文献速递骨肌CTDR

无龙门架高分辨率锥形束CT:对桡骨远端与舟骨骨折检出及特征描述的效能

Gantry-Free High-Resolution Cone-Beam CT: Efficacy for Distal Radius and Scaphoid Fracture Detection and Characterization.
PubMed 文献 · Acad Radiol · 2022-09-25

本研究纳入12个月内113例严重腕部外伤患者,术前均行X线摄影与无龙门架锥形束CT(CBCT)检查,以手术报告为参考标准。CBCT对舟骨骨折、桡骨远端多块骨折及关节面受累的准确性高于X线摄影;CBCT中位有效剂量为3.65 µSv,X线摄影为0.16 µSv。

要点

  • CBCT对舟骨骨折诊断准确性高于X线摄影(阅片者1:100.0% vs 75.0%;阅片者2:100.0% vs 65.0%)。
  • CBCT对桡骨远端多块骨折(100.0% vs 90.5%;100.0% vs 93.7%)及关节面受累(99.0% vs 84.2%;100.0% vs 83.2%)准确性更高。
  • CBCT中位有效剂量3.65 µSv,X线摄影0.16 µSv,原文称极低剂量即可维持高图像质量。

核心结论

  • 无龙门架CBCT对桡骨远端与舟骨骨折形态学评估具有极佳诊断准确性,即使活动受限患者以极低辐射剂量即可维持高图像质量。

研究设计与样本

  • 12个月内113例严重腕部外伤患者(共115只腕,含2例双侧损伤)术前均行X线摄影与无龙门架多用途扫描仪CBCT,两名放射科医生回顾性分析桡骨远端骨折(n=95)与舟骨骨折(n=20)形态,以手术报告为参考标准。

诊断标准与关键数值

  • 舟骨骨折诊断准确性:阅片者1 CBCT 100.0% vs X线摄影75.0%;阅片者2 CBCT 100.0% vs X线摄影65.0%。
  • 桡骨远端多块骨折准确性:阅片者1 CBCT 100.0% vs X线摄影90.5%;阅片者2 CBCT 100.0% vs X线摄影93.7%。
  • 关节面受累准确性:阅片者1 CBCT 99.0% vs X线摄影84.2%;阅片者2 CBCT 100.0% vs X线摄影83.2%。
  • 舟骨近极或腰部受累诊断准确性:阅片者1 CBCT 100.0% vs X线摄影70.0%;阅片者2 CBCT 100.0% vs X线摄影65.0%。
  • 舟骨粉碎性骨折诊断准确性:阅片者1 CBCT 100.0% vs X线摄影70.0%;阅片者2 CBCT 100.0% vs X线摄影75.0%。

临床应用与阅片要点

  • 无龙门架CBCT可用于严重腕部外伤术前评估桡骨远端与舟骨骨折形态。
  • CBCT中位有效剂量3.65 µSv,标准X线摄影0.16 µSv,原文称即使活动受限患者极低剂量也足以维持高图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Gantry-free cone-beam CT (CBCT) allows for comfortable patient positioning due to an open scanner architecture. Since CBCT without gantry is not yet established for clinical wrist trauma imaging, this study's aim was to investigate its diagnostic value in the preoperative workup of patients with distal radius and scaphoid fractures.

METHODS: Within a 12-month period, 113 patients with severe wrist trauma underwent both radiography and CBCT with the same gantry-free multi-use scanner before surgery. Two radiologists retrospectively analyzed all datasets for the morphology of distal radius (n = 95) and scaphoid fractures (n = 20). In all 115 wrists (two bilateral injuries), surgical reports served as the standard of reference.

RESULTS: While accuracy for distal radius fractures was comparable among CBCT and radiographs, the former was superior with regard to scaphoid fractures (Reader 1: 100.0% vs. 75.0%; Reader 2: 100.0% vs. 65.0%). Accuracy for multi-fragmentary radius injuries (100.0% vs. 90.5%; 100.0% vs. 93.7%), and articular affliction (99.0% vs. 84.2%; 100.0% vs. 83.2%) was also higher in CBCT. Regarding scaphoid fractures, CBCT proved superior for diagnosis of proximal pole or waist involvement (100.0% vs. 70.0%; 100.0% vs. 65.0%) and comminuted patterns (100.0% vs. 70.0%; 100.0% vs. 75.0%). Median effective dose of CBCT was as low as 3.65 µSv compared with 0.16 µSv for standard radiography.

CONCLUSION: Gantry-free CBCT allows for excellent diagnostic accuracy in the assessment of distal radius and scaphoid fracture morphology. Even in patients with limited mobility, very low radiation dose is sufficient to maintain high image quality.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Grunz JP, Jordan MC, Schmitt R, et al. Gantry-Free High-Resolution Cone-Beam CT: Efficacy for Distal Radius and Scaphoid Fracture Detection and Characterization. Acad Radiol. 2022. PMID: 36167629. doi:10.1016/j.acra.2022.08.030

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。