卒中取栓术中的动脉塌陷:人脑与体内模型的临床证据和实验发现
研究报道一例既往植入血流导向装置的动脉塌陷临床病例,并在兔模型(n=3)和人尸体脑模型中评估抽吸取栓时的动脉壁反应。人脑模型中M2段动脉塌陷发生率为98%,完全近端血流停止时全部发生;更大口径抽吸导管、更低脑灌注压及后循环与更高动脉塌陷概率相关。
要点
- 抽吸取栓过程中可发生颅内动脉塌陷,M2段发生率达98%。
- 完全近端血流停止时全部病例出现动脉塌陷。
- 更大口径导管、更低灌注压及后循环与动脉塌陷概率更高相关。
核心结论
- 动脉塌陷确实在抽吸取栓术中发生,且更可能发生于更大口径导管、更低灌注压和更小动脉。
研究设计与样本
- 研究包含一例临床病例报告、兔模型(n=3)及人尸体脑模型实验;人脑模型共进行168次直接抽吸取栓测试,改变导管内径(0.064英寸、0.068英寸、0.070英寸)、脑灌注压(平均约60和90 mm Hg)及前循环与后循环。
关键影像表现/征象
- 临床病例中既往植入血流导向装置处出现动脉塌陷。
- 兔模型经透视和血管内光学相干断层扫描评估真空效应。
- 人脑模型中经直接腔内观察记录并分级动脉壁反应。
诊断标准与关键数值
- 人脑模型中M2段动脉塌陷发生率为98%。
- 完全近端血流停止时全部病例发生动脉塌陷。
- 更大口径抽吸导管、更低脑灌注压及后循环(相较前循环)与更高动脉塌陷概率相关。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Aspiration thrombectomy has become a preferred approach to recanalize large-vessel occlusion in stroke with a growing trend toward using larger-bore catheters and stronger vacuum pumps. However, the mechanical response of the delicate cerebral arteries to aspiration force has not been evaluated. Here, we provide preclinical and clinical evidence of intracranial arterial collapse in aspiration thrombectomy.
MATERIALS AND METHODS: We presented a clinical case of arterial collapse with previously implanted flow diverters. We then evaluated the effect of vacuum with conventional aspiration catheters (with and without stent retrievers) in a rabbit model (n = 3) using fluoroscopy and intravascular optical coherence tomography. Then, in a validated human cadaveric brain model, we conducted 168 tests of direct aspiration thrombectomy following an experimental design modifying the catheter inner diameter (0.064 inch, 0.068 inch, and 0.070 inch), cerebral perfusion pressures (mean around 60 and 90 mm Hg), and anterior-versus-posterior circulation. Arterial wall response was recorded and graded via direct transluminal observation.
RESULTS: Arterial collapse was observed in both the patient and preclinical experimental models. In the human brain model, arterial collapse was observed in 98% of cases in the M2 and in all the cases with complete proximal flow arrest. A larger bore size of the aspiration catheter, a lower cerebral perfusion pressure, and the posterior circulation in comparison with the anterior circulation were associated with a higher probability of arterial collapse.
CONCLUSIONS: Arterial collapse does occur during aspiration thrombectomy and is more likely to happen with larger catheters, lower perfusion pressure, and smaller arteries.
引用
- Liu Y, Gebrezgiabhier D, Zheng Y, et al. Arterial Collapse during Thrombectomy for Stroke: Clinical Evidence and Experimental Findings in Human Brains and In Vivo Models. AJNR Am J Neuroradiol. 2022. PMID: 35027348. doi:10.3174/ajnr.A7389
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。