双源光子计数探测器CT扫描仪高螺距多能冠状动脉CT血管成像在低碘对比剂剂量下的应用
在双源光子计数探测器CT上对27例患者用30 mL碘海醇350 mg/mL行高螺距(3.2)多能冠状动脉CTA,并与26例常规60 mL剂量对照。低对比剂组120 kV和50 keV图像各测量点平均CT值均>300 HU,50 keV较120 kV CT值增加96%;50 keV的CNR显著高于120 kV和100 keV。多能虚拟单能图像显著提高钙化斑块及支架处管腔狭窄评估信心,并使9例患者CAD-RADS分类发生改变。
要点
- 低对比剂组120 kV和50 keV图像在升主动脉、左主干及前降支远端平均CT值均>300 HU。
- 50 keV虚拟单能图像CT值较120 kV增加96%,CNR显著高于120 kV和100 keV。
- 加用100 keV虚拟单能图像可提高钙化斑块和支架处管腔狭窄评估信心(读者1),并使9例CAD-RADS分类改变。
核心结论
- 高螺距多能光子计数探测器CT模式可在低辐射和低碘对比剂剂量下获得诊断质量冠状动脉CTA图像,多能虚拟单能图像显著改善CNR、整体图像质量及钙化斑块和支架处管腔狭窄评估信心,并使9例患者CAD-RADS分类发生改变。
研究设计与样本
- 在临床双源光子计数探测器CT上对27例患者行高螺距(3.2)多能冠状动脉CTA(低对比剂剂量,30 mL碘海醇350 mg/mL),另26例为常规对比剂剂量(60 mL);重建低能阈值120 kV图像及50 keV、100 keV虚拟单能图像,矩阵1024×1024、层厚0.6 mm。
诊断标准与关键数值
- 低对比剂组120 kV图像平均CT值:左主干382.9±76.2 HU、前降支远端341.0±53.9 HU、升主动脉399.5±76.1 HU,各测量点均>300 HU。
- 低对比剂组50 keV图像平均CT值:左主干667.5±139.9 HU、前降支远端578.1±121.5 HU、升主动脉700.8±142.5 HU,各测量点均>300 HU。
- 50 keV图像CT值较120 kV增加96%。
- 50 keV图像CT值显著高于120 kV和100 keV虚拟单能图像(P<0.0001),CNR亦显著高于两者(P<0.0001)。
- 钙化斑块处管腔狭窄评估信心:读者1加用100 keV后中位评分100,高于单用50 keV(中位70)和120 kV(中位70),差异显著(P=0.001);读者2各图像类型中位评分均为100,无显著差异。
临床应用与阅片要点
- 低对比剂剂量方案(30 mL碘海醇350 mg/mL)下,50 keV虚拟单能图像可用于提升血管强化和整体图像质量的主观评分。
- 评估钙化斑块处管腔狭窄时,加用100 keV虚拟单能图像可提高读者1的评估信心(中位评分100 vs 单用50 keV和120 kV均为70)。
- 评估支架内管腔狭窄时,加用100 keV图像可提高两位读者的信心(读者1:50+100 keV中位100 vs 50 keV 82.5 vs 120 kV 82.5;读者2:50+100 keV中位100 vs 50 keV 90 vs 120 kV 90)。
- 50 keV虚拟单能图像噪声高于120 kV图像,但伪影评分与120 kV相当,锐利度非显著高于120 kV。
- 多能虚拟单能图像的可用性使9例患者的CAD-RADS分类发生改变。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to evaluate the high-helical pitch, multienergy (ME) scanning mode of a clinical dual-source photon-counting detector (PCD) computed tomography (CT) and the benefit of virtual monoenergetic images (VMIs) for low-contrast-dose coronary CT angiography (CTA).
MATERIALS AND METHODS: High-pitch (3.2) ME coronary CTA was performed in PCD-CT in 27 patients using low contrast dose (30 mL of iohexol 350 mg/mL) and in 26 patients at routine contrast dose (60 mL). Low-energy-threshold 120 kV images (also known as T3D images) and 50 kiloelectron volts (50 keV) and 100 kiloelectron volts (100 keV) VMIs were reconstructed using a 1024 × 1024 matrix and 0.6-mm slices. The CT numbers, noise, and contrast-to-noise ratio (CNR) were measured in the ascending aorta (AA), left main coronary artery (LMCA), and distal left anterior descending (LAD) artery. Confidence in grading luminal stenosis with calcific plaque, noncalcific plaque, and stent was evaluated by 2 independent readers on a 0-100 scale (0 the lowest), and a CAD-RADS score was assigned. Image contrast enhancement, sharpness, noise, artifacts, and overall image quality were rated using a 5-point ordinal scale (1 the lowest).
RESULTS: The radiation doses (CTDI) in low- and routine-contrast cohorts were 2.5 ± 0.6 mGy and 3.1 ± 1.7 mGy, respectively ( P = 0.12). At all measured locations, the mean CT number was >300 HU in 120 kV (LMCA 382.9 ± 76.2, distal LAD 341.0 ± 53.9, AA 399.5 ± 76.1) and 50 keV images (LMCA 667.5 ± 139.9, distal LAD 578.1 ± 121.5, AA 700.8 ± 142.5) in the low-contrast cohort, with a 96% increase in CT numbers for 50 keV over 120 kV. The CT numbers were significantly higher ( P < 0.0001) in 50 keV than 120 kV and 100 keV VMI. The CNR was also significantly ( P < 0.0001) higher in 50 keV than 120 kV and 100 keV images in all vessels. Confidence in the assessment of luminal stenosis in the presence of calcific plaque was significantly higher ( P = 0.001) with the addition of 100 keV VMI (median score, 100) than using 50 keV alone (median score, 70) and 120 kV (median score, 70) for reader 1, but no significant differences were seen for reader 2 who had same median scores of 100 for all image types. The confidence in the assessment of luminal stenosis within a stent improved with the use of 100 keV images for both readers (reader 1: median scores for 50 + 100 keV = 100, 50 keV = 82.5, 120 kV = 82.5; reader 2: 50 + 100 keV = 100, 50 keV = 90, 120 kV = 90). There were no significant differences in confidence scores for assessment of luminal stenosis from noncalcific plaques for both readers. The reader-averaged qualitative scores for vascular enhancement and overall image quality were significantly higher for 50 keV VMI than for 120 kV images in both low- and routine-contrast dose cohorts. The image sharpness was nonsignificantly higher at 50 keV VMI than 120 kV images, and the artifact score was comparable for 50 keV VMI and 120 kV images. The noise was higher in 50 keV VMI than in 120 kV images.
CONCLUSIONS: High-pitch ME PCD-CT mode produced diagnostic quality coronary CTA images at low radiation and iodinated contrast doses. The availability of ME VMIs significantly improved the CNR, overall image quality, and confidence in assessment of luminal stenosis in the presence of calcific plaques and stents, and resulted in change of CAD-RADS categories in 9 patients.
引用
- Rajiah PS, Dunning CAS, Rajendran K, et al. High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose. Invest Radiol. 2023. PMID: 36822655. doi:10.1097/RLI.0000000000000961
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。