乳腺癌非金属闪烁标记的超声检测与手术取出:初步研究

Ultrasonographic Detection and Surgical Retrieval of a Nonmetallic Twinkle Marker in Breast Cancer: Pilot Study.
PubMed 文献 · Radiol Imaging Cancer · 2022-01-01

这项前瞻性初步研究纳入8例完成化疗的淋巴结阳性乳腺癌患者,在术前放射性粒子定位时放置非金属闪烁标记,用五点评分法比较闪烁标记与标准活检夹、放射性粒子在超声、钼靶、标本X线及病理标本上的可识别性。闪烁标记超声中位评分2.0,优于活检夹的2.4,6/8例超声上最易识别,8枚标记均在定位后1-3天被外科取出。

要点

  • 非金属闪烁标记在彩色多普勒超声上的可识别性中位评分为2.0,优于标准活检夹的2.4。
  • 8例中有6例闪烁标记在超声上最易识别,8枚标记均在定位后1–3天被手术取出。
  • 标本X线16次判读中,粒子、活检夹、闪烁标记均评为非常容易识别。

核心结论

  • 这项初步研究提示,闪烁组织标记的短期安全性及超声检测的简便性可能与活检夹相当。

研究设计与样本

  • 前瞻性初步研究,2020年11月至2021年7月纳入8例完成化疗的淋巴结阳性乳腺癌患者,在术前放射性粒子定位时放置非金属闪烁标记,采用五点评分法(1=最容易,5=最难)评价超声、钼靶、标本X线及大体病理标本上的识别难易,使用描述性统计。

关键影像表现/征象

  • 超声可识别性中位评分:放射性粒子3.9(范围3.7–5.0),活检夹2.4(范围1.0–5.0),闪烁标记2.0(范围1.0–4.3)。
  • 8例中6例闪烁标记在超声上最易识别。
  • 定位后钼靶8例中7例的粒子、活检夹、闪烁标记均评为「非常容易」识别。
  • 标本X线共16次判读,粒子、活检夹、闪烁标记均评为非常容易识别。
  • 病理检查中活检夹在全部患者中均为最难识别的器械,闪烁标记在8例中7例为最易识别。

诊断标准与关键数值

  • 五点评分法:1=最容易识别,5=最难识别,用于评价超声、钼靶、标本X线及大体病理标本上的识别难易。

临床应用与阅片要点

  • 作者认为闪烁组织标记的短期安全性及超声检测的简便性可能与活检夹相当。

局限性

  • 本文为初步研究(pilot study)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Purpose To evaluate the short-term safety of a nonmetallic twinkle marker and compare its conspicuity at color Doppler US with that of standard breast biopsy clips and radioactive seeds by using B-mode US in axillary lymph nodes. Materials and Methods This prospective study (November 2020-July 2021) of participants with node-positive breast cancer who completed chemotherapy involved placing a twinkle marker at the time of preoperative radioactive seed localization. A five-point scoring system (1 = easiest, 5 = most difficult) was used to rate the ease of identifying the clip, seed, and twinkle marker on postlocalization sonograms, mammograms, specimen radiographs, and gross pathologic specimens. Descriptive statistics were used. Results Eight women (mean age, 57 years ± 16 [SD]) were enrolled. The median scores for US conspicuity of each device were 3.9 (range, 3.7-5.0) for the radioactive seed, 2.4 (range, 1.0-5.0) for the clip, and 2.0 (range, 1.0-4.3) for the twinkle marker. In six of eight participants, the twinkle marker was the most identifiable at US. The seeds, clips, and twinkle markers were scored "very easy" to identify on seven of eight postlocalization mammograms. The surgeon retrieved all eight twinkle markers 1-3 days after localization. In all 16 interpretations, the seeds, clips, and twinkle markers were rated as very easy to identify on specimen radiographs. The clip was the most difficult device to identify at pathologic examination in all participants, and the twinkle marker was the easiest to identify in seven of eight participants. Conclusion This pilot study demonstrates that the safety and ease of US detection of a twinkling tissue marker may be comparable to a biopsy clip. Keywords: Ultrasonography, US-Doppler, Breast, Localization, Surgery Clinical trial registration no. NCT04674852 © RSNA, 2022.