常规与光子计数CT评估冠状动脉斑块及支架:高分辨率光子计数CT的优势
在冠状动脉体模模型中比较EID CT、PCCT与高分辨率PCCT对斑块和支架伪影的显示能力。HR-PCCT的径线测量更准确,支架所致狭窄百分比变化最小,观察者认为其斑块显示更清晰且受支架伪影影响最小,读者间一致性高(ICC=0.85)。
要点
- HR-PCCT在无支架和有支架体模中的径线测量均较EID和PCCT更准确
- 加入支架后HR-PCCT狭窄百分比平均仅变化-5.5%,EID为+90.5%,PCCT为+313%
- 观察者认为HR-PCCT斑块显示更清晰且受支架伪影影响最小,读者一致性ICC=0.85
核心结论
- 尽管噪声增加,HR-PCCT较EID或PCCT重建能更好地显示冠状动脉斑块并减少支架伪影。
研究设计与样本
- 体模研究,使用带EID和PCCT子系统的研究型扫描仪对含模拟不同斑块成分圆柱探针的冠状动脉体模成像,分别在有和无冠状动脉支架条件下扫描,图像采用临床心脏内核及额外HR-PCCT内核重建,由专家读者进行定性比较,采用线性混合效应模型比较直径结果,Shrout-Fleiss组内相关系数评估读者一致性。
关键影像表现/征象
- HR-PCCT无支架体模径线测量偏差4.4%±1.1,有支架体模-9.4%±4.6,较EID和PCCT更接近物理尺寸
- EID无支架体模径线偏差15.5%±4.0,有支架体模-19.5%±5.8
- PCCT无支架体模径线偏差19.4%±2.5,有支架体模-18.3%±4.4
- 加入支架后狭窄百分比平均变化:HR-PCCT -5.5%,EID +90.5%,PCCT +313%
- 观察者认为HR-PCCT图像斑块显示更清晰且受支架伪影影响最小,读者一致性高(ICC=0.85)
诊断标准与关键数值
- 方差分析显示无支架和有支架条件下各图像组直径测量值不同(P < .001)
- 读者一致性采用组内相关系数评估,ICC=0.85
临床应用与阅片要点
- 在体模模型中,HR-PCCT对斑块径线和狭窄程度的测量较EID和PCCT更准确,受支架伪影影响更小
- 观察者主观评价HR-PCCT斑块显示更清晰且受支架伪影影响最小,但图像噪声增加
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To compare the performance of energy-integrating detector (EID) CT, photon-counting detector CT (PCCT), and high-resolution PCCT (HR-PCCT) for the visualization of coronary plaques and reduction of stent artifacts in a phantom model.
MATERIALS AND METHODS: An investigational scanner with EID and PCCT subsystems was used to image a coronary artery phantom containing cylindrical probes simulating different plaque compositions. The phantom was imaged with and without coronary stents using both subsystems. Images were reconstructed with a clinical cardiac kernel and an additional HR-PCCT kernel. Regions of interest were drawn around probes and evaluated for in-plane diameter and a qualitative comparison by expert readers. A linear mixed-effects model was used to compare the diameter results, and a Shrout-Fleiss intraclass correlation coefficient was used to assess consistency in the reader study.
RESULTS: Comparing in-plane diameter to the physical dimension for nonstented and stented phantoms, measurements of the HR-PCCT images were more accurate (nonstented: 4.4% ± 1.1 [standard deviation], stented: -9.4% ± 4.6) than EID (nonstented: 15.5% ± 4.0, stented: -19.5% ± 5.8) and PCCT (nonstented: 19.4% ± 2.5, stented: -18.3% ± 4.4). Our analysis of variance found diameter measurements to be different across image groups for both nonstented and stented cases (P < .001). HR-PCCT showed less change on average in percent stenosis due to the addition of a stent (-5.5%) than either EID (+90.5%) or PCCT (+313%). For both nonstented and stented phantoms, observers rated the HR-PCCT images as having higher plaque conspicuity and as being the image type that was least impacted by stent artifacts, with a high level of agreement (interclass correlation coefficient = 0.85).
CONCLUSION: Despite increased noise, HR-PCCT images were able to better visualize coronary plaques and reduce stent artifacts compared with EID or PCCT reconstructions.Keywords: CT-Spectral Imaging (Dual Energy), Phantom Studies, Cardiac, Physics, Technology Assessment© RSNA, 2021.