在维持结节检出能力的前提下降低辐射剂量的可行性:采用不同重建方法的胸部CT体模研究中银与铜X射线能谱调制滤过器的比较

Capability for dose reduction while maintaining nodule detection: Comparison of silver and copper X-ray spectrum modulation filters for chest CT using a phantom study with different reconstruction methods.
PubMed 文献 · Eur J Radiol · 2023-07-10

该体模研究用320排CT分别加银滤过器(0.6、1.6、2.5 mGy)和铜滤过器(0.6、1.6、2.5、9.6 mGy)扫描胸部体模,并以FBP、混合迭代重建和深度学习重建(DLR)重建。各剂量下银滤过器的SNR、CNR和AUC均显著高于铜滤过器(p<0.05),DLR的上述数值也显著高于其他重建方法(p<0.05)。结论认为银滤过器和DLR可在各所用剂量下显著改善图像质量和结节检出能力。

要点

  • 相同剂量下银滤过器的SNR、CNR和结节检出AUC均显著高于铜滤过器(p<0.05)。
  • 各剂量及任一种滤过器下,DLR的SNR、CNR和AUC均显著高于其他重建方法(p<0.05)。
  • 体模研究提示银滤过器联合DLR有利于在维持结节检出能力的同时降低辐射剂量。

核心结论

  • 银滤过器和DLR在各所用辐射剂量下均可较铜滤过器和其他重建方法显著改善图像质量和结节检出能力。

研究设计与样本

  • 胸部CT体模研究,采用320排CT,银滤过器剂量为0.6、1.6和2.5 mGy,铜滤过器剂量为0.6、1.6、2.5和9.6 mGy,分别以FBP、混合迭代重建和DLR重建。

关键影像表现/征象

  • 以5点视觉评分评估任意结节的检出概率,并据此进行ROC分析比较结节检出能力。
  • 各剂量下银滤过器的SNR、CNR和AUC均显著高于或大于铜滤过器(p<0.05)。
  • 各剂量及任一种滤过器下,DLR的SNR、CNR和AUC均显著高于或大于其他重建方法(p<0.05)。

诊断标准与关键数值

  • 结节检出能力以5点视觉评分后的ROC分析所得AUC评估。
  • 各剂量下银滤过器的AUC显著大于铜滤过器(p<0.05)。
  • 各剂量及任一种滤过器下DLR的AUC显著大于其他重建方法(p<0.05)。

临床应用与阅片要点

  • 体模研究显示,相同剂量下银滤过器的SNR、CNR和结节检出AUC均高于铜滤过器(p<0.05)。
  • 各剂量及任一种滤过器下,DLR的SNR、CNR和AUC均高于其他重建方法(p<0.05)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the capability of CTs obtained with a silver or copper x-ray beam spectral modulation filter (Ag filter and Cu filter) and reconstructed with FBP, hybrid-type IR and deep learning reconstruction (DLR) for radiation dose reduction for lung nodule detection using a chest phantom study.

MATERIALS AND METHODS: A chest CT phantom was scanned with a 320-detector row CT with Ag filter at 0.6, 1.6 and 2.5 mGy and Cu filters at 0.6, 1.6, 2.5 and 9.6 mGy, and reconstructed with the aforementioned methods. To compare image quality of all the CT data, SNRs and CNRs for any nodule were calculated for all protocols. To compare nodule detection capability among all protocols, the probability of detection of any nodule was assessed with a 5-point visual scoring system. Then, ROC analyses were performed to compare nodule detection capability of Ag and Cu filters for each radiation dose data with the same method and of the three methods for any radiation dose data and obtained with either filter.

RESULTS: At any of the doses, SNR, CNR and area under the curve for the Ag filter were significantly higher or larger than those for the Cu filter (p < 0.05). Moreover, with DLR, those values were significantly higher or larger than all the others for CTs obtained with any of the radiation doses and either filter (p < 0.05).

CONCLUSION: The Ag filter and DLR can significantly improve image quality and nodule detection capability compared with the Cu filter and other reconstruction methods at each of radiation doses used.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Oshima Y, Ohno Y, Takenaka D, et al. Capability for dose reduction while maintaining nodule detection: Comparison of silver and copper X-ray spectrum modulation filters for chest CT using a phantom study with different reconstruction methods. Eur J Radiol. 2023. PMID: 37454556. doi:10.1016/j.ejrad.2023.110969