用于肝脏脂肪定量的衰减系数估计的预发布算法与置信图
该单中心研究在30名参与者中比较了Philips Epiq Elite实验软件与Canon Aplio i800已发布软件测量肝脏衰减系数的一致性,结果显示两系统高度一致(CCC 0.792),但基于置信图确定的最佳测量深度可能不同。
要点
- 两系统衰减系数测量高度一致,CCC为0.792(95%CI 0.666-0.918)。
- 基于Philips置信图确定的最佳测量深度为距肝包膜3.5至4.0 cm,可能与Canon系统显著不同。
- 测量感兴趣区放置应基于置信图,且两系统间深度和大小需保持一致。
核心结论
- 两系统衰减系数估计高度一致,但基于置信图确定测量框放置的质量评估可能在两系统间不同。
研究设计与样本
- 单中心IRB批准、HIPAA合规研究,共30名参与者,使用Philips Epiq Elite实验软件和Canon Aplio i800已发布软件,每系统取5次衰减系数测量并计算中位数和IQR/M,感兴趣区放置基于置信图且两系统深度和大小一致。
诊断标准与关键数值
- Philips系统衰减系数范围为0.45至1.0 dB/cm/MHz,Canon系统为0.30至0.96 dB/cm/MHz;Lin一致性相关系数CCC为0.792(95%CI 0.666-0.918),Pearson r为0.839,偏差校正因子Cb为0.944,平均差异为0.03(-0.101;0.162);基于Philips置信图确定的最佳测量深度为距肝包膜3.5至4.0 cm。
临床应用与阅片要点
- 基于Philips置信图可确定测量感兴趣区的最佳位置为距肝包膜3.5至4.0 cm,但该深度可能与Canon系统显著不同,测量时两系统需保持深度和大小一致。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To compare the estimates of attenuation coefficient (AC) for liver fat quantification between 2 Ultrasound systems and to evaluate the quality measure of a pre-released software.
METHODS: AC were obtained in 30 participants in this single-center IRB-approved, HIPAA compliant study. Images were obtained on the Philips Epiq Elite system using experimental software and the Canon Medical Systems Aplio i800 with released software. Five AC measurements were taken and the median and IQR/M were calculated. Region of interest placement was based on a confidence map. ROI was at the same depth and size for each system. The concordance was estimated using the Lin's concordance correlation coefficient (CCC), the r Pearson's correlation coefficient, the bias-correction factor (Cb), and the Bland-Altman method.
RESULTS: The ACs varied from 0.45 to 1.0 dB/cm/MHz for the Philips system and 0.30 to 0.96 dB/cm/MHz for the Canon system. The CCC (95% CI) was 0.792 (0.666-0.918), Pearson's r was 0.839 with Cb of 0.944, and the mean difference was 0.03 (-0.101; 0.162) suggesting the 2 methods are considered to be in agreement. Based on a Philips confidence map to determine the best location for performing the measurements, a depth of 3.5 to 4.0 cm from the liver capsule was determined, which might be significantly different than that of the Canon system.
CONCLUSIONS: Estimation of the AC of the 2 systems showed a high agreement, that is, a similar trend. Assessment of the placement of the measurement box based on the quality of the measurement might be different between the 2 systems.
引用
- Barr RG, Cestone A, De Silvestri A, et al. A Pre-Release Algorithm With a Confidence Map for Estimating the Attenuation Coefficient for Liver Fat Quantification. J Ultrasound Med. 2021. PMID: 34730847. doi:10.1002/jum.15870
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。