利用双翻转角估算DCE-MRI中的对比剂浓度
本研究在C57BL/6J小鼠中采用超短回波时间与两个翻转角的三段式DCE-MRI扫描方案,无需预对比T1测量即可估算血浆钆浓度;以电感耦合等离子体质谱为参考,双翻转角法估算的钆浓度与质谱结果强相关(r=0.85),且GL261脑肿瘤的对比剂动力学模型参数与使用预对比T1测量时无显著差异。
要点
- 双翻转角法估算的钆浓度与质谱参考值强相关(r=0.85),平均百分比误差为15%±44%。
- 使用文献常数T1(2.1秒)时相关性弱(r=0.52),误差较大(33%±24%)。
- GL261脑肿瘤的对比剂动力学模型参数在双翻转角法与预对比T1测量法之间无显著差异。
核心结论
- 本研究证实了在DCE-MRI扫描中使用双翻转角方法估算血浆钆浓度的可行性,无需额外扫描进行预对比T1测量。
研究设计与样本
- 动物实验研究,使用C57BL/6J小鼠,采用超短回波时间与两个翻转角的三段式单次扫描DCE-MRI方案。
诊断标准与关键数值
- 以电感耦合等离子体质谱测定的血样钆浓度为参考标准;双翻转角法估算的钆浓度与质谱值强相关(r=0.85,平均百分比误差±标准差为15%±44%),预对比T1测量法为r=0.70(16%±56%),文献常数T1(2.1秒)法为r=0.52(33%±24%)。
临床应用与阅片要点
- 在DCE-MRI中采用双翻转角方案可在不进行预对比T1测量的情况下估算血浆钆浓度,可能减少扫描步骤。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: The aim of this study was to investigate the feasibility of using 2 flip angles (FAs) with an ultrashort echo time during dynamic contrast-enhanced (DCE)-magnetic resonance imaging (MRI) for estimation of plasma gadolinium (Gd) concentration without using a precontrast longitudinal relaxation time T1 (T10) measurement.
METHODS: T1-weighted DCE-MRI experiments were carried out with C57BL/6J mice using the scan protocol with 2 FAs over 3 sequential segments during 1 scan. The data with 2 FAs were used to estimate T10 (T1T) during conversion of a time-intensity curve to the time-concentration curve. Three dosages of gadolinium-based contrast agent were used to achieve a wide range of variability in Gd concentrations when measured at 10 minutes postinjection: 0.05 mmol/kg (n = 6), 0.1 mmol/kg (n = 11), and 0.15 mmol/kg (n = 7). For comparison, the signal-to-concentration conversion was also conducted using the T10 measured from the precontrast scan (T1M) as well as a constant T10 (2.1 seconds) from the literature (T1C). The Gd concentrations ([Gd]) estimated using DCE-MRI data for the time of retro-orbital blood collection ([Gd]T1T, [Gd]T1M, and [Gd]T1C, respectively) were compared against the [Gd] of the blood samples measured by inductively coupled plasma mass spectrometry ([Gd]MS). In addition, contrast kinetic model analysis was conducted on mice with GL261 brain tumors (n = 5) using the 3 different methods for T10.
RESULTS: T1T strongly correlated with T1M (r = 0.81). [Gd]T1M and [Gd]T1T were significantly different from [Gd]T1C. [Gd]T1M and [Gd]T1T were in good agreement with [Gd]MS with strong correlations (mean percentage error ± standard deviation) of r = 0.70 (16% ± 56%) and r = 0.85 (15% ± 44%), respectively. In contrast, [Gd]T1C had a weak correlation of r = 0.52 with larger errors of 33% ± 24%. The contrast kinetic model parameters of GL261 brain tumors using T1T were not significantly different from those using T1M.
CONCLUSIONS: This study substantiates the feasibility of using the 2-FA approach during DCE-MRI scan to estimate [Gd] in the plasma without using an extra scan to perform precontrast T1 measurements.