以氧15水PET为参考验证双源CT绝对心肌血流量定量准确性的临床研究
本研究前瞻性纳入34例已知或疑似冠心病受试者,随机分为试验组(n=17)确定校正方程和验证组(n=17)验证校正后的CT心肌血流量(MBFCT-corrected)与氧15水PET心肌血流量(MBFPET)的一致性。试验组中MBFCT低于MBFPET(1.24 mL/min/g ± 0.28 vs 2.51 mL/min/g ± 0.89,P < .001),基于广义Renkin-Crone模型得出校正方程;验证组中MBFCT-corrected为2.66 mL/min/g ± 1.93,MBFPET为2.68 mL/min/g ± 1.87,组内相关系数=0.93(95% CI: 0.87, 0.96),测量偏差为-0.02 mL/min/g ± 0.74。结论:基于心肌对比剂动力学的校正方法成功校正了CT对MBF的低估。
要点
- 动态CT灌注测得的心肌血流量(MBFCT)较氧15水PET参考值偏低。
- 基于广义Renkin-Crone模型的校正方程可校正CT对MBF的低估。
- 校正后MBFCT与MBFPET一致性优秀,组内相关系数=0.93。
核心结论
- 基于心肌对比剂动力学的校正方法成功校正了CT对心肌血流量的低估。
研究设计与样本
- 前瞻性研究(编号2466),34例已知或疑似冠心病受试者(平均年龄70岁±8,27例男性),2014年1月至2018年12月间接受动态负荷CT灌注和负荷氧15水PET,随机分为试验组(n=17)和验证组(n=17)。
诊断标准与关键数值
- 试验组中MBFCT低于MBFPET(1.24 mL/min/g ± 0.28 vs 2.51 mL/min/g ± 0.89,P < .001);校正关系为MBFCT = MBFCT-corrected × {1-exp[-(0.11 × MBFCT-corrected + 1.54)/MBFCT-corrected]};验证组中MBFCT-corrected为2.66 mL/min/g ± 1.93,MBFPET为2.68 mL/min/g ± 1.87,组内相关系数=0.93(95% CI: 0.87, 0.96),测量偏差为-0.02 mL/min/g ± 0.74。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To determine the fitting equation that can correct for the underestimation of myocardial blood flow (MBF) measurement by using dynamic CT perfusion (CTP) with dual-source CT (MBFCT), using MBF with oxygen 15-labeled water (15O-water) PET (MBFPET) as a reference, and to determine the accuracy of corrected MBFCT (MBFCT-corrected) compared with MBFPET in a separate set of participants.
MATERIALS AND METHODS: In this prospective study (reference no. 2466), 34 participants (mean age, 70 years ± 8 [standard deviation]; 27 men) known or suspected to have coronary artery disease underwent dynamic stress CTP and stress 15O-water PET between January 2014 and December 2018. The participants were randomly assigned to either a pilot group (n = 17), to determine the fitting equation on the basis of the generalized Renkin-Crone model that can explain the relation between MBFCT and MBFPET, or to a validation group (n = 17), to validate MBFCT-corrected compared with MBFPET. The agreement between MBFCT-corrected and MBFPET was evaluated by intraclass correlation and Bland-Altman analysis.
RESULTS: In the pilot group, MBFCT was lower than MBFPET (1.24 mL/min/g ± 0.28 vs 2.51 mL/min/g ± 0.89, P < .001) at the segment level. The relationship between MBFCT and MBFCT-corrected was represented as MBFCT = MBFCT-corrected × {1-exp[-(0.11 × MBFCT-corrected + 1.54)/MBFCT-corrected]}. In the validation group, MBFCT-corrected was 2.66 mL/min/g ± 1.93, and MBFPET was 2.68 mL/min/g ± 1.87 at the vessel level. MBFCT-corrected showed an excellent agreement with MBFPET (intraclass correlation coefficient = 0.93 [95% CI: 0.87, 0.96]). The measurement bias of MBFCT-corrected and MBFPET was -0.02 mL/min/g ± 0.74.
CONCLUSION: Underestimation of MBF by CT was successfully corrected with a correction method that was based on contrast kinetics in the myocardium.Keywords: CT, CT-Perfusion, PET, Cardiac, Heart Supplemental material is available for this article. © RSNA, 2021.