深度学习重建联合超分辨率处理加速3T乳腺MRI扩散加权成像的前瞻性对比研究
前瞻性单中心研究纳入65例受试者,比较深度学习加速DWI(DWI DL)与标准DWI(DWI STD)的采集时间和图像质量。DWI DL将采集时间从5:02分钟缩短至2:44分钟,虽乳腺组织SNR较低且伪影评分较高,但病变对比度更高、病变显著性更优,ADC值无显著差异,整体图像质量维持较高水平。
要点
- DWI DL采集时间约缩短一半(2:44 vs 5:02分钟),ADC值无显著差异,可常规用于乳腺MRI方案。
- DWI DL病变对比度更高、病变显著性评分更优,但乳腺组织SNR较低,阅片需注意伪影评分较高。
- 良恶性病变ADC值在两种序列间无显著差异,DWI DL不改变基于ADC的病变定性判断。
关键影像表现/征象
- DWI DL所有病变类型的对比度显著高于DWI STD(P < 0.001),而SNR和对比噪声比在两种序列间无显著差异。
- DWI DL的病变显著性最高评分更常见(P < 0.001),但伪影评分更高(P < 0.001),总体未发现额外伪影。
- DWI STD的主观图像质量高评分(29/65)多于DWI DL(20/65),差异有统计学意义(P < 0.001)。
诊断标准与关键数值
- 浸润性乳腺癌平均ADC值:DWI STD 0.77×10⁻³±0.13 mm²/s,DWI DL 0.75×10⁻³±0.12 mm²/s,无显著差异(P = 0.32)。
- 良性病变平均ADC值:DWI STD 1.32×10⁻³±0.48 mm²/s,DWI DL 1.39×10⁻³±0.54 mm²/s(P = 0.12)。
- 囊肿平均ADC值:DWI STD 2.18×10⁻³±0.49 mm²/s,DWI DL 2.31×10⁻³±0.43 mm²/s。
- 平均采集时间:DWI STD 5:02分钟,DWI DL 2:44分钟(P < 0.001)。
- 观察者间信度k = 0.68~1.0,为充分至优秀。
临床应用与阅片要点
- DWI DL可显著缩短乳腺MRI扫描时间,适合纳入常规乳腺MRI协议以提升效率。
- 阅片时注意DWI DL伪影评分较高,但未发现额外伪影类型,仍需结合ADC图及动态增强序列综合判断。
- DWI DL病变显著性更优,有助于病灶检出,但乳腺组织SNR较低,评估背景实质时需谨慎。
局限性
- 单中心前瞻性研究,样本量65例,乳腺癌患病率23%,结果外推需多中心验证。
- 未报告DWI DL的敏感度、特异度或AUC等诊断效能指标。
- 伪影评分较高虽未发现额外伪影,但可能影响部分图像的主观质量评价。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: Diffusion-weighted imaging (DWI) enhances specificity in multiparametric breast MRI but is associated with longer acquisition time. Deep learning (DL) reconstruction may significantly shorten acquisition time and improve spatial resolution. In this prospective study, we evaluated acquisition time and image quality of a DL-accelerated DWI sequence with superresolution processing (DWI DL ) in comparison to standard imaging including analysis of lesion conspicuity and contrast of invasive breast cancers (IBCs), benign lesions (BEs), and cysts.
MATERIALS AND METHODS: This institutional review board-approved prospective monocentric study enrolled participants who underwent 3 T breast MRI between August and December 2022. Standard DWI (DWI STD ; single-shot echo-planar DWI combined with reduced field-of-view excitation; b-values: 50 and 800 s/mm 2 ) was followed by DWI DL with similar acquisition parameters and reduced averages. Quantitative image quality was analyzed for region of interest-based signal-to-noise ratio (SNR) on breast tissue. Apparent diffusion coefficient (ADC), SNR, contrast-to-noise ratio, and contrast (C) values were calculated for biopsy-proven IBCs, BEs, and for cysts. Two radiologists independently assessed image quality, artifacts, and lesion conspicuity in a blinded independent manner. Univariate analysis was performed to test differences and interrater reliability.
RESULTS: Among 65 participants (54 ± 13 years, 64 women) enrolled in the study, the prevalence of breast cancer was 23%. Average acquisition time was 5:02 minutes for DWI STD and 2:44 minutes for DWI DL ( P < 0.001). Signal-to-noise ratio measured in breast tissue was higher for DWI STD ( P < 0.001). The mean ADC values for IBC were 0.77 × 10 -3 ± 0.13 mm 2 /s in DWI STD and 0.75 × 10 -3 ± 0.12 mm 2 /s in DWI DL without significant difference when sequences were compared ( P = 0.32). Benign lesions presented with mean ADC values of 1.32 × 10 -3 ± 0.48 mm 2 /s in DWI STD and 1.39 × 10 -3 ± 0.54 mm 2 /s in DWI DL ( P = 0.12), and cysts presented with 2.18 × 10 -3 ± 0.49 mm 2 /s in DWI STD and 2.31 × 10 -3 ± 0.43 mm 2 /s in DWI DL . All lesions presented with significantly higher contrast in the DWI DL ( P < 0.001), whereas SNR and contrast-to-noise ratio did not differ significantly between DWI STD and DWI DL regardless of lesion type. Both sequences demonstrated a high subjective image quality (29/65 for DWI STD vs 20/65 for DWI DL ; P < 0.001). The highest lesion conspicuity score was observed more often for DWI DL ( P < 0.001) for all lesion types. Artifacts were scored higher for DWI DL ( P < 0.001). In general, no additional artifacts were noted in DWI DL . Interrater reliability was substantial to excellent (k = 0.68 to 1.0).
CONCLUSIONS: DWI DL in breast MRI significantly reduced scan time by nearly one half while improving lesion conspicuity and maintaining overall image quality in a prospective clinical cohort.
引用
- Wilpert C, Neubauer C, Rau A, et al. Accelerated Diffusion-Weighted Imaging in 3 T Breast MRI Using a Deep Learning Reconstruction Algorithm With Superresolution Processing: A Prospective Comparative Study. Invest Radiol. 2023. PMID: 37428618. doi:10.1097/RLI.0000000000000997
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。