在颅内动脉内皮化硅胶模型中评估模拟导管使用后的血管损伤
本研究构建了颈内动脉和大脑中动脉的复杂内皮化硅胶模型,用神经血管导管模拟操作后通过定量图像分析和视觉评分评估血管损伤。结果显示3次通过造成的损伤多于1次通过,损伤随器械数量和尺寸增加,提示该模型可用于早期体外评估导管相关血管损伤。
要点
- 内皮化硅胶模型可模拟神经血管导管操作并观察血管壁损伤。
- 3次通过造成的血管损伤多于1次通过,损伤随器械数量和尺寸增加。
- 透明硅胶允许无透视下模拟操作,损伤可被成像和定量分析。
核心结论
- 复杂硅胶神经血管模型可内皮化并用于体外评估和比较神经血管导管使用所致的血管损伤。
研究设计与样本
- 体外研究:硅胶制作颈内动脉和大脑中动脉模型,经灭菌、纤连蛋白包被、生物反应器培养后用人脐静脉内皮细胞内皮化;维持流动3天和7天后评估细胞衬里;随后制作24条硅胶血管,用导丝、微导管和/或抽吸导管单次(1-pass)或三次(3-pass)通过,固定后行定量图像分析和视觉评分评估损伤。
关键影像表现/征象
- 透明硅胶允许无透视下模拟导管使用,血管壁损伤可被观察、成像和表征。
- 3次通过的血管损伤多于1次通过。
- 损伤随器械数量和尺寸增加。
临床应用与阅片要点
- 作者认为该模型支持用于神经血管导管相关血管损伤的早期体外评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: Neurothrombectomy catheters can disrupt or injure the vessel wall. This potential injury is often studied in animal or cadaver models, but prior work suggests that endothelialized silicone models may be an option for early in vitro assessment. The purpose of this work was to create a complex, clinically-relevant endothelialized neurovascular silicone model, and to determine the utility of the model for evaluating vessel injury due to catheter simulated use.
METHODS: Models of the ICA and MCA were fabricated out of silicone, sterilized, coated with fibronectin, placed in bioreactors, and endothelialized with HUVECs. These silicone vessels were maintained under flow for 3 and 7 days, and cellular linings were assessed. Subsequently, 24 silicone vessels were created and treated with neurovascular catheters. Vessels were accessed with a guidewire, microcatheter, and/or aspiration catheter, either once (1-pass) or three times (3-pass). Vessels were then fixed, and injury was evaluated through quantitative image analysis and a visual scoring system.
RESULTS: Complex silicone models were successfully endothelialized and maintained with consistent cell linings. The transparent silicone permitted catheter simulated use without fluoroscopy, and injury to the vessel wall was observed and successfully imaged and characterized. Vessels subjected to 3-passes exhibited more injury than 1-pass, and injury increased with the number and size of devices. These results illustrated expected trends and support use of these models for early assessment of vessel injury.
CONCLUSION: Complex silicone neurovascular models can be endothelialized and used in vitro to assess and compare injury due to the use of neurovascular catheters.
引用
- McCulloch A, Yang B, Frenklakh S, et al. Evaluation of vessel injury after simulated catheter use in an endothelialized silicone model of the intracranial arteries. Neuroradiology. 2023. PMID: 37458786. doi:10.1007/s00234-023-03197-8
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。