文献速递心血管CT

光子计数探测器CT虚拟非碘重建算法用于体外和体内冠状动脉钙化积分:虚拟单能级和量子迭代重建的影响

Photon Counting Detector CT-Based Virtual Noniodine Reconstruction Algorithm for In Vitro and In Vivo Coronary Artery Calcium Scoring: Impact of Virtual Monoenergetic and Quantum Iterative Reconstructions.
PubMed 文献 · Invest Radiol · 2023-01-01

在模拟3种体模及61例患者(最终分析34例有冠脉钙化者)中,评估光子计数探测器CT虚拟非碘重建的冠脉钙化积分(CACS VNI)与真实平扫(CACS TNC)的一致性。CACS VNI与CACS TNC强相关(r>0.9,P<0.001)且一致性极佳(ICC>0.9),但CACS VNI随keV降低而显著升高;不同量子迭代重建水平的影响不一致,似取决于个体条件,需进一步研究。

要点

  • CACS VNI与CACS TNC在所有VMI和QIR水平均强相关(r>0.9)且一致性极佳(ICC>0.9)。
  • CACS VNI随keV降低而显著升高,55 keV时部分病例出现高估。
  • 不同QIR水平对CACS VNI的影响不一致,似乎取决于多种个体条件。

核心结论

  • 虚拟单能级重建对CACS VNI有显著影响;不同QIR水平的效果不一致,似乎取决于多种个体条件,需进一步研究。

研究设计与样本

  • 体模研究结合患者研究:使用 anthropomorphic 胸部体模通过2个延伸环模拟3种患者体型(小:300×200 mm,中:350×250 mm,大:400×300 mm),并纳入61例患者(最终仅分析有冠脉钙化的34例;65.4±10.0岁;73.5%男性),均行光子计数探测器系统非增强和增强心电门控心脏CT;以70 keV、QIR关闭的TNC扫描为参考。

关键影像表现/征象

  • 体外和体内CACS VNI在所有研究的VMI和QIR水平均与CACS TNC强相关(r>0.9,P<0.001)且一致性极佳(ICC>0.9)。
  • 体模和患者CACS VNI随keV降低而显著升高(体外:80 keV 475.2±26.3升至55 keV 652.5±42.2;体内:80 keV 142.5 [7.4/737.7]升至55 keV 248.1 [31.2/1144];P<0.001)。
  • 55 keV时CACS VNI部分病例较70 keV CACS TNC高估(体外625.8±24.4;体内225.4 [35.1/959.7])。
  • 体外低keV下CACS随QIR升高而增加;体内仅60和80 keV时QIR 1的评分显著高于QIR 4(60 keV:220.3 [29.6-1060] vs 219.5 [23.7/1048];80 keV:152.0 [12.0/735.6] vs 142.5 [7.4/737.7];P<0.001)。

诊断标准与关键数值

  • CACS VNI与CACS TNC最接近的条件:小体模60 keV、QIR 2(+0.1%);中体模55 keV、QIR 1(±0%);大体模55 keV、QIR 4(-0.1%);患者60 keV、QIR 1(-2.3%)。
  • 参考标准为70 keV、QIR关闭的TNC扫描(CACS TNC)。

临床应用与阅片要点

  • 采用VNI重建行CACS时,虚拟单能级水平显著影响CACS VNI,55 keV时部分病例出现高估。
  • 不同QIR水平的效果不一致,似乎取决于多种个体条件,需进一步研究。

局限性

  • 不同QIR水平的效果不一致,似乎取决于多种个体条件,需进一步研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to evaluate the impact of virtual monoenergetic imaging (VMI) and quantum iterative reconstruction (QIR) on the accuracy of coronary artery calcium scoring (CACS) using a virtual noniodine (VNI) reconstruction algorithm on a first-generation, clinical, photon counting detector computed tomography system.

MATERIALS AND METHODS: Coronary artery calcium scoring was evaluated in an anthropomorphic chest phantom simulating 3 different patient sizes by using 2 extension rings (small: 300 × 200 mm, medium: 350 × 250 mm, large: 400 × 300 mm) and in patients (n = 61; final analyses only in patients with coronary calcifications [n = 34; 65.4 ± 10.0 years; 73.5% male]), who underwent nonenhanced and contrast-enhanced, electrocardiogram-gated, cardiac computed tomography on a photon counting detector system. Phantom and patient data were reconstructed using a VNI reconstruction algorithm at different VMI (55-80 keV) and QIR (strength 1-4) levels (CACS VNI ). True noncontrast (TNC) scans at 70 keV and QIR "off" were used as reference for phantom and patient studies (CACS TNC ).

RESULTS: In vitro and in vivo CACS VNI showed strong correlation ( r > 0.9, P < 0.001 for all) and excellent agreement (intraclass correlation coefficient > 0.9 for all) with CACS TNC at all investigated VMI and QIR levels. Phantom and patient CACS VNI significantly increased with decreasing keV levels (in vitro: from 475.2 ± 26.3 at 80 keV up to 652.5 ± 42.2 at 55 keV; in vivo: from 142.5 [7.4/737.7] at 80 keV up to 248.1 [31.2/1144] at 55 keV; P < 0.001 for all), resulting in an overestimation of CACS VNI at 55 keV compared with CACS TNC at 70 keV in some cases (in vitro: 625.8 ± 24.4; in vivo: 225.4 [35.1/959.7]). In vitro CACS increased with rising QIR at low keV. In vivo scores were significantly higher at QIR 1 compared with QIR 4 only at 60 and 80 keV (60 keV: 220.3 [29.6-1060] vs 219.5 [23.7/1048]; 80 keV: 152.0 [12.0/735.6] vs 142.5 [7.4/737.7]; P < 0.001). CACS VNI was closest to CACS TNC at 60 keV, QIR 2 (+0.1%) in the small; 55 keV, QIR 1 (±0%) in the medium; 55 keV, QIR 4 (-0.1%) in the large phantom; and at 60 keV, QIR 1 (-2.3%) in patients.

CONCLUSIONS: Virtual monoenergetic imaging reconstructions have a significant impact on CACS VNI . The effects of different QIR levels are less consistent and seem to depend on several individual conditions, which should be further investigated.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Fink N, Zsarnoczay E, Schoepf UJ, et al. Photon Counting Detector CT-Based Virtual Noniodine Reconstruction Algorithm for In Vitro and In Vivo Coronary Artery Calcium Scoring: Impact of Virtual Monoenergetic and Quantum Iterative Reconstructions. Invest Radiol. 2023. PMID: 36822677. doi:10.1097/RLI.0000000000000959

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。