1.5 T 与 3 T 下深度学习加速单次屏气腹部 HASTE 的综合临床评估
本研究为单中心回顾性研究,纳入 2020 年 8 月至 2021 年 2 月间在 1.5 T 和 3 T 下接受临床指征肝脏 MRI 的 320 例患者,比较深度学习加速单次屏气 HASTE(HASTEDL)与标准多激发涡轮自旋回波 T2 加权抑脂序列的图像质量、诊断信心和病灶检出能力,并记录采集时间缩短幅度。
要点
- HASTEDL 整体图像质量中位评分为 4(四分位距 3–4),表现优秀。
- BLADE 在整体及肾上腺图像质量、对比度、锐利度、伪影和诊断信心上评分显著更高(P < .001)。
- 噪声、肝胰脾肾图像质量及病灶数目与最大径无显著差异(P = 0.886;P = 0.120–0.366;ICC = 0.972–1.0)。
核心结论
- HASTEDL 相比标准 T2 序列提供优秀图像质量、良好诊断信心和病灶检出能力,并大幅缩短采集时间。
研究设计与样本
- 单中心回顾性研究,纳入 320 例在 1.5 T 和 3 T 下接受临床指征肝脏 MRI 的患者,比较 HASTEDL 与标准 T2 加权抑脂多激发涡轮自旋回波序列(BLADE)。
关键影像表现/征象
- HASTEDL 整体图像质量中位评分为 4(四分位距 3–4)。
- BLADE 整体图像质量中位评分为 4(四分位距 4–4),显著高于 HASTEDL(P < .001)。
- BLADE 在肾上腺图像质量、对比度、锐利度、伪影和诊断信心上评分显著更高(P < .001)。
- HASTEDL 与 BLADE 在噪声上无显著差异(P = 0.886)。
- HASTEDL 与 BLADE 在肝、胰、脾、肾图像质量上无显著差异(P = 0.120–0.366)。
诊断标准与关键数值
- 图像质量、噪声、对比度、锐利度、伪影、诊断信心及病灶可检出性采用 1–4 分 Likert 量表(4 分为最佳)。
- HASTEDL 整体图像质量中位评分 4(四分位距 3–4),BLADE 中位评分 4(四分位距 4–4)。
- 病灶数目和最大病灶直径的组间一致性 ICC = 0.972–1.0。
- 采集时间缩短至少 89%(1.5 T 图像)和 86%(3 T 图像)。
临床应用与阅片要点
- HASTEDL 可提供优秀图像质量和良好诊断信心,并大幅缩短采集时间。
- BLADE 在整体及肾上腺图像质量、对比度、锐利度、伪影和诊断信心上评分显著更高(P < .001)。
- HASTEDL 与 BLADE 在噪声、肝胰脾肾图像质量及病灶检出数目和最大径上无显著差异。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
To evaluate the clinical performance of a deep learning-accelerated single-breath-hold half-Fourier acquisition single-shot turbo spin echo (HASTEDL)-sequence for T2-weighted fat-suppressed MRI of the abdomen at 1.5 T and 3 T in comparison to standard T2-weighted fat-suppressed multi-shot turbo spin echo-sequence. A total of 320 patients who underwent a clinically indicated liver MRI at 1.5 T and 3 T between August 2020 and February 2021 were enrolled in this single-center, retrospective study. HASTEDL and standard sequences were assessed regarding overall and organ-based image quality, noise, contrast, sharpness, artifacts, diagnostic confidence, as well as lesion detectability using a Likert scale ranging from 1 to 4 (4 = best). The number of visible lesions of each organ was counted and the largest diameter of the major lesion was measured. HASTEDL showed excellent image quality (median 4, interquartile range 3-4), although BLADE (median 4, interquartile range 4-4) was rated significantly higher for overall and organ-based image quality of the adrenal gland (P < .001), contrast (P < 0.001), sharpness (P < 0.001), artifacts (P < 0.001), as well as diagnostic confidence (P < .001). No significant differences were found concerning noise (P = 0.886), organ-based image quality of the liver, pancreas, spleen, and kidneys (P = 0.120-0.366), number and measured diameter of the detected lesions (ICC = 0.972-1.0). Reduction of the aquisition time (TA) was at least 89% for 1.5 T images and 86% for 3 T images. HASTEDL provided excellent image quality, good diagnostic confidence and lesion detection compared to a standard T2-sequences, allowing an eminent reduction of the acquisition time.
引用
- Herrmann J, Wessling D, Nickel D, et al. Comprehensive Clinical Evaluation of a Deep Learning-Accelerated, Single-Breath-Hold Abdominal HASTE at 1.5 T and 3 T. Acad Radiol. 2022. PMID: 35469719. doi:10.1016/j.acra.2022.03.018
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。