对比增强超声的伪影与技术考量

Artifacts and Technical Considerations at Contrast-enhanced US.
PubMed 文献 · Radiographics · 2023-01-01

本文回顾对比增强超声(CEUS)中因微泡造影剂特性及成像技术产生的伪影,包括灰阶超声常见伪影在减影图像上的表现及非线性伪影和无意微泡破坏所致假性廓清。作者强调识别并减轻这些伪影可提高检查质量和诊断信心。

要点

  • 减影图像上可见灰阶超声常见伪影如声影、混响、折射和穿透不良。
  • 非线性伪影和无意微泡破坏可引起假性廓清。
  • 微泡本身可致假性增强、信号饱和与衰减伪影并干扰多普勒。

核心结论

  • 识别并理解CEUS相关伪影及其减轻方法可能提高影像检查质量、增强使用者信心并改善患者照护。

研究设计与样本

  • 综述;回顾超声造影剂原理、声-微泡相互作用及图像生成技术,讨论技术因素与具体伪影的识别及纠正措施。

关键影像表现/征象

  • 灰阶超声常见伪影如声影、声增强、混响、折射、反射和穿透不良可出现在仅造影图像上。
  • 非线性伪影和无意微泡破坏可导致假性廓清。
  • 微泡本身可产生假性增强、信号饱和以及衰减和声影伪影。

临床应用与阅片要点

  • 需识别非线性伪影和假性廓清,避免误判为真实廓清。
  • 注意微泡所致假性增强和信号饱和可影响强化评估。
  • 使用彩色和频谱多普勒时需注意微泡伪影的干扰。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Contrast-enhanced US (CEUS), similar to other radiologic modalities, requires specific technical considerations and is subject to image artifacts. These artifacts may affect examination quality, negatively impact diagnostic accuracy, and decrease user comfort when using this emerging technique. Some artifacts are related to commonly known gray-scale US artifacts that can also appear on the contrast-only image (tissue-subtracted image obtained with the linear responses from background tissues nulled). These may include acoustic shadowing and enhancement; reverberation, refraction, and reflection; and poor penetration. Other artifacts are exclusive to CEUS owing to the techniques used for contrast mode image generation and the unique properties of the microbubbles that constitute ultrasound-specific contrast agents (UCAs). UCA-related artifacts may appear on the contrast-only image, the gray-scale image, or various Doppler mode images. Artifacts related to CEUS may include nonlinear artifacts and unintentional microbubble destruction resulting in pseudowashout. The microbubbles themselves may result in specific artifacts such as pseudoenhancement, signal saturation, and attenuation and shadowing and can confound the use of color and spectral Doppler US. Identifying and understanding these artifacts and knowing how to mitigate them may improve the quality of the imaging study, increase user confidence, and improve patient care. The authors review the principles of UCAs and the sound-microbubble interaction, as well as the technical aspects of image generation. Technical considerations, including patient positioning, depth, acoustic window, and contrast agent dose, also are discussed. Specific artifacts are described, with tips on how to identify and, if necessary, apply corrective measures, with the goal of improving examination quality. © RSNA, 2022 Online supplemental material and the slide presentation from the RSNA Annual Meeting are available for this article.

来源:Radiographics,查看原文。版权归原作者所有。

引用

  1. Fetzer DT, Vijay K, Caserta MP, et al. Artifacts and Technical Considerations at Contrast-enhanced US. Radiographics. 2023. PMID: 36563094. doi:10.1148/rg.220093

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。