基于3D打印模型的血管迂曲度对血管内血栓切除技术疗效的体外分析
本研究利用3D打印的非迂曲与急性成角迂曲脑血管解剖流动模型,在球囊导引导管近端保护下测试接触抽吸、支架取栓器及联合方法三种血栓切除技术的首过效应和远端栓塞率。共完成30次血栓切除实验,非迂曲解剖的首过效应总体发生率为80.0% versus 迂曲解剖46.7%,远端栓塞总体发生率非迂曲26.7% versus 迂曲46.7%。结论认为迂曲血管解剖可能恶化首过效应和远端栓塞率,联合技术因取栓过程中血栓剪切脱落现象未能显示结局改善。
要点
- 迂曲血管解剖下首过效应总体发生率为46.7%,低于非迂曲解剖的80.0%。
- 迂曲血管解剖下远端栓塞总体发生率为46.7%,高于非迂曲解剖的26.7%。
- 迂曲模型中接触抽吸技术首过效应80.0%、远端栓塞率20%,优于其他技术。
核心结论
- 迂曲血管解剖可能恶化首过效应和远端栓塞率;联合技术因取栓过程中血栓剪切脱落现象未能显示结局改善。
研究设计与样本
- 体外研究,利用真实患者血管造影数据制造3D打印非迂曲与急性成角迂曲血管解剖流动模型,在球囊导引导管近端保护下测试接触抽吸、支架取栓器和联合方法三种血栓切除技术,共完成30次血栓切除实验。
诊断标准与关键数值
- 非迂曲解剖首过效应总体发生率为80.0%,迂曲解剖为46.7%。
- 非迂曲解剖远端栓塞总体发生率为26.7%,迂曲解剖为46.7%。
- 迂曲模型中接触抽吸技术首过效应为80.0%,远端栓塞率为20%。
临床应用与阅片要点
- 联合技术在非迂曲和迂曲解剖中均未显示首过效应和远端栓塞的显著优势。
- 联合组血栓切除失败的主要机制为血栓剪切脱落。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Achieving complete recanalization with the front-line endovascular thrombectomy device improves the outcome of acute stroke. The aim of this study was to evaluate whether various thrombectomy techniques including contact aspiration, stent retriever thrombectomy, and combination therapy differ in first-pass effect and distal emboli in acute large-vessel occlusion simulated using 3D printed nontortuous and tortuous cerebrovascular anatomy models.
MATERIALS AND METHODS: 3D printed flow models were manufactured using angiographic data of nontortuous and acutely angulated tortuous vascular anatomy from real patients. Three thrombectomy techniques, contact aspiration, stent retriever, and combined methods, were tested under proximal protection with the balloon-guiding catheter. The first-pass effect and distal emboli rates were analyzed in addition to the thrombectomy-failure mechanisms of the respective techniques.
RESULTS: A total of 30 thrombectomy experiments were performed. The overall incidence of first-pass effect in the nontortuous and tortuous anatomy was 80.0% versus 46.7%. The overall incidence of distal emboli in the nontortuous and tortuous anatomy was 26.7% versus 46.7%. The contact aspiration technique showed better first-pass effect (80.0%) and distal emboli rates (20%) in the tortuous model compared with other techniques. The combined technique did not show remarkable superiority of the first-pass effect and distal emboli in either the nontortuous or tortuous anatomy. Shearing off of the thrombus was the main mechanism of thrombectomy failure in the combined group.
CONCLUSIONS: The tortuous vascular anatomy may worsen the first-pass effect and distal emboli rates. The combined techniques failed to show improvement in outcome due to the shearing-off phenomenon of the thrombus during retrieval.
引用
- Kim JH, Kim BM, Kim DJ, et al. In Vitro Analysis of the Efficacy of Endovascular Thrombectomy Techniques according to the Vascular Tortuosity Using 3D Printed Models. AJNR Am J Neuroradiol. 2022. PMID: 36574330. doi:10.3174/ajnr.A7633
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。