急性卒中患者最新一代血管造影系统的有效剂量测量:与最新多排CT的对比
本研究使用体模和金属氧化物半导体场效应晶体管剂量计,估算新一代平板探测器CT血管造影和灌注方案的有效剂量,并与多排CT方案比较。结果显示平板探测器CT灌注有效剂量为4.52 mSv(无准直)和2.88 mSv(准直),多排CT灌注为2.17 mSv;平板探测器CT血管造影为0.91 mSv,多排CT血管造影为1.35 mSv。结论认为平板探测器CT灌注和血管造影的有效剂量与多排CT方案无实质性差异。
要点
- 平板探测器CT灌注有效剂量:无准直4.52 mSv,准直2.88 mSv;多排CT灌注2.17 mSv。
- 平板探测器CT血管造影有效剂量0.91 mSv,多排CT血管造影1.35 mSv。
- 现代双翼血管造影系统上行平板探测器CT灌注和血管造影,有效剂量与多排CT方案无实质性差异。
核心结论
- 现代双翼血管造影系统上行平板探测器CT灌注和血管造影,患者有效剂量与多排CT方案无实质性差异。
研究设计与样本
- 体模剂量学研究,使用人体模体和金属氧化物半导体场效应晶体管剂量计,按国际放射防护委员会103号出版物估算有效剂量;体模内设100个测量点,表面放置18个剂量计;所有方案遵循制造商规格,患者体位和准直按常规临床实践执行;测量在最新一代血管造影和多排CT系统上以相同剂量计放置方式完成。
诊断标准与关键数值
- 平板探测器CT灌注(无准直)有效剂量4.52 mSv。
- 平板探测器CT灌注(准直)有效剂量2.88 mSv。
- 多排CT灌注有效剂量2.17 mSv。
- 平板探测器CT血管造影(纵向覆盖范围与多排CT血管造影相当)有效剂量0.91 mSv。
- 多排CT血管造影有效剂量1.35 mSv。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Patients with acute ischemic stroke are increasingly triaged with one-stop management approaches, resulting in baseline imaging with a flat detector CT scanner. This study aimed to estimate the effective dose to a patient of a novel cervical and intracranial flat detector CT angiography and a flat detector CT perfusion protocol and to compare it with the effective dose of analogous multidetector row CT protocols.
MATERIALS AND METHODS: We estimated the effective dose to the patient according to the International Commission on Radiological Protection 103 using an anthropomorphic phantom with metal oxide semiconductor field effect transistor dosimeters. Placement was according to the organ map provided by the phantom manufacturer. We used 100 measurement points within the phantom, and 18 metal oxide semiconductor field effect transistor dosimeters were placed on the surface of the phantom. All protocols followed the manufacturer's specifications, and patient positioning and collimation were performed as in routine clinical practice. Measurements were obtained on the latest-generation angiography and multidetector row CT systems with identical placement of the metal oxide semiconductor field effect transistor dosimeters.
RESULTS: The estimated effective doses of the investigated perfusion protocols were 4.52 mSv (flat detector CT perfusion without collimation), 2.88 mSv (flat detector CT perfusion with collimation), and 2.17 mSv (multidetector row CT perfusion). A novel protocol called portrait flat detector CT angiography that has a z-axis coverage area comparable with that of multidetector row CT angiography had an estimated effective dose of 0.91 mSv, while the dose from multidetector row CT was 1.35 mSv.
CONCLUSIONS: The estimated effective dose to the patient for flat detector CT perfusion and angiography on a modern biplane angiography system does not deviate substantially from that of analogous multidetector row CT protocols.