且能更高比例获得可解读的NIFECG曲线。

要点

  • 完全滤波PRSA的AAC和ADC在cCTG与NIFECG间强相关(r=0.879和r=0.895,P<0.001)。
  • 锚点检测数<100的NIFECG曲线PRSA指数与cCTG偏离较大。
  • PRSA可能等效于STV用于胎儿评估,且可提高可解读NIFECG曲线比例。

核心结论

  • PRSA似乎是等效于STV的胎儿评估方法,且因能消除伪差,能以更高比例生成可解读的NIFECG曲线;这些发现提出自我居家或远程胎心率监测配合自动报告的可能性。

研究设计与样本

  • 前瞻性观察研究,纳入285名孕28+0周以上单胎孕妇,于伦敦教学医院进行cCTG评估,同期进行cCTG与NIFECG监测最长60分钟,共采集306对同步记录。

关键影像表现/征象

  • 完全滤波NIFECG PRSA(eAAC/eADC)结果由65%的曲线生成,cCTG PRSA(cAAC/cADC)输出由全部曲线生成。
  • NIFECG锚点检测随信号丢失增加而显著减少,当曲线检出锚点少于100个时,NIFECG PRSA指数与cCTG偏离较大。
  • 去除NIFECG锚点滤波生成部分滤波PRSA输出后,相关性减弱:AAC r=0.505(P<0.001),ADC r=0.560(P<0.001)。
  • 将锚点阈值降至100后,eAAC和eADC输出产率增至约74%,同时与cAAC(r=0.839,P<0.001)和cADC(r=0.815,P<0.001)保持强相关。
  • cAAC与cCTG STV呈极强线性关系(r=0.928,P<0.001),cADC与STV亦呈极强线性关系(r=0.911,P<0.001)。

诊断标准与关键数值

  • cCTG与NIFECG的AAC强相关:r=0.879(P<0.001)。
  • cCTG与NIFECG的ADC强相关:r=0.895(P<0.001)。
  • cAAC与cCTG STV极强相关:r=0.928(P<0.001);cADC与STV极强相关:r=0.911(P<0.001)。
  • eAAC与STV相关:r=0.825(P<0.001);eADC与STV相关:r=0.827(P<0.001)。
  • 锚点阈值降至100时,eAAC与cAAC相关r=0.839(P<0.001),eADC与cADC相关r=0.815(P<0.001)。

临床应用与阅片要点

  • PRSA似乎是等效于STV的胎儿评估方法,但原文未据此给出具体阅片流程推荐。
  • 作者认为这些发现提出自我居家或远程胎心率监测配合自动报告的可能性。
  • 原文结论未给出关于锚点阈值选择的推荐。

局限性

  • 完全滤波NIFECG PRSA结果仅由65%的曲线生成,cCTG PRSA输出由全部曲线生成。
  • NIFECG锚点检测随信号丢失增加而显著减少,锚点少于100个时PRSA指数与cCTG偏离较大。
  • 去除锚点滤波生成部分滤波PRSA输出后相关性减弱(AAC r=0.505,ADC r=0.560,均P<0.001)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To establish the correlation between phase-rectified signal averaging (PRSA) outputs obtained from a novel self-applicable non-invasive fetal electrocardiography (NIFECG) monitor with those from computerized cardiotocography (cCTG). A secondary objective was to evaluate the potential for remote assessment of fetal wellbeing by determining the relationship between PRSA and short-term variation (STV).

METHODS: This was a prospective observational study of women with a singleton pregnancy over 28 + 0 weeks' gestation attending a London teaching hospital for cCTG assessment. Participants underwent concurrent cCTG and NIFECG monitoring for up to 60 min. Averaged accelerative (AAC) and decelerative (ADC) capacities and STV were derived by postprocessing and filtration of signals, generating fully (F) and partially (P) filtered results. Linear correlation and accuracy and precision analysis were performed to assess the relationship between PRSA outputs from cCTG and NIFECG, using varying anchor thresholds, and their association with STV.

RESULTS: A total of 306 concurrent cCTG and NIFECG traces were collected from 285 women. F-filtered NIFECG PRSA (eAAC/eADC) results were generated from 65% of traces, whereas cCTG PRSA (cAAC/cADC) outputs were generated from all. Strong correlations were observed between cAAC and F-filtered eAAC (r = 0.879, P < 0.001) and between cADC and F-filtered eADC (r = 0.895, P < 0.001). NIFECG anchor detection decreased significantly with increasing signal loss, and NIFECG PRSA indices showed considerable deviation from those of cCTG when derived from traces in which fewer than 100 anchors were detected. Removing anchor filters from NIFECG traces to generate P-filtered PRSA outputs weakened the correlation (AAC: r = 0.505, P < 0.001; ADC: r = 0.560, P < 0.001). Lowering the anchor threshold to 100 increased the yield of eAAC and eADC outputs to approximately 74%, whilst maintaining strong correlation with cAAC (r = 0.839, P < 0.001) and cADC (r = 0.815, P < 0.001), respectively. Both cAAC and cADC showed a very strong linear relationship with cCTG STV (r = 0.928, P < 0.001 and r = 0.911, P < 0.001, respectively). Similar findings were observed with eAAC (r = 0.825, P < 0.001) and eADC (r = 0.827, P < 0.001).

CONCLUSIONS: PRSA appears to be a method of fetal assessment equivalent to STV, but, due to its innate ability to eliminate artifacts, it generates interpretable NIFECG traces with high accuracy at a higher rate. These findings raise the possibility of self-applied at-home or remote fetal heart-rate monitoring with automated reporting, thus enabling increased surveillance in high-risk women without impacting on service demand. © 2023 The Authors. Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of International Society of Ultrasound in Obstetrics and Gynecology.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。

引用

  1. Liu B, Thilaganathan B, Bhide A, et al. Phase-rectified signal averaging: correlation between two monitors and relationship with short-term variation of fetal heart rate. Ultrasound Obstet Gynecol. 2023. PMID: 36864541. doi:10.1002/uog.26192

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。