一种自动计算噪声、空间分辨率和对比度影像质量指标的新算法:体模与临床腹部CT中的概念验证及与主观评分的一致性

A New Algorithm for Automatically Calculating Noise, Spatial Resolution, and Contrast Image Quality Metrics: Proof-of-Concept and Agreement With Subjective Scores in Phantom and Clinical Abdominal CT.
PubMed 文献 · Invest Radiol · 2023-01-01

本研究开发了一种概念验证计算机算法,用于自动计算腹部门静脉期CT的噪声、空间分辨率和对比度影像质量指标,并在体模和患者数据集上评估其与放射科医生主观评分的一致性。算法能100%自动计算所有指标,多个指标与主观评分显著相关,且在「不足」与「充足」影像质量分类间存在显著差异,展示了算法客观评估影像质量的可行性与潜力。

要点

  • 算法可自动计算体模和临床腹部CT的噪声、空间分辨率及对比度影像质量指标,覆盖全部扫描。
  • 体模数据中噪声4/5、空间分辨率1/1、对比度10/13个参数与主观评分显著相关。
  • 患者数据中噪声4/5、空间分辨率1/1、对比度16/18个参数与主观评分显著相关。

核心结论

  • 所开发算法能成功计算体模和临床腹部CT扫描的客观噪声、空间分辨率和对比度影像质量指标;三类多个指标与放射科医生主观评分一致,且在「不足」与「充足」影像质量扫描间显著不同,表明算法客观评估影像质量的可行性与潜力,可能有助于日常临床实践中的优化和质量控制。

研究设计与样本

  • 概念验证研究,开发计算机算法并在2个体模CT数据集(各n=57,来自2台不同扫描仪)和1个患者腹部CT数据集(n=510)上应用;体模数据集通过改变扫描仪设置(管电压、管电流、重建算法)获得从高到低影像质量,患者数据集将30例连续门静脉期扫描按较低mAs重建获得低影像质量图像;以5点Likert量表主观评分为参照。

诊断标准与关键数值

  • 体模数据集计算5个噪声、1个空间分辨率和13个对比度参数,患者数据集计算5个噪声、1个空间分辨率和18个对比度参数;体模中与主观评分显著相关的为噪声4/5(R²范围0.55-0.70)、空间分辨率1/1(R²=0.21和0.26)、对比度10/13(R²范围0.11-0.73);患者中为噪声4/5(R²范围0.019-0.096)、空间分辨率1/1(R²=0.11)、对比度16/18(R²范围0.008-0.116)。

临床应用与阅片要点

  • 算法可自动计算腹部门静脉期CT的客观影像质量指标,多个指标与放射科医生主观评分一致,作者认为该算法应适用于任何扫描,并可能通过日常临床实践中的自动影像质量评估帮助优化和质量控制。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aims of this study were to develop a proof-of-concept computer algorithm to automatically determine noise, spatial resolution, and contrast-related image quality (IQ) metrics in abdominal portal venous phase computed tomography (CT) imaging and to assess agreement between resulting objective IQ metrics and subjective radiologist IQ ratings.

MATERIALS AND METHODS: An algorithm was developed to calculate noise, spatial resolution, and contrast IQ parameters. The algorithm was subsequently used on 2 datasets of anthropomorphic phantom CT scans, acquired on 2 different scanners (n = 57 each), and on 1 dataset of patient abdominal CT scans (n = 510). These datasets include a range of high to low IQ: in the phantom dataset, this was achieved through varying scanner settings (tube voltage, tube current, reconstruction algorithm); in the patient dataset, lower IQ images were obtained by reconstructing 30 consecutive portal venous phase scans as if they had been acquired at lower mAs. Five noise, 1 spatial, and 13 contrast parameters were computed for the phantom datasets; for the patient dataset, 5 noise, 1 spatial, and 18 contrast parameters were computed. Subjective IQ rating was done using a 5-point Likert scale: 2 radiologists rated a single phantom dataset each, and another 2 radiologists rated the patient dataset in consensus. General agreement between IQ metrics and subjective IQ scores was assessed using Pearson correlation analysis. Likert scores were grouped into 2 categories, "insufficient" (scores 1-2) and "sufficient" (scores 3-5), and differences in computed IQ metrics between these categories were assessed using the Mann-Whitney U test.

RESULTS: The algorithm was able to automatically calculate all IQ metrics for 100% of the included scans. Significant correlations with subjective radiologist ratings were found for 4 of 5 noise ( R2 range = 0.55-0.70), 1 of 1 spatial resolution ( R2 = 0.21 and 0.26), and 10 of 13 contrast ( R2 range = 0.11-0.73) parameters in the phantom datasets and for 4 of 5 noise ( R2 range = 0.019-0.096), 1 of 1 spatial resolution ( R2 = 0.11), and 16 of 18 contrast ( R2 range = 0.008-0.116) parameters in the patient dataset. Computed metrics that significantly differed between "insufficient" and "sufficient" categories were 4 of 5 noise, 1 of 1 spatial resolution, 9 and 10 of 13 contrast parameters for phantom the datasets and 3 of 5 noise, 1 of 1 spatial resolution, and 10 of 18 contrast parameters for the patient dataset.

CONCLUSION: The developed algorithm was able to successfully calculate objective noise, spatial resolution, and contrast IQ metrics of both phantom and clinical abdominal CT scans. Furthermore, multiple calculated IQ metrics of all 3 categories were in agreement with subjective radiologist IQ ratings and significantly differed between "insufficient" and "sufficient" IQ scans. These results demonstrate the feasibility and potential of algorithm-determined objective IQ. Such an algorithm should be applicable to any scan and may help in optimization and quality control through automatic IQ assessment in daily clinical practice.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Jeukens CRLPN, Brauer MTH, Mihl C, et al. A New Algorithm for Automatically Calculating Noise, Spatial Resolution, and Contrast Image Quality Metrics: Proof-of-Concept and Agreement With Subjective Scores in Phantom and Clinical Abdominal CT. Invest Radiol. 2023. PMID: 36719964. doi:10.1097/RLI.0000000000000954