双能CT虚拟平扫检测急性缺血性卒中早期缺血改变:一项非劣效性分析
本研究纳入193例疑似急性缺血性卒中成人患者,以DWI为参考标准,比较虚拟平扫与常规平扫基于ASPECTS检测早期缺血改变的表现。虚拟平扫与参考标准的一致性相关系数为0.44,常规平扫为0.23,两者差值0.20(95%CI 0.01-0.39),未跨越预设非劣效界值-0.10,提示虚拟平扫非劣于常规平扫。
要点
- 虚拟平扫ASPECTS与DWI一致性相关系数0.44,高于常规平扫的0.23。
- 虚拟平扫与常规平扫一致性差值0.20(95%CI 0.01-0.39),未跨越非劣效界值-0.10。
- 100例(52%)患者DWI证实缺血,虚拟平扫可用于早期缺血改变评估。
核心结论
- 双能虚拟平扫在基于ASPECTS检测早期缺血改变方面非劣于常规平扫。
研究设计与样本
- 非劣效性分析,纳入193例疑似急性缺血性卒中成人患者,均行双能平扫、CTA及48小时内脑MR成像,以DWI为参考标准;常规平扫由标准线性融合图像重建模拟,虚拟平扫由CTA重建。
诊断标准与关键数值
- 虚拟平扫与参考标准ASPECTS一致性相关系数0.44(95%CI 0.33-0.53),常规平扫为0.23(95%CI 0.15-0.32);两者差值0.20(95%CI 0.01-0.39),未跨越预设非劣效界值-0.10。
临床应用与阅片要点
- 作者结论认为双能虚拟平扫在基于ASPECTS检测早期缺血改变方面非劣于常规平扫,原文未据此给出具体阅片操作推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Dual-energy virtual NCCT has the potential to replace conventional NCCT to detect early ischemic changes in acute ischemic stroke. In this study, we evaluated whether virtual NCCT is noninferior compared with standard linearly blended NCCT, a surrogate of conventional NCCT, regarding the detection of early ischemic changes with ASPECTS.
MATERIALS AND METHODS: Adult patients who presented with suspected acute ischemic stroke and who underwent dual-energy NCCT and CTA and brain MR imaging within 48 hours were included. Standard linearly blended images were reconstructed to match a conventional NCCT. Virtual NCCT images were reconstructed from CTA. ASPECTS was evaluated on conventional NCCT, virtual NCCT, and DWI, which served as the reference standard. Agreement between CT assessments and the reference standard was evaluated with the Lin concordance correlation coefficient. Noninferiority was assessed with bootstrapped estimates of the differences in ASPECTS between conventional and virtual NCCT with 95% CIs.
RESULTS: Of the 193 included patients, 100 patients (52%) had ischemia on DWI. Compared with the reference standard, the ASPECTS concordance correlation coefficient for conventional and virtual NCCT was 0.23 (95% CI, 0.15-0.32) and 0.44 (95% CI, 0.33-0.53), respectively. The difference in the concordance correlation coefficient between virtual and conventional NCCT was 0.20 (95% CI, 0.01-0.39) and did not cross the prespecified noninferiority margin of -0.10.
CONCLUSIONS: Dual-energy virtual NCCT is noninferior compared with conventional NCCT for the detection of early ischemic changes with ASPECTS.