文献速递腹部外周血管介入CTUS

改良CT血管造影标准提高肝移植术后疑似多普勒超声异常患者肝动脉闭塞诊断特异度

Improving the Specificity of CT Angiography for the Diagnosis of Hepatic Artery Occlusion after Liver Transplantation in Suspected Patients with Doppler Ultrasound Abnormalities.
PubMed 文献 · Korean J Radiol · 2022-01-01

本研究纳入134例肝移植术后多普勒超声怀疑肝动脉闭塞的成人受者,比较传统标准(仅吻合口异常)与改良标准(吻合口异常加远端流出道异常)的诊断效能。改良标准将特异度从74.5%提高至93.6%、准确率从82.1%提高至93.3%,敏感度从100%略降至92.5%但无统计学显著差异。

要点

  • 改良CTA标准同时评估吻合口异常与远端流出道异常,特异度93.6%高于传统标准74.5%。
  • 传统标准敏感度100%,改良标准敏感度92.5%,下降无统计学显著差异(p=0.250)。
  • 改良标准准确率93.3%,较传统标准82.1%显著提高(p=0.002)。

核心结论

  • 改良标准同时考虑吻合口和远端流出道异常,提高了CTA对多普勒超声异常疑似患者肝动脉闭塞的诊断效能,尤其提高了特异度。

研究设计与样本

  • 纳入134例成人肝移植受者(88男46女,平均年龄52.7岁),多普勒超声怀疑肝动脉闭塞,参照标准判定40例肝动脉闭塞、94例非肝动脉闭塞,比较传统与改良CTA标准的诊断效能,采用McNemar检验。

关键影像表现/征象

  • 吻合口异常包括吻合口处狭窄≥50%或移植肝侧与受体侧吻合口周围弥漫性狭窄。
  • 远端流出道异常包括远端血管不可见或强化不规则、微弱、不连续。

诊断标准与关键数值

  • 传统标准(仅吻合口异常)诊断肝动脉闭塞:敏感度100%(40/40)、特异度74.5%(70/94)、阳性预测值62.5%(40/64)、阴性预测值100%(70/70)、准确率82.1%(110/134)。
  • 改良标准(吻合口异常加远端流出道异常)诊断肝动脉闭塞:敏感度92.5%(37/40)、特异度93.6%(88/94)、准确率93.3%(125/134)。
  • 改良标准较传统标准特异度提高(p=0.001)、准确率提高(p=0.002),敏感度下降无统计学显著差异(p=0.250)。

临床应用与阅片要点

  • 对多普勒超声怀疑肝动脉闭塞的肝移植受者,CTA阅片时除吻合口异常外应同时评估远端流出道强化是否不可见、不规则、微弱或不连续。
  • 采用改良标准可减少仅凭吻合口异常造成的假阳性,提高诊断特异度。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To investigate whether the diagnostic performance of CT angiography (CTA) could be improved by modifying the conventional criterion (anastomosis site abnormality) to diagnose hepatic artery occlusion (HAO) after liver transplantation (LT) in suspected patients with Doppler ultrasound (US) abnormalities.

MATERIALS AND METHODS: One hundred thirty-four adult LT recipients (88 males and 46 females; mean age, 52.7 years) with suspected HAO on Doppler US (40 HAO and 94 non-HAO according to the reference standards) were included. We evaluated 1) abnormalities in the HA anastomosis, categorized as a cutoff, ≥ 50% stenosis at the anastomotic site, or diffuse stenosis at both graft and recipient sides around the anastomosis, and 2) abnormalities in the distal run-off, including invisibility or irregular, faint, and discontinuous enhancement. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of the conventional (considering anastomosis site abnormalities alone) and modified CTA criteria (abnormalities in both the anastomosis site and distal run-off) for the diagnosis of HAO were calculated and compared using the McNemar test.

RESULTS: By using the conventional criterion to diagnose HAO, the sensitivity, specificity, PPV, NPV, and accuracy were 100% (40/40), 74.5% (70/94), 62.5% (40/64), 100% (70/70), and 82.1% (110/134), respectively. The modified criterion for diagnosing HAO showed significantly increased specificity (93.6%, 88/94) and accuracy (93.3%, 125/134) compared to that with the conventional criterion (p = 0.001 and 0.002, respectively), although the sensitivity (92.5%, 37/40) decreased slightly without statistical significance (p = 0.250).

CONCLUSION: The modified criterion considering abnormalities in both the anastomosis site and distal run-off improved the diagnostic performance of CTA for HAO in suspected patients with Doppler US abnormalities, particularly by increasing the specificity.

来源:Korean J Radiol,查看原文。版权归原作者所有。

引用

  1. Kim JS, Kim DW, Kim KW, et al. Improving the Specificity of CT Angiography for the Diagnosis of Hepatic Artery Occlusion after Liver Transplantation in Suspected Patients with Doppler Ultrasound Abnormalities. Korean J Radiol. 2022. PMID: 34983093. doi:10.3348/kjr.2021.0266

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。