自由呼吸与屏气心脏CTA在TAVR术前瓣环测量中的对比评估
本研究纳入60例连续TAVR候选者行自由呼吸宽探测器心脏CTA,并与60例既往屏气采集患者对照,比较瓣环测量准确性及临床结局。结果显示自由呼吸组图像质量相当,CT建议瓣膜尺寸与实际植入尺寸一致率87%(κ=0.79),与屏气组82%(κ=0.78)可比,操作均成功且30天死亡率或不良事件未增加。
要点
- 自由呼吸心脏CTA可用于TAVR术前瓣环测量,图像质量与屏气采集相当。
- CT建议瓣膜尺寸与实际植入尺寸一致率自由呼吸组87%、屏气组82%。
- 无法配合屏气的老年患者不应被排除在TAVR术前CT评估之外。
核心结论
- 自由呼吸心脏CTA能够准确测量主动脉瓣环,不损害图像质量或TAVR术后患者结局。
研究设计与样本
- 前瞻性纳入60例连续TAVR候选者行自由呼吸宽探测器心脏CTA,并与60例既往屏气采集对照患者比较。
- 以实际植入经导管心脏瓣膜尺寸作为参考标准,评估CT建议瓣膜尺寸与实际植入尺寸的一致性。
- 收集操作成功率和30天结局,并按主观和客观标准评估瓣环图像质量。
关键影像表现/征象
- 所有CT采集均提供足够图像质量以进行精确瓣环测量。
- 自由呼吸组平均衰减值(p < 0.001)和图像噪声(p = 0.01)较高,但图像质量主观评价相当(p = 0.36)。
诊断标准与关键数值
- CT建议瓣膜尺寸与实际植入尺寸一致率:自由呼吸组87%(κ = 0.79,95%CI 0.566, 0.908),屏气组82%(κ = 0.78,95%CI 0.634, 0.904)。
临床应用与阅片要点
- 对于因呼吸困难、不适或听力损失无法配合屏气的老年患者,可采用自由呼吸采集行TAVR术前CT评估。
- 自由呼吸采集图像噪声和衰减值较高,但主观图像质量与屏气采集相当,不影响瓣环测量和瓣膜尺寸选择。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To assess feasibility and accuracy of aortic annulus measurements using cardiac computed tomography angiography (CTA) performed during free-breathing prior to transcatheter aortic valve replacement (TAVR).
MATERIALS AND METHODS: Sixty consecutive TAVR candidates underwent free-breathing wide-detector cardiac CTA, followed by a percutaneous valve replacement. For each, a theoretical valve size was suggested through CT measurements of the annulus, then compared to the size of the actual implanted transcatheter heart valve (THV). The procedural success and the 30-day outcomes were collected. Image quality of the annulus was also studied according to subjective and objective criteria. Data of a control group of 60 patients previously evaluated on breath-holding were also evaluated.
RESULTS: A total of 120 patients (mean age, 83 years ± 7, 60 men) were evaluated. All CT acquisitions provided sufficient image quality allowing precise annulus measurements. Mean attenuation (p < 0.001) and image noise (p = 0.01) were higher in the free-breathing group, while image quality was comparable (p = 0.36). The agreement rate between CT-suggested valve size and THV implanted size was comparable, estimated at 87% (κ = 0.79, 95%CI 0.566, 0.908) on free-breathing vs. 82% (κ = 0.78, 95%CI 0.634, 0.904) on breath-holding. The procedure was successful for all patients without increase in 30-day mortality or adverse events.
CONCLUSIONS: Free-breathing cardiac CTA allows accurate aortic annulus measurements without compromising image quality or patients' outcome after TAVR. Elderly patients experiencing dyspnea, discomfort, or hearing loss that could prevent proper breath-holding should not be excluded from CT prior to TAVR.
CLINICAL RELEVANCE STATEMENT: To decrease elderly patients' discomfort, MDCT evaluation prior to transcatheter aortic valve replacement (TAVR) may be performed on quiet breathing with no significant impact on the outcome.
KEY POINTS: • Adhering to CT breathing commands can be challenging for patients with dyspnea, hearing impairment, agitation, or pulmonary diseases. • Free-breathing cardiac CT may be an alternative to breath-holding for patients unable to follow the breathing commands. • Wide-detector CT acquisition on free-breathing does not impair annulus measurements and prosthesis sizing in patients scheduled for TAVR.