使用Definity微泡的次谐波辅助压力估测技术无创评估心腔压力
在非随机、经机构审查委员会批准的临床试验(NCT03243942)中,对计划接受左心和/或右心导管术的患者输注Definity微泡,用SonixTablet扫描仪同步采集次谐波辅助压力估测(SHAPE)数据并与压力导管数据对比。左心室最小压和舒张末压的绝对误差均值分别为2.9±2.0和1.7±1.2 mmHg(n=26),右心室收缩压为2.2±1.5 mmHg(n=11)。作者认为SHAPE技术联合Definity微泡无创、准确获取心内压的结果令人鼓舞。
要点
- 左心室最小压和舒张末压的绝对误差均值分别为2.9±2.0和1.7±1.2 mmHg(n=26)。
- 右心室收缩压的绝对误差均值为2.2±1.5 mmHg(n=11)。
- Definity输注期间发生2例不良事件,均已缓解。
核心结论
- 作者认为,使用Definity微泡的SHAPE技术无创、准确获取心内压的结果令人鼓舞。
研究设计与样本
- 非随机、经机构审查委员会批准的临床试验(NCT03243942),纳入计划接受左心和/或右心导管术并签署书面知情同意的患者,使用SonixTablet扫描仪(BK Medical Systems)采集SHAPE数据,以压力导管数据为参考。
诊断标准与关键数值
- 左心室最小压绝对误差均值2.9±2.0 mmHg(n=26),左心室舒张末压绝对误差均值1.7±1.2 mmHg(n=26)。
- 右心室收缩压绝对误差均值2.2±1.5 mmHg(n=11)。
- SHAPE数据与SphygmoCor XCEL PWA设备(ATCOR Medical)测量结合,导出转换因子(mmHg/dB),再联合左和/或右心室SHAPE数据获得临床相关的收缩期和舒张期心室压。
局限性
- Definity输注期间发生2例不良事件,均已缓解。
- 原文结论未据此给出临床推荐。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Noninvasive and accurate assessment of intracardiac pressures has remained an elusive goal of noninvasive cardiac imaging.
OBJECTIVES: The purpose of this study was to investigate if errors in intracardiac pressures obtained noninvasively using contrast microbubbles and the subharmonic-aided pressure estimation (SHAPE) technique are <5 mm Hg.
METHODS: In a nonrandomized institutional review board-approved clinical trial (NCT03243942), patients scheduled for a left-sided and/or right-sided heart catheterization procedure and providing written informed consent were included. A standard-of-care catheterization procedure was performed advancing clinically used pressure catheters into the left and/or right ventricles and/or the aorta. After pressure catheter placement, patients received an infusion of Definity microbubbles (n = 56; 2 vials diluted in 50 mL of saline; infusion rate: 4-10 mL/min) (Lantheus Medical Imaging). Then SHAPE data was acquired using a validated interface developed on a SonixTablet scanner (BK Medical Systems) synchronously with the pressure catheter data. A conversion factor (mm Hg/dB) was derived from SHAPE data and measurements with a SphygmoCor XCEL PWA device (ATCOR Medical) and was combined with SHAPE data from the left and/or the right ventricles to obtain clinically relevant systolic and diastolic ventricular pressures.
RESULTS: The mean value of absolute errors for left ventricular minimum and end diastolic pressures were 2.9 ± 2.0 and 1.7 ± 1.2 mm Hg (n = 26), respectively, and for right ventricular systolic pressures was 2.2 ± 1.5 mm Hg (n = 11). Two adverse events occurred during Definity infusion; both were resolved.
CONCLUSIONS: These results indicate that the SHAPE technique with Definity microbubbles is encouragingly efficacious for obtaining intracardiac pressures noninvasively and accurately. (Noninvasive, Subharmonic Intra-Cardiac Pressure Measurement; NCT03243942).