新型可穿戴胎儿心电图监测仪所测短期变异与计算机化胎心宫缩图的 相关性研究
前瞻性队列研究纳入285例孕28+0周以上孕妇,同步采集306段NIFECG与cCTG曲线,比较两者短期变异(STV)。完全滤波后eSTV与cSTV强相关(r=0.911),但仅46.4%曲线可生成结果;去除信号丢失>60%的曲线后65%曲线可生成结果且相关性更强(r=0.839)。
要点
- 完全滤波eSTV与cSTV强相关(r=0.911),但仅46.4%的1小时曲线能生成结果。
- 信号丢失>60%的曲线eSTV与cSTV差异大;剔除后65%曲线仍可生成eSTV。
- NIFECG需信号滤波、剔除低质量曲线及信号丢失校正后才能可靠计算STV。
核心结论
- NIFECG与cCTG计算的STV存在差异,需信号滤波、剔除低质量曲线及eSTV校正;校正后NIFECG可生成与cCTG强相关的STV值。
研究设计与样本
- 前瞻性队列研究,纳入285例孕28+0周以上单胎妊娠孕妇,在伦敦一家三级医院行cCTG评估(2021年6月至2022年6月),同步进行NIFECG与cCTG监测最多1小时,共采集306段同步曲线。
诊断标准与关键数值
- 完全滤波eSTV与cSTV相关系数r=0.911(P<0.001),但仅142/306(46.4%)的1小时曲线可生成结果;部分滤波生成98.4%的eSTV数据,与cSTV相关性弱(r=0.337,P<0.001);剔除信号丢失>60%的曲线后,eSTV对65%的曲线仍可生成结果,且与cSTV相关性更强;对剩余eSTV值用线性回归校正信号丢失后,与cSTV相关性进一步提高(r=0.839,P<0.001)。
局限性
- 部分滤波生成更多eSTV数据(98.4%),但与cSTV相关性弱(r=0.337,P<0.001)。
- 信号丢失>60%的曲线中,eSTV与cSTV差异变得高度不一致。
- 完全滤波虽与cSTV强相关(r=0.911),但仅46.4%的1小时曲线能生成结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare short-term variation (STV) outputs from a novel self-applied non-invasive fetal electrocardiography (NIFECG) device with those obtained on computerized cardiotocography (cCTG). Technological and algorithmic limitations and mitigation strategies were also evaluated.
METHODS: This was a prospective cohort study of women with a singleton pregnancy over 28 + 0 weeks' gestation attending a tertiary London hospital for cCTG assessment between June 2021 and June 2022. Women underwent concurrent monitoring with both NIFECG and cCTG for up to 1 h. Postprocessing of NIFECG data using various filtering methods produced NIFECG-STV (eSTV) values, which were compared with cCTG-STV (cSTV) outputs. Linear correlation, mean bias, precision and limits of agreement were assessed for STV derived by the different methods of computation and mathematical correction.
RESULTS: Overall, 306 concurrent NIFECG and cCTG traces were collected from 285 women. Fully filtered eSTV was correlated very strongly with cSTV (r = 0.911, P < 0.001), but generated results only in 142/306 (46.4%) 1-h traces owing to the removal of traces with lower-quality signals. Partial filtering generated more eSTV data (98.4%), but with a weak correlation with cSTV (r = 0.337, P < 0.001). The difference in STV between the monitors (eSTV - cSTV) increased with signal loss; in traces with > 60% signal loss, the values became highly discrepant. Removal of traces with > 60% signal loss resulted in a stronger correlation with cSTV, while still generating eSTV results for 65% of traces. Correcting these remaining eSTV values for signal loss using linear regression analysis further improved correlation with cSTV (r = 0.839, P < 0.001).
CONCLUSIONS: The discrepancy between STV computed by NIFECG and cCTG necessitates signal filtering, exclusion of poor-quality traces and eSTV correction. This study demonstrates that, with such correction, NIFECG is able to produce STV values that are strongly correlated with those of cCTG. This evidence base for NIFECG monitoring and interpretation is a promising step forward in the development of safe and effective at-home fetal heart-rate monitoring. © 2023 The Authors. Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of International Society of Ultrasound in Obstetrics and Gynecology.
引用
- Liu B, Thilaganathan B, Bhide A, et al. Correlation of short-term variation derived from novel ambulatory fetal electrocardiography monitor with computerized cardiotocography. Ultrasound Obstet Gynecol. 2023. PMID: 36864543. doi:10.1002/uog.26191
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。