压缩感知MP2RAGE作为3T神经成像中MPRAGE的替代方案

The Compressed Sensing MP2RAGE as a Surrogate to the MPRAGE for Neuroimaging at 3 T.
PubMed 文献 · Invest Radiol · 2022-01-14

在13名健康志愿者中于3T临床扫描仪上比较压缩感知MP2RAGE与标准MPRAGE序列,两者空间分辨率均为0.8毫米各向同性、扫描时间均为6分21秒。CS-MP2RAGE提供的形态学图像质量与MPRAGE相同且对比度更高,脑结构分割的个体内变异性相似,并同时获得高分辨率T1图,白质、灰质及多个深部灰质核团的T1值与文献一致且变异性低于1%。

要点

  • CS-MP2RAGE与MPRAGE空间分辨率(0.8毫米各向同性)和扫描时间(6分21秒)相同
  • CS-MP2RAGE形态学图像质量与MPRAGE相同,对比度更高
  • CS-MP2RAGE分割的个体内变异性与MPRAGE相似,并同时提供高分辨率T1图

核心结论

  • CS-MP2RAGE可在未来方案中用于快速获得三维形态学图像和定量T1图,同时保持体积测量和弛豫时间的高重复性。

研究设计与样本

  • 研究纳入13名健康志愿者,在3T临床扫描仪上间隔数天扫描3或4次,比较CS-MP2RAGE与MPRAGE两种序列;两者空间分辨率均为0.8毫米各向同性,扫描时间均为6分21秒。

关键影像表现/征象

  • CS-MP2RAGE获得的形态学图像质量与标准MPRAGE图像相同,且对比度更高。
  • CS-MP2RAGE分割所得的个体内变异性与MPRAGE分割相似。
  • CS-MP2RAGE可同时获得高分辨率T1图。

诊断标准与关键数值

  • CS-MP2RAGE输出(包括2幅回波图像、MP2RAGE图像和T1图)的重建时间为3分33秒,使用在Gadgetron框架中实现的开源内部算法。
  • 白质、灰质及多个深部灰质核团的T1值与文献一致,变异性低于1%。

临床应用与阅片要点

  • CS-MP2RAGE可在与MPRAGE相同的扫描时间(6分21秒)和空间分辨率(0.8毫米各向同性)下同时提供形态学图像和定量T1图。
  • 基于CAT12和VolBrain的脑结构体积测量在CS-MP2RAGE与MPRAGE之间个体内变异性相似。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The magnetization-prepared 2 rapid acquisition gradient echo (MP2RAGE) sequence provides quantitative T1 maps in addition to high-contrast morphological images. Advanced acceleration techniques such as compressed sensing (CS) allow its acquisition time to be compatible with clinical applications. To consider its routine use in future neuroimaging protocols, the repeatability of the segmented brain structures was evaluated and compared with the standard morphological sequence (magnetization-prepared rapid gradient echo [MPRAGE]). The repeatability of the T1 measurements was also assessed.

MATERIALS AND METHODS: Thirteen healthy volunteers were scanned either 3 or 4 times at several days of interval, on a 3 T clinical scanner, with the 2 sequences (CS-MP2RAGE and MPRAGE), set with the same spatial resolution (0.8-mm isotropic) and scan duration (6 minutes 21 seconds). The reconstruction time of the CS-MP2RAGE outputs (including the 2 echo images, the MP2RAGE image, and the T1 map) was 3 minutes 33 seconds, using an open-source in-house algorithm implemented in the Gadgetron framework.Both precision and variability of volume measurements obtained from CAT12 and VolBrain were assessed. The T1 accuracy and repeatability were measured on phantoms and on humans and were compared with literature.Volumes obtained from the CS-MP2RAGE and the MPRAGE images were compared using Student t tests (P < 0.05 was considered significant).

RESULTS: The CS-MP2RAGE acquisition provided morphological images of the same quality and higher contrasts than the standard MPRAGE images. Similar intravolunteer variabilities were obtained with the CS-MP2RAGE and the MPRAGE segmentations. In addition, high-resolution T1 maps were obtained from the CS-MP2RAGE. T1 times of white and gray matters and several deep gray nuclei are consistent with the literature and show very low variability (<1%).

CONCLUSIONS: The CS-MP2RAGE can be used in future protocols to rapidly obtain morphological images and quantitative T1 maps in 3-dimensions while maintaining high repeatability in volumetry and relaxation times.