经会阴超声容积中最小裂孔尺寸切面的自动识别与分割
本研究在有症状盆腔器官脱垂的116例女性经会阴超声容积中手动选择最小裂孔尺寸切面并分割泌尿生殖裂孔,训练两个深度学习算法实现自动选择与分割,在30个容积测试集上自动与手动选择切面平均绝对距离0.20 cm,分割重叠度均大于0.85,测量组内相关系数泌尿生殖裂孔面积0.94、前后径0.92、冠状径0.82,结论认为算法可可靠自动选择切面并分割裂孔。
要点
- 自动选择最小裂孔尺寸切面与手动位置平均绝对距离0.20 cm。
- 自动与手动泌尿生殖裂孔分割重叠度均大于0.85。
- 自动与手动测量组内相关系数面积0.94、前后径0.92、冠状径0.82。
核心结论
- 深度学习算法可在有症状盆腔器官脱垂女性的经会阴超声容积中可靠地自动选择最小裂孔尺寸切面并分割泌尿生殖裂孔。
研究设计与样本
- 在116例有症状盆腔器官脱垂女性的经会阴超声容积中手动选择最小裂孔尺寸切面并分割泌尿生殖裂孔,用于训练两个深度学习算法;在30个经会阴超声容积测试集上评估自动与手动选择切面的平均绝对距离、分割重叠度和测量组内相关系数。
诊断标准与关键数值
- 自动与手动选择最小裂孔尺寸切面平均绝对距离0.20 cm。
- 自动与手动泌尿生殖裂孔分割所有重叠度值均大于0.85。
- 自动与手动泌尿生殖裂孔面积测量组内相关系数0.94(95%CI 0.87-0.97)。
- 自动与手动泌尿生殖裂孔前后径测量组内相关系数0.92(95%CI 0.78-0.97)。
- 自动与手动泌尿生殖裂孔冠状径测量组内相关系数0.82(95%CI 0.66-0.91)。
临床应用与阅片要点
- 算法可整合入经会阴超声设备软件,以减少临床分析时间并简化经会阴超声数据的研究与临床用途检查。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To develop and validate a tool for automatic selection of the slice of minimal hiatal dimensions (SMHD) and segmentation of the urogenital hiatus (UH) in transperineal ultrasound (TPUS) volumes.
METHODS: Manual selection of the SMHD and segmentation of the UH was performed in TPUS volumes of 116 women with symptomatic pelvic organ prolapse (POP). These data were used to train two deep-learning algorithms. The first algorithm was trained to provide an estimation of the position of the SMHD. Based on this estimation, a slice was selected and fed into the second algorithm, which performed automatic segmentation of the UH. From this segmentation, measurements of the UH area (UHA), anteroposterior diameter (APD) and coronal diameter (CD) were computed automatically. The mean absolute distance between manually and automatically selected SMHD, the overlap (dice similarity index (DSI)) between manual and automatic UH segmentation and the intraclass correlation coefficient (ICC) between manual and automatic UH measurements were assessed on a test set of 30 TPUS volumes.
RESULTS: The mean absolute distance between manually and automatically selected SMHD was 0.20 cm. All DSI values between manual and automatic UH segmentations were above 0.85. The ICC values between manual and automatic UH measurements were 0.94 (95% CI, 0.87-0.97) for UHA, 0.92 (95% CI, 0.78-0.97) for APD and 0.82 (95% CI, 0.66-0.91) for CD, demonstrating excellent agreement.
CONCLUSIONS: Our deep-learning algorithms allowed reliable automatic selection of the SMHD and UH segmentation in TPUS volumes of women with symptomatic POP. These algorithms can be implemented in the software of TPUS machines, thus reducing clinical analysis time and simplifying the examination of TPUS data for research and clinical purposes. © 2021 The Authors. Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of International Society of Ultrasound in Obstetrics and Gynecology.
引用
- van den Noort F, Manzini C, van der Vaart CH, et al. Automatic identification and segmentation of slice of minimal hiatal dimensions in transperineal ultrasound volumes. Ultrasound Obstet Gynecol. 2022. PMID: 34767663. doi:10.1002/uog.24810
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。